510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Axiom Artis - Modular Angiographic System

K052202510(k) TraditionalCleared Mar 7, 2006207-day reviewClass II21 CFR 892.1650Recalls 29

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Mar 7, 2006 as substantially equivalent to K994231 (Siemens Medical Corp., 2000) and 1 other predicate after a 207-day review. Its family tree reaches back 5 generations to 1980. 4 of the 188 devices in its family, counting itself, have a Class I recall. 154 later devices descend from it. It has been recalled 29 times.

Decided
Mar 7, 2006
Review
207 days
Class
Class II
Product code
OWB
Recalls
29
Predicates
3

Its longest line of descent runs back 5 generations to K801623 (1980). 33 ancestors in all. 154 later devices descend from it, 2 citing it directly. 4 of the 188 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2004: 2 other ancestors2002: 2 other ancestors2000: 2 other ancestors1999: 1 other ancestor1997: 2 other ancestors1996: 3 other ancestors1995: 9 other ancestors1994: 1 other ancestor1993: 1 other ancestor1991: 3 other ancestors1992: 1 other ancestor1988: 1 other ancestor2006: 1 descendant cleared2008: 5 descendants cleared2015: 7 descendants cleared2007: 1 descendant cleared2009: 3 descendants cleared2010: 5 descendants cleared2013: 10 descendants cleared2011: 5 descendants cleared2018: 11 descendants cleared2022: 10 descendants cleared2019: 7 descendants cleared2014: 11 descendants cleared2012: 4 descendants cleared2017: 6 descendants cleared, 2 with a Class I recall22024: 7 descendants cleared2016: 11 descendants cleared, 1 with a Class I recall12020: 12 descendants cleared, 1 with a Class I recall12021: 9 descendants cleared2025: 13 descendants cleared2023: 7 descendants cleared2026: 9 descendants clearedK801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996×2: 2 generations cleared close together, K972114 (1998), K981993 (1998) K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 200019801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 11 other ancestors4 generations back: 9 other ancestors5 generations back: 4 other ancestors 1 generation forward: 2 descendants2 generations forward: 7 descendants3 generations forward: 25 descendants4 generations forward: 34 descendants5 generations forward: 36 descendants6 generations forward: 22 descendants, 2 with a Class I recall27 generations forward: 14 descendants, 1 with a Class I recall18 generations forward: 5 descendants, 1 with a Class I recall19 generations forward: 4 descendants10 generations forward: 5 descendants K994231 · Siemens Integrated Operating System (Sios) · Siemens Medical Corp. · 2000K981993 · Olympus Integrated Endosurgery System Endoalpha · The Olympus Optical Co. · 1998K972114 · Olympus Sonosurg System · The Olympus Optical Co. · 1998K962952 · Olympus Ultrasonic Surgical System (Usu) · Olympus Optical, Co. · 1996K801623 · Cavitron Ultrasonic Surg. Aspir. Cusa · Cavitron Corp. · 1980−50+5+10
  1. Root · 5 generations backK801623 Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp. · 1980
  2. 4 generations backK962952 Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co. · 1996
  3. 3 generations backK972114 Olympus Sonosurg SystemThe Olympus Optical Co. · 1998
  4. 2 generations backK981993 Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co. · 1998
  5. 1 generation backK994231 Siemens Integrated Operating System (Sios)Siemens Medical Corp. · 2000
  6. This deviceK052202 Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc. · 2006
  7. Descendant · cites it directly · 80 of its ownK061054 Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  8. Descendant · 2 generations on · 78 of its ownK062623 Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · Recalled
  9. Descendant · cites it directly · 72 of its ownK073290 Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  10. Descendant · 2 generations on · 69 of its ownK090745 Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · Recalled
  11. Descendant · 2 generations on · 43 of its ownK081722 YSIOSiemens Medical Solutions USA, Inc. · 2008 · Recalled

Showing 5 of 154 descendants. See all descendants

Descendants cleared per year

2006: 12007: 12008: 52009: 32010: 52011: 52012: 42013: 102014: 112015: 72016: 112017: 62018: 112019: 72020: 122021: 92022: 102023: 72024: 72025: 132026: 9
2006high 13 (2025) · low 1 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 154 · See all
  • K061054Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · cites it directly80descendants
  • K062623Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · 2 generations on78descendants
  • K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc. · 2008 · cites it directly72descendants
  • K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc. · 2009 · 2 generations on69descendants
  • K081722YSIOSiemens Medical Solutions USA, Inc. · 2008 · 2 generations on43descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K801623Cavitron Ultrasonic Surg. Aspir. CusaCavitron Corp.1980—
−4K962952Olympus Ultrasonic Surgical System (Usu)Olympus Optical, Co.1996—
−3K972114Olympus Sonosurg SystemThe Olympus Optical Co.1998—
−2K981993Olympus Integrated Endosurgery System EndoalphaThe Olympus Optical Co.1998—
−1K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
0 (this device)K052202Axiom Artis - Modular Angiographic SystemSiemens Medical Solutions USA, Inc.2006—
+1K061054Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc.20063
+2K062623Axiom Luminos DRFSiemens Medical Solutions USA, Inc.200720
+1K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200839
+2K090745Artis Zee / Zeegos with SW VC14Siemens Medical Solutions USA, Inc.20097
+2K081722YSIOSiemens Medical Solutions USA, Inc.200811

