510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

YSIO

K081722510(k) TraditionalCleared Aug 22, 200865-day reviewClass II21 CFR 892.1680Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · KPR System, X-Ray, Stationary

Cleared on Aug 22, 2008 as substantially equivalent to K061054 (Siemens Medical Solutions USA, Inc., 2006) and 2 other predicates after a 65-day review. Its family tree reaches back 8 generations to 1980. 43 later devices descend from it. It has been recalled 11 times.

Decided
Aug 22, 2008
Review
65 days
Class
Class II
Product code
KPR
Recalls
11
Predicates
4

Its longest line of descent runs back 8 generations to K801623 (1980). 63 ancestors in all. 43 later devices descend from it, 7 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2006: 1 other ancestor2005: 1 other ancestor2002: 5 other ancestors1999: 3 other ancestors2004: 2 other ancestors2003: 1 other ancestor2001: 1 other ancestor2000: 3 other ancestors1997: 4 other ancestors1995: 13 other ancestors1994: 2 other ancestors1996: 4 other ancestors1993: 3 other ancestors1992: 2 other ancestors1989: 1 other ancestor1986: 1 other ancestor1991: 4 other ancestors1988: 2 other ancestors1981: 1 other ancestor2022: 3 descendants cleared2011: 1 descendant cleared2019: 6 descendants cleared2014: 3 descendants cleared2012: 1 descendant cleared2017: 1 descendant cleared2020: 4 descendants cleared2015: 2 descendants cleared2016: 3 descendants cleared2018: 3 descendants cleared2023: 4 descendants cleared2024: 2 descendants cleared2025: 4 descendants cleared2026: 3 descendants cleared2021: 3 descendants clearedK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000×2: 2 generations cleared close together, K052202 (2006), K061054 (2006) K062623 · Axiom Luminos DRF · Siemens Medical Solutions USA, Inc. · 2007 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 6 other ancestors3 generations back: 9 other ancestors4 generations back: 10 other ancestors5 generations back: 16 other ancestors6 generations back: 9 other ancestors7 generations back: 4 other ancestors 1 generation forward: 7 descendants2 generations forward: 10 descendants3 generations forward: 10 descendants4 generations forward: 6 descendants5 generations forward: 3 descendants6 generations forward: 2 descendants7 generations forward: 2 descendants8 generations forward: 3 descendants K062623 · Axiom Luminos DRF · Siemens Medical Solutions USA, Inc. · 2007 · recalledK061054 · Axiom Aristos FX Plus · Siemens Medical Solutions USA, Inc. · 2006 · recalledK052202 · Axiom Artis - Modular Angiographic System · Siemens Medical Solutions USA, Inc. · 2006 · recalledK994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−50+5
  1. Root · 8 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 7 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 6 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 5 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 4 generations backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. 3 generations backK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  7. 2 generations backK061054 Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  8. 1 generation backK062623 Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · Recalled
  9. This deviceK081722 YSIOSiemens Medical Solutions USA, Inc. · 2008
  10. Descendant · cites it directly · 32 of its ownK121513 Multix FusionSiemens Medical Solution · 2012 · Recalled
  11. Descendant · cites it directly · 23 of its ownK133259 Ysio MaxSiemens Medical Solutions USA, Inc. · 2014 · Recalled
  12. Descendant · 2 generations on · 18 of its ownK142049 Multix Fusion VA30 & VA40Siemens Medical Solutions USA, Inc. · 2014
  13. Descendant · 3 generations on · 10 of its ownK162971 Multix Fusion MaxSiemens Medical Solutions USA, Inc. · 2016
  14. Descendant · 4 generations on · 8 of its ownK182517 MULTIX ImpactSiemens Medi Cal Solutions, Inc. · 2019

Showing 5 of 43 descendants. See all descendants

Descendants cleared per year

2011: 12012: 12014: 32015: 22016: 32017: 12018: 32019: 62020: 42021: 32022: 32023: 42024: 22025: 42026: 3
2011high 6 (2019) · low 0 (2013)2026

