510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Amicas Pacs, Version 6.0

K082144510(k) TraditionalCleared Sep 5, 200837-day reviewClass II21 CFR 892.2050Recalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Amicas, Inc. · LLZ System, Image Processing, Radiological

Cleared on Sep 5, 2008 as substantially equivalent to K073526 (Amicas, Inc., 2008) after a 37-day review. Its family tree reaches back 6 generations to 1996. 3 later devices descend from it. It has been recalled 13 times.

Decided
Sep 5, 2008
Review
37 days
Class
Class II
Product code
LLZ
Recalls
13
Predicates
1

Its longest line of descent runs back 6 generations to K954678 (1996). 3 later devices descend from it, 2 citing it directly.

  1. Root · 6 generations backK954678 Pro VisionAlgotec Systems, Ltd. · 1996
  2. 5 generations backK980648 Pro VisionAlgotec Systems, Ltd. · 1998
  3. 4 generations backK992654 Plug 'N View 3D, Version 1.0Voxar Limited · 1999
  4. 3 generations backK022970 Amicas Light Beam Workstation, Model 1.0Amicas, Inc. · 2002
  5. 2 generations backK062477 Amicas Vision Series Pacs, Model 4.3Amicas, Inc. · 2006
  6. 1 generation backK073526 Amicas Vision Series Pacs, Model 5.5Amicas, Inc. · 2008
  7. This deviceK082144 Amicas Pacs, Version 6.0Amicas, Inc. · 2008
  8. Descendant · cites it directly · 1 of its ownK173475 Merge PACSMerge Healthcare Incorporated · 2017 · Recalled
  9. Descendant · 2 generations on · 0 of its ownK192455 Merge PACSMerge Healthcare Incorporated · 2019 · Recalled
  10. Descendant · cites it directly · 0 of its ownK124048 ImagequbeIntuitive Imaging Informatics, LLC · 2013

Most influential descendants

Showing 3 of 3 · See all
  • K173475Merge PACSMerge Healthcare Incorporated · 2017 · cites it directly1descendant
  • K192455Merge PACSMerge Healthcare Incorporated · 2019 · 2 generations on0descendants
  • K124048ImagequbeIntuitive Imaging Informatics, LLC · 2013 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K954678Pro VisionAlgotec Systems, Ltd.1996—
−5K980648Pro VisionAlgotec Systems, Ltd.1998—
−4K992654Plug 'N View 3D, Version 1.0Voxar Limited1999—
−3K022970Amicas Light Beam Workstation, Model 1.0Amicas, Inc.2002—
−2K062477Amicas Vision Series Pacs, Model 4.3Amicas, Inc.2006—
−1K073526Amicas Vision Series Pacs, Model 5.5Amicas, Inc.2008—
0 (this device)K082144Amicas Pacs, Version 6.0Amicas, Inc.2008—
+1K173475Merge PACSMerge Healthcare Incorporated20172
+2K192455Merge PACSMerge Healthcare Incorporated20194
+1K124048ImagequbeIntuitive Imaging Informatics, LLC2013—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201710
201910
FIG. 1 — Lineage of K082144, 1996–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K073526Amicas Vision Series Pacs, Model 5.5
Summary page 1
…AMICAS Inc. 20 Guest Street Boston, MA 021235 Predicate Device 5 10(k) number: K073526 Date received: 12/17/2007 Decision date: 03/12/2008 Decision:…
Amicas, Inc.2008—