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199130
199210
199310
199410
199590
199630
199720
199910
200020
200220
200420

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200710
200850
200930
201050
201150
201240
2013100
2014110
201570
2016111
201762
2018110
201970
2020121
202190
2022100
202370
202470
2025130
202690
FIG. 1 — Lineage of K052202, 1980–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK031359FDA’s K031359 is S-Monovette Edta K2-Gel
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
Sarstedt, Inc.2003—
*K994231Siemens Integrated Operating System (Sios)Siemens Medical Corp.2000—
*K042540Philips IU22 Ultrasound SystemPhilips Ultrasound, Inc.20044

Also cleared against K994231: 1 other device →

What descends from it? (154)

See all 154

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K061054Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc.200680
K073290Artis Zee and ZeegoSiemens Medical Solutions USA, Inc.200872

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 30, 2021Class IIZ-1496-2021Equipment maintenanceDue to inappropriate cleaning, some Artis systems show unexpected corrosion of visible belts which are needed to move system parts (e.g. C-Arm), Increased corrosion may lead to a malfunction of the belts which may result in limited…
Jun 12, 2019Class IIZ-2018-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2019-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2016-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2023-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2027-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2015-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2025-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2013-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2024-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2012-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2017-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2014-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2026-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2020-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2021-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Jun 12, 2019Class IIZ-2022-2019Component design/selectionA tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue…
Sep 28, 2017Class IIZ-0711-2018Device DesignThe possibility exists for the wireless foot switch to fail due to impacts by external factors, such as electrostatic discharge that exceed a certain intensity. If the foot switch fails, it will no longer be able to be used to release…
Jun 6, 2017Class IIZ-0708-2018Device DesignPotential for liquid entry
Feb 27, 2017Class IIZ-1468-2017Process controlA gap in the housing of the wireless foot switch could potentially result in liquids penetrating the interior. Such liquids may include disinfectants, cleaning agents and/ or bodily fluids. In rare cases, the introduction of liquids in…
Nov 1, 2016Class IIZ-0603-2017Software in the Use EnvironmentThe possibility exists that the semi-automatic focus switch may not work as intended when pressing the footswitch multiple times.
Jan 16, 2012Not classifiedZ-0927-2012Software change controlThere is a potential misalignment in the connection of the remote foot switch and the stationary unit. This potential malfunction may occur if either the wireless foot switch or the stationary unit is equipped with improper firmware…
May 1, 2009Not classifiedZ-1441-2009OtherImproper installation of accessory support arm, which could potentially result in the loosening of the arm over time causing it to fall.
Aug 25, 2008Not classifiedZ-0316-2009Software designSystem may switch to emergency fluoroscopy mode.
Aug 25, 2008Not classifiedZ-0313-2009Software designSystem may switch to emergency fluoroscopy mode.
Aug 25, 2008Not classifiedZ-0315-2009Software designSystem may switch to emergency fluoroscopy mode.
Aug 25, 2008Not classifiedZ-0312-2009Software designSystem may switch to emergency fluoroscopy mode.
Aug 25, 2008Not classifiedZ-0314-2009Software designSystem may switch to emergency fluoroscopy mode.
Aug 25, 2008Not classifiedZ-0317-2009Software designSystem may switch to emergency fluoroscopy mode.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K162859Allura Xper FD series and Allura Xper OR Table seriesPhilips Medical Systems Nederland B.V. · 2016 · 6 generations on · ◆ in the 2016 descendants bar of FIG. 112 Class I recalls
  • K163715Allura Xper R9Philips Medical Systems Nederland B.V. · 2017 · 6 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls
  • K172822Azurion series R1.2Philips Medical Systems Nederland B.V. · 2017 · 7 generations on · ◆ in the 2017 descendants bar of FIG. 19 Class I recalls
  • K200917Azurion R2.1Philips Medical Systems Nederland B.V. · 2020 · 8 generations on · ◆ in the 2020 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 26 years and passes through 5 companies; a new generation was cleared every 2.0 years on median.
  • 1 of the 33 ancestors has a recall of any class.
  • Of its descendants, K061054 (Siemens Medical Solutions USA, Inc., 2006) has gone furthest, with 80 descendants of its own.
  • 1 other device also cites its predicate K994231.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Procedures that can be performed with the AXIOM Artis family include cardiac angiography, neuro-angiography, general angiography, operating room angiography, multipurpose angiography and radiographic/fluoroscopic procedures e.g. gastro-intestinal imaging, skeletal imaging etc. AXIOM Artis can also support the…

Read the full statement (100 words)

AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures.

Procedures that can be performed with the AXIOM Artis family include cardiac angiography, neuro-angiography, general angiography, operating room angiography, multipurpose angiography and radiographic/fluoroscopic procedures e.g. gastro-intestinal imaging, skeletal imaging etc.

AXIOM Artis can also support the acquisition of position triggered imaging for spatial data synthesis.

The intended use and indications for use of the AXIOM Artis with wireless footswitch and with voice control as described in its labeling have not changed from its predicate device the AXIOM Artis Modular Angiographic System.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052202: Axiom Artis - Modular Angiographic System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K052202

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026