Most influential descendants

Showing 5 of 43 · See all
  • K121513Multix FusionSiemens Medical Solution · 2012 · cites it directly32descendants
  • K133259Ysio MaxSiemens Medical Solutions USA, Inc. · 2014 · cites it directly23descendants
  • K142049Multix Fusion VA30 & VA40Siemens Medical Solutions USA, Inc. · 2014 · 2 generations on18descendants
  • K162971Multix Fusion MaxSiemens Medical Solutions USA, Inc. · 2016 · 3 generations on10descendants
  • K182517MULTIX ImpactSiemens Medi Cal Solutions, Inc. · 2019 · 4 generations on8descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−7K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−6K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−5K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−4K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
−3K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.200629
−2K061054Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc.20063
−1K062623Axiom Luminos DRFSiemens Medical Solutions USA, Inc.200720
0 (this device)K081722YSIOSiemens Medical Solutions USA, Inc.2008—
+1K121513Multix FusionSiemens Medical Solution20124
+1K133259Ysio MaxSiemens Medical Solutions USA, Inc.20147
+2K142049Multix Fusion VA30 & VA40Siemens Medical Solutions USA, Inc.2014—
+3K162971Multix Fusion MaxSiemens Medical Solutions USA, Inc.2016—
+4K182517MULTIX ImpactSiemens Medi Cal Solutions, Inc.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198610
198820
198910
199140
199220
199330
199420
1995130
199640
199740
199930
200030
200110
200250
200310
200420
200510
200610
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201210
201430
201520
201630
201710
201830
201960
202040
202130
202230
202340
202420
202540
202630
FIG. 1 — Lineage of K081722, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPR device

  1. #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK983732FDA’s K983732 is Thorax FD and Multix FD
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
Siemens Medical Solutions USA, Inc.19991
*K061054Axiom Aristos FX Plus
Summary page 2
…following STEMENS Medical Systems devices: _510(!() Number Date of Clearance Device Name ‘ KY83732 April 21, 1999 Thorax FD and Multix FD (AXIOM Aristos TX / MX) K061054 May 09, 2006 AXIOM Aristos FX Plus (Wircless Remote) K062623 August 22, 2007 AXIOM Luminos dRF Many of the components used…
Siemens Medical Solutions USA, Inc.20063
*K062623Axiom Luminos DRF
Summary page 2
…e of Clearance Device Name ‘ KY83732 April 21, 1999 Thorax FD and Multix FD (AXIOM Aristos TX / MX) K061054 May 09, 2006 AXIOM Aristos FX Plus (Wircless Remote) K062623 August 22, 2007 AXIOM Luminos dRF Many of the components used on the Ysio i.e. x-ray generator, x-ray tube, user interface, di…
Siemens Medical Solutions USA, Inc.200720
*K070505Krystal-X Wifi
Summary page 2
…uivalent to the wireless interface of the Krystal-X WiFi wireless digital X-ray sensor from Video Dental Concepts, Inc. 510{(k) Number Date of Clearance Device Name K070505 April 5, 2007 Krystal-X WiFi I A dectailed Substantial Equivalence Comparison is provided in Section 12. 510(k) Ysio Page 10…
Owandy2007—

Also cleared against K061054: 3 other devices →

What descends from it? (43)

See all 43

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K121513Multix FusionSiemens Medical Solution201232
K133259Ysio MaxSiemens Medical Solutions USA, Inc.201423
K162529DigiX FDXAllengers Medical Systems Limited20176
K192541DigiX FDXAllengers Medical Systems Limited20195
K223060DigiX FDXAllengers Medical Systems Limited20224
K111912Mobilett MiraSiemens Medical Solutions USA, Inc.20113
K220149General Radiography X-ray System (Model: GR10X 40K), General Radiography X-ray System…Xera Medical Systems & Technology, Ltd.20220