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K173475Merge PACSMerge Healthcare Incorporated20171
K124048ImagequbeIntuitive Imaging Informatics, LLC20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 24, 2017Class IIZ-2601-2017Software design (manufacturing process)Studies that have QC, PDE, or HL7 ADT updates applied may fail to propagate via telmed protocol to the destination server and may fail to import on the destination server.
May 8, 2017Class IIZ-2590-2017Software design (manufacturing process)Merge PACS did not show unviewed images when the last view was skipped. There is a potential risk to health of a physician misdiagnosis because not all images are available for viewing.
May 8, 2017Class IIZ-2591-2017Software design (manufacturing process)When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.
Apr 4, 2016Class IIZ-1431-2017Software designStudies that have QC, PDE, or HL7 ADT updates applied may fail to propagate via telmed protocol to the destination server and may fail to import on the destination server.
Apr 4, 2016Class IIZ-1176-2017Software designPotential exists for an incorrect patient image being displayed which could result in the delay in diagnosis or treatment.
Mar 28, 2016Class IIZ-0599-2017Software designThe software did not show unviewed images when the last view was skipped. The physician may potentially misdiagnose when not all images are available for viewing.
Mar 28, 2016Class IIZ-0726-2017Component design/selectionCut lines on the image may present horizontally rather than vertically.
Feb 5, 2016Class IIZ-2363-2016Software designWhen measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same.
Jan 30, 2016Class IIZ-2301-2016Software designStudies coming over via telmed were missing patient's DOB, procedure, and referring physician.
Jan 30, 2016Class IIZ-2303-2016Software designA migrated study that has annotations will display the annotations, but when a prior study is loaded that should have annotations, the second study will not display annotations.
Jan 30, 2016Class IIZ-2379-2016Software designThe patient name in the Halo title bar and the thumbnails do not match the name on displayed images.
Jan 30, 2016Class IIZ-2531-2016Software designStudies that are viewed in iConnect Access that originate on Merge PACS / Ortho PACS may not display the correct current patient demographics if there has been a PDE (Patient / Study Demographic Edit) performed after the images were…
Jan 30, 2016Class IIZ-2532-2016Software designStudies that are viewed in iConnect Access that originate on Merge PACS / Ortho PACS may not display the correct current patient demographics if there has been a PDE (Patient / Study Demographic Edit) performed after the images were…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 2.5 years on median.
  • None of the 6 ancestors has a recall.
  • Of its descendants, K173475 (Merge Healthcare Incorporated, 2017) has gone furthest, with 1 descendant of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

AMICAS PACS 6.0 is designed and marketed for soft copy reading, communication and storage of studies produced by digital modalities, to include Digital Mammography. AMICAS PACS 6.0 receives images acquired from DICOM-compliant medical imaging systems, data from FDA-cleared Computer-Aided Detection systems and other FDA-cleared Image processing systems. AMICAS PACS 6.0 imports images and render said images, upon request, within the viewer component…

Read the full statement (332 words)

AMICAS PACS 6.0 is designed and marketed for soft copy reading, communication and storage of studies produced by digital modalities, to include Digital Mammography. AMICAS PACS 6.0 receives images acquired from DICOM-compliant medical imaging systems, data from FDA-cleared Computer-Aided Detection systems and other FDA-cleared Image processing systems.

AMICAS PACS 6.0 imports images and render said images, upon request, within the viewer component utilizing both lossless (reversible) and lossy (irreversible) compression.

To support the diagnostic interpretation of Mammography studies, AMICAS PACS 6.0 will display the full fidelity DICOM image in a non-compressed format. Images will be rendered with patient and clinical information clearly displayed as part of the DICOM Overlay as required by MQSA, on monitors cleared by FDA for use in Digital Mammography. Lossy compressed mammography images and digitized film screen images must not be used for the purpose of primary diagnosis. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

Within AMICAS PACS 6.0, the AMICAS Real Time Worklist offers real-time status of radiology activity and provides customizable workflow management capabilities. Communication of critical results is facilitated and documented through optional, and configurable, components within the Real Time Worklist.

AMICAS PACS 6.0 also offers the Patient Dashboard which provides a single view of all patient data, both imaging and non-imaging.

AMICAS Reach is an optional component within AMICAS PACS 6.0 which provides clinicians secure, proactive communication and access to clinical reports and images.

Order and Report information generated by HIS/RIS and report creation systems are received and displayed in PACS via the transmission of HL7 messaging. For this data, AMICAS PACS 6.0 is not the creator, but instead the downstream recipient which relies on the validity of data from said systems.

AMICAS PACS 6.0 must be installed on suitable, commercial-standard hardware. It is the user's responsibility to ensure monitor quality, ambient light conditions and image compression ratios are consistent with the clinical application.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K082144: Amicas Pacs, Version 6.0.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K082144

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.