Strata NSC Burr Hole Valve, Model 42335; Shunt Assembly, Model 46635

K082127510(k) SpecialCleared Aug 27, 200829-day reviewClass II21 CFR 882.5550Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Neurosurgery · JXG Shunt, Central Nervous System And Components

Cleared on Aug 27, 2008 as substantially equivalent to K033850 (Medtronic Neurosurgery, 2004) and 2 other predicates after a 29-day review. Its family tree reaches back 6 generations to 1995. 1 later device descends from it. It has been recalled once.

Decided
Aug 27, 2008
Review
29 days
Class
Class II
Product code
JXG
Recalls
1
Predicates
3

Its longest line of descent runs back 6 generations to K944222 (1995). 13 ancestors in all, the oldest K831678 (1983). 1 later device descends from it, 1 citing it directly.

  1. Root · 6 generations backK944222 Codman-Medos Valve Shunt SystemJohnson & Johnson Professionals, Inc. · 1995
  2. 5 generations backK973774 Codman Hakim Micro Precision ValveJohnson & Johnson Professionals, Inc. · 1997
  3. 4 generations backK980778 Hakim Micro Programmable Valve SystemJohnson & Johnson Professionals, Inc. · 1998
  4. 3 generations backK012052 Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical · 2002
  5. 2 generations backK033850 Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and…Medtronic Neurosurgery · 2004 · Recalled
  6. 1 generation backK042465 PS Medical Strata II ValveMedtronic Neurosurgery · 2004 · Recalled
  7. This deviceK082127 Strata NSC Burr Hole Valve, Model 42335; Shunt Assembly, Model 46635Medtronic Neurosurgery · 2008
  8. Descendant · cites it directly · 0 of its ownK083076 Strata Burr Hole Valve, Model 42836 and Strata Burr Hole Shunt…Medtronic Neurosurgery · 2008 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K083076Strata Burr Hole Valve, Model 42836 and Strata Burr Hole Shunt Assembly, Model 46836Medtronic Neurosurgery · 2008 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K944222Codman-Medos Valve Shunt SystemJohnson & Johnson Professionals, Inc.1995—
−5K973774Codman Hakim Micro Precision ValveJohnson & Johnson Professionals, Inc.1997—
−4K980778Hakim Micro Programmable Valve SystemJohnson & Johnson Professionals, Inc.1998—
−3K012052Medtronic PS Medical Strata Valve, Models 42855, 42865, 92865, 92855Medtronic PS Medical2002—
−2K033850Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and Without…Medtronic Neurosurgery20042
−1K042465PS Medical Strata II ValveMedtronic Neurosurgery20044
0 (this device)K082127Strata NSC Burr Hole Valve, Model 42335; Shunt Assembly, Model 46635Medtronic Neurosurgery2008—
+1K083076Strata Burr Hole Valve, Model 42836 and Strata Burr Hole Shunt Assembly, Model…Medtronic Neurosurgery20081

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
198610
199010
199510
199810
200010
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
FIG. 1 — Lineage of K082127, 1995–2008Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JXG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JXG device

  1. #1 K243531 Codman Libertís Shunt Catheter with Bactiseal…score 11.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K033850Medtronic PS Medical Strata NSC Valve and Shunt Assemblies with and Without Bioglide
Our reading of the summary gave K038850, which is not a number in FDA’s lists. Corrected to the number one digit away whose FDA name carries the summary’s words for the device, “Medtronic PS Medical Strata NSC Valve and Shunt Assemblies”.
Medtronic Neurosurgery20042
*K042465PS Medical Strata II Valve
Summary page 1
…edtronic PS Medical Strata IIValves and Shunt Assemblies with and without BioGlide (K042465) * CSF-Flow Control Valve, Burr Hole (K831678…
Medtronic Neurosurgery20044
*K831678P-S Medical CSF-Flow Control Valve-Bur
Summary page 1
…BioGlide (K042465) * CSF-Flow Control Valve, Burr Hole (K831678) Description: The PS Medical® Strata® NSC Burr Hole Valve is an adjustable non-siphon control valve. The valve is manufactured o…
Pudenz-Schulte Medical Research Corp.19831

Also cleared against K033850: 5 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K083076Strata Burr Hole Valve, Model 42836 and Strata Burr Hole Shunt Assembly, Model 46836Medtronic Neurosurgery20080

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 17, 2017Class IIZ-1121-2017OtherMedtronic Neurosurgery initiated a recall after receiving five (5) Product Experiences between February 15, 2015 and August 23, 2016 reporting of discrepancies between the pressure/performance level (PL) at which the valve was set and the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 2.4 years on median.
  • 5 of the 13 ancestors have a recall of any class.
  • Of its descendants, K083076 (Medtronic Neurosurgery, 2008) has gone furthest, with 0 descendants of its own.
  • 5 other devices also cite its predicate K033850.

How does its category compare?

All JXG devices

JXG Shunt, Central Nervous System And Components: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · JXG

8in 2025

2007: 32008: 62009: 52010: 32011: 72012: 32013: 32015: 62016: 62017: 72018: 62019: 42020: 22021: 12022: 42023: 52024: 32025: 82026 (partial year): 2
2007high 8 (2025) · low 0 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JXG

255in 2025

2007: 184 days median2008: 60 days median2009: 106 days median2010: 30 days median2011: 136 days median2012: 78 days median2013: 108 days median2015: 276 days median2016: 143 days median2017: 90 days median2018: 102 days median2019: 101 days median2022: 101 days median2023: 134 days median2024: 90 days median2025: 255 days median
2007high 276 (2015) · low 30 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JXG

3in 2024

2008: 12009: 302011: 52012: 12013: 142014: 62017: 12020: 82023: 22024: 32026 (partial year): 1
2007high 30 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JXG

1,198in 2025

2007: 6492008: 6302009: 6292010: 8122011: 7872012: 9772013: 1,0042014: 1,2512015: 9842016: 1,0912017: 1,3352018: 1,4842019: 1,2952020: 1,2062021: 1,4672022: 1,3292023: 1,2212024: 1,2142025: 1,198
2007high 1,484 (2018) · low 629 (2009)2025

MAUDE reports naming product code JXG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731840649
20086601630
2009510630629
20103300812
201171365787
20123781977
20133108141,004
20140—61,251
201562760984
2016614301,091
201779011,335
2018610201,484
2019410101,295
20202—81,206
20211—01,467
2022410101,329
2023513421,221
202439031,214
2025825501,198
20262—1—

What is it indicated for?

The Strata NSC Burr Hole Valves and Shunt Assemblies are shunt components designed to provide continued cerebrospinal fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The Strata NSC Burr Hole Valve allows the physician to non-invasively adjust the pressure/ flow performance level pre- and post-implantation without the need for radiographic confirmation inorder to address changing patient needs.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K082127: Strata NSC Burr Hole Valve, Model 42335; Shunt Assembly, Model 46635.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K082127

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026