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 31, 2022Class IIZ-0841-2022Software designSiemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If the user activates a…
Feb 10, 2020Class IIZ-1501-2020Component design/selectionThe Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head or the remote…
Aug 6, 2018Class IIZ-0299-2019Device DesignIn rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope…
Aug 27, 2014Class IIZ-2680-2014Under Investigation by firmSiemens discovered that an unlikely error may occur on the Ysio system with fixed detector in the wall stand. This error may result in line artifacts in the image. If these artifacts appear in the region of interest, the examination may…
Jul 7, 2014Class IIZ-2292-2014Device DesignIt was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in longer radiation…
May 9, 2014Class IIZ-1717-2014Use errorExcessive usage of cleaning liquid on Siemens Ysio Systems may cause infiltration of the generator console, which may lead to a short circuit causing an unintended release of radiation.
Sep 3, 2013Class IIZ-0374-2014Software designA potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required and overexposed…
Aug 28, 2013Class IIZ-1845-2014Software designWhen using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in the image appearing very dark, which may prompt unnecessary repetition of the examination.
May 6, 2013Class IIZ-0016-2014Software designSiemens has discovered through product monitoring a potential malfunction when using system operating with software versions VB10D, VB10F, VB10C and VB10. During a RAD examination the possibility exists that radiation may continue longer…
Feb 15, 2012Not classifiedZ-1159-2012Software designFirm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A. Due to an error in the image pipeline, images may display dark…
Apr 28, 2011Not classifiedZ-2787-2011OtherThere is a potential hazard to staff or patients when lowering the table height on the Ysio system. If staff or patients lower extremities are placed underneath the table and the user lowers the table top, it is possible that parts of the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 5 companies; a new generation was cleared every 1.5 years on median.
  • 18 of the 63 ancestors have a recall of any class.
  • Of its descendants, K121513 (Siemens Medical Solution, 2012) has gone furthest, with 32 descendants of its own.
  • 3 other devices also cite its predicate K061054.

How does its category compare?

All KPR devices

KPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · KPR

8in 2025

2007: 122008: 152009: 92010: 92011: 92012: 112013: 52014: 102015: 182016: 112017: 62018: 152019: 122020: 162021: 62022: 82023: 92024: 112025: 82026 (partial year): 6
2007high 18 (2015) · low 5 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPR

155in 2025

2007: 31 days median2008: 48 days median2009: 35 days median2010: 126 days median2011: 152 days median2012: 58 days median2013: 163 days median2014: 108 days median2015: 81 days median2016: 84 days median2017: 75 days median2018: 30 days median2019: 33 days median2020: 39 days median2021: 55 days median2022: 43 days median2023: 76 days median2024: 162 days median2025: 155 days median2026 (partial year): 197 days median
2007high 197 (2026) · low 30 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPR

7in 2025

2007: 102008: 102009: 52010: 32011: 52012: 12013: 52014: 82015: 12016: 42017: 32018: 42019: 12020: 12022: 22023: 42024: 12025: 7
2007high 10 (2007) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPR

39in 2025

2007: 1202008: 952009: 612010: 822011: 1032012: 742013: 242014: 342015: 592016: 182017: 312018: 362019: 772020: 412021: 372022: 272023: 392024: 342025: 39
2007high 120 (2007) · low 18 (2016)2025

MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123110120
200815481095
2009935561
20109126382
201191525103
20121158174
20135163524
201410108834
20151881159
20161184418
2017675331
20181530436
20191233177
20201639141
2021655037
2022843227
2023976439
202411162134
20258155739
202661970—

What is it indicated for?

The Ysio (New RAD -FAMILY) systems are radiographic systems used in hospitals, clinics, and medical practices. Ysio enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with…

Read the full statement (117 words)

The Ysio (New RAD -FAMILY) systems are radiographic systems used in hospitals, clinics, and medical practices. Ysio enables radiographic and tomographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatric patients. It can also be used for intravenous, small interventions (like biopsy, punctures, etc.) and emergency (trauma, critical ill) applications. Exposures may be taken with the patient sitting, standing, or in the prone position. The Ysio system is not meant for mammography.

The Ysio uses an integrated or portable digital detector for generating diagnostic images by converting x-rays into electronic signals. Ysio is also designed to be used with conventional film/screen or Computed Radiography (CR) casseltes.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K081722: YSIO.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K081722

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026