510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

S5/S5I Series Intravascular Imaging and Pressure System, Combomap Pressure and Flow System

K082229510(k) SpecialCleared Oct 8, 200862-day reviewClass II21 CFR 892.1560Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Volcano Corporation · IYO System, Imaging, Pulsed Echo, Ultrasonic

Cleared on Oct 8, 2008 as substantially equivalent to K071554 (Volcano Corporation, 2008) and 6 other predicates after a 62-day review. Its family tree reaches back 6 generations to 1995. 25 later devices descend from it. It has been recalled 3 times.

Decided
Oct 8, 2008
Review
62 days
Class
Class II
Product code
IYO
Recalls
3
Predicates
7

Its longest line of descent runs back 6 generations to K934139 (1995). 44 ancestors in all, the oldest K823434 (1982). 25 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 2 other ancestors2003: 2 other ancestors2002: 1 other ancestor1998: 3 other ancestors1997: 3 other ancestors2007: 1 other ancestor2006: 1 other ancestor2005: 3 other ancestors2000: 2 other ancestors1996: 3 other ancestors1995: 2 other ancestors1994: 3 other ancestors1993: 3 other ancestors1992: 5 other ancestors1991: 1 other ancestor1990: 1 other ancestor1986: 1 other ancestor1982: 1 other ancestor2011: 1 descendant cleared2013: 4 descendants cleared2012: 1 descendant cleared2014: 1 descendant cleared2015: 3 descendants cleared2019: 4 descendants cleared2018: 1 descendant cleared2017: 3 descendants cleared2024: 1 descendant cleared2025: 1 descendant cleared2022: 1 descendant cleared2020: 1 descendant cleared2023: 1 descendant cleared2021: 1 descendant cleared2026: 1 descendant clearedK934139 · Tomtec Echo-View · Tomtec Tomographic Technologies GmbH · 1995K993398 · Echo View, Easy View, Omni View · Tomtec Imaging Systems GmbH · 1999K022824 · Echo-View 5.X, Easy-View 2.X, Omni View 2.X, Cardio-View 1.X, LV Analysis 1.X & Surgical View 1.X · Tomtec Imaging Systems GmbH · 2002K042188 · Volcano Therapeutics, Inc. Volcano Ivus System · Volcano Therapeutics, Inc. · 2004K051337 · Eagle Eye Gold Ivus Imaging Catheter, Model 85900/859XX; Volcano VH-Ivus System, Model 8049XX-001 · Volcano Corporation · 2005K071554 · Volcano S5/S5I Intravascular Imaging and Pressure System · Volcano Corporation · 2008 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 6 other ancestors2 generations back: 20 other ancestors3 generations back: 5 other ancestors4 generations back: 7 other ancestors 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 4 descendants4 generations forward: 6 descendants5 generations forward: 10 descendants6 generations forward: 2 descendants K071554 · Volcano S5/S5I Intravascular Imaging and Pressure System · Volcano Corporation · 2008 · recalledK051337 · Eagle Eye Gold Ivus Imaging Catheter, Model 85900/859XX; Volcano VH-Ivus System, Model 8049XX-001 · Volcano Corporation · 2005K042188 · Volcano Therapeutics, Inc. Volcano Ivus System · Volcano Therapeutics, Inc. · 2004K022824 · Echo-View 5.X, Easy-View 2.X, Omni View 2.X, Cardio-View 1.X, LV Analysis 1.X & Surgical View 1.X · Tomtec Imaging Systems GmbH · 2002K993398 · Echo View, Easy View, Omni View · Tomtec Imaging Systems GmbH · 1999K934139 · Tomtec Echo-View · Tomtec Tomographic Technologies GmbH · 1995−50+5
  1. Root · 6 generations backK934139 Tomtec Echo-ViewTomtec Tomographic Technologies GmbH · 1995
  2. 5 generations backK993398 Echo View, Easy View, Omni ViewTomtec Imaging Systems GmbH · 1999
  3. 4 generations backK022824 Echo-View 5.X, Easy-View 2.X, Omni View 2.X, Cardio-View 1.X, LV…Tomtec Imaging Systems GmbH · 2002
  4. 3 generations backK042188 Volcano Therapeutics, Inc. Volcano Ivus SystemVolcano Therapeutics, Inc. · 2004
  5. 2 generations backK051337 Eagle Eye Gold Ivus Imaging Catheter, Model 85900/859XX; Volcano…Volcano Corporation · 2005
  6. 1 generation backK071554 Volcano S5/S5I Intravascular Imaging and Pressure SystemVolcano Corporation · 2008 · Recalled
  7. This deviceK082229 S5/S5I Series Intravascular Imaging and Pressure System, Combomap…Volcano Corporation · 2008
  8. Descendant · cites it directly · 24 of its ownK111706 Volcano S5IINTRAVASCULAR Imaging and Pressure SystemVolcano Corporation · 2011 · Recalled
  9. Descendant · 2 generations on · 21 of its ownK123898 Volcano S5/S5I Intravascular Ultrasound Imaging and Pressure SystemVolcano Corporation · 2013 · Recalled
  10. Descendant · 3 generations on · 18 of its ownK133323 Volcano Ifr ModalityVolcano Corporation · 2014
  11. Descendant · 4 generations on · 9 of its ownK150441 Volcano iFR Modality, iFR Scout featureVolcano Corporation · 2015
  12. Descendant · 5 generations on · 5 of its ownK170133 FFR V2.5Philips Volcano · 2017 · Recalled

Showing 5 of 25 descendants. See all descendants

Descendants cleared per year

2011: 12012: 12013: 42014: 12015: 32017: 32018: 12019: 42020: 12021: 12022: 12023: 12024: 12025: 12026: 1
2011high 4 (2013) · low 0 (2016)2026

Most influential descendants

Showing 5 of 25 · See all
  • K111706Volcano S5IINTRAVASCULAR Imaging and Pressure SystemVolcano Corporation · 2011 · cites it directly24descendants
  • K123898Volcano S5/S5I Intravascular Ultrasound Imaging and Pressure SystemVolcano Corporation · 2013 · 2 generations on21descendants
  • K133323Volcano Ifr ModalityVolcano Corporation · 2014 · 3 generations on18descendants
  • K150441Volcano iFR Modality, iFR Scout featureVolcano Corporation · 2015 · 4 generations on9descendants
  • K170133FFR V2.5Philips Volcano · 2017 · 5 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K934139Tomtec Echo-ViewTomtec Tomographic Technologies GmbH1995—
−5K993398Echo View, Easy View, Omni ViewTomtec Imaging Systems GmbH1999—
−4K022824Echo-View 5.X, Easy-View 2.X, Omni View 2.X, Cardio-View 1.X, LV Analysis 1.X…Tomtec Imaging Systems GmbH2002—
−3K042188Volcano Therapeutics, Inc. Volcano Ivus SystemVolcano Therapeutics, Inc.2004—
−2K051337Eagle Eye Gold Ivus Imaging Catheter, Model 85900/859XX; Volcano VH-Ivus…Volcano Corporation2005—
−1K071554Volcano S5/S5I Intravascular Imaging and Pressure SystemVolcano Corporation20082
0 (this device)K082229S5/S5I Series Intravascular Imaging and Pressure System, Combomap Pressure and…Volcano Corporation2008—
+1K111706Volcano S5IINTRAVASCULAR Imaging and Pressure SystemVolcano Corporation20111
+2K123898Volcano S5/S5I Intravascular Ultrasound Imaging and Pressure SystemVolcano Corporation201310
+3K133323Volcano Ifr ModalityVolcano Corporation2014—
+4K150441Volcano iFR Modality, iFR Scout featureVolcano Corporation2015—
+5K170133FFR V2.5Philips Volcano20171

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
198610
199010
199110
199250
199330
199430
199520
199630
199730
199830
200020
200210
200320
200420
200530
200610
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201210
201340
201410
201530
201730
201810
201940
202010
202110
202210
202310
202410
202510
202610
FIG. 1 — Lineage of K082229, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYO device

  1. #1 K243937 Accuro 3Sscore 20.5 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (7)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K071554Volcano S5/S5I Intravascular Imaging and Pressure System
Summary page 2
…Volcano s5/s~~~~i Series ~~s5/807300-001 Intravascular and ImagingK071554 s5i1807400-001 ComboMapR Pressure and Flow K041134 6800…
Volcano Corporation20082
*K041134Combomap Pressure and Flow System
Summary page 2
…Intravascular and ImagingK071554 s5i1807400-001 ComboMapR Pressure and Flow K041134 6800 System Other prior predicates: WaveMap cleared under K965140; SmartMap cleared under…
Volcano Therapeutics, Inc.2004—
*K965140Cardiometrics Wavewire/Wavemap Pressure System
Summary page 2
…s5i1807400-001 ComboMapR Pressure and Flow K041134 6800 System Other prior predicates: WaveMap cleared under K965140; SmartMap cleared under K021219; ColorFlo Option for the Oracle InVision Intravascular Ultrasound Imaging System cleared under K9…
Cardio Metrics, Inc.1997—
*K021219Smartwire/Smartmap Pressure System
Summary page 2
…ressure and Flow K041134 6800 System Other prior predicates: WaveMap cleared under K965140; SmartMap cleared under K021219; ColorFlo Option for the Oracle InVision Intravascular Ultrasound Imaging System cleared under K963290 on August 6, 1997; the Res…
Jomed, Inc.2002—
*K963290Colorflo Option for the Oracle Invision Intravascular Ultrasonic Imaging System
Summary page 2
…predicates: WaveMap cleared under K965140; SmartMap cleared under K021219; ColorFlo Option for the Oracle InVision Intravascular Ultrasound Imaging System cleared under K963290 on August 6, 1997; the Resolve Option for the Oracle InVision Intravascular Ultrasound Imaging System cleared under K965223 on Ju…
Endosonics Corp.1997—
*K965223Resolve Option for the Oracle in Vision Intravascular Ultrasonic Imaging System
Summary page 2
…cular Ultrasound Imaging System cleared under K963290 on August 6, 1997; the Resolve Option for the Oracle InVision Intravascular Ultrasound Imaging System cleared under K965223 on June 29, 1998, the InVision Imaging System cleared under K031148 on May 28, 2003. NOTE: Volcano Therapeutics, Inc. (now Volca…
Endosonics Corp.1998—
*K031148IN-Vision Imaging System
Summary page 2
…997; the Resolve Option for the Oracle InVision Intravascular Ultrasound Imaging System cleared under K965223 on June 29, 1998, the InVision Imaging System cleared under K031148 on May 28, 2003. NOTE: Volcano Therapeutics, Inc. (now Volcano Corporation) purchased the assets of JOMED Inc. on July 21, 2003.…
Jomed, Inc.2003—

Also cleared against K071554: 1 other device →

What descends from it? (25)

See all 25

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111706Volcano S5IINTRAVASCULAR Imaging and Pressure SystemVolcano Corporation201124

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 23, 2013Class IIZ-2045-2015Nonconforming Material/ComponentA limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano…
Aug 24, 2012Class IIZ-2345-2012Software designThe display on the system has the potential to change. In house testing eventually identified the unique sequence of events that must occur in order for this issue to manifest itself.
Jun 29, 2010Not classifiedZ-0168-2011Software designSoftware defect results in improper image brightness, which may result in introduction of artifacts, which may cause misclassification of results and affect classification of tissue types. This may result in misdiagnosis and inappropriate…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 2.4 years on median.
  • 3 of the 44 ancestors have a recall of any class.
  • Of its descendants, K111706 (Volcano Corporation, 2011) has gone furthest, with 24 descendants of its own.
  • 1 other device also cites its predicate K071554.

How does its category compare?

All IYO devices

IYO System, Imaging, Pulsed Echo, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.

FDA decisions per year · IYO

7in 2025

2007: 202008: 142009: 132010: 152011: 112012: 152013: 172014: 142015: 192016: 102017: 192018: 142019: 132020: 112021: 152022: 52023: 112024: 102025: 72026 (partial year): 10
2007high 20 (2007) · low 5 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYO

171in 2025

2007: 17 days median2008: 36 days median2009: 79 days median2010: 113 days median2011: 69 days median2012: 68 days median2013: 50 days median2014: 63 days median2015: 46 days median2016: 36 days median2017: 72 days median2018: 56 days median2019: 59 days median2020: 57 days median2021: 64 days median2022: 190 days median2023: 177 days median2024: 220 days median2025: 171 days median2026 (partial year): 139 days median
2007high 220 (2024) · low 17 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYO

1in 2022

2007: 42008: 52009: 92010: 82012: 22013: 32014: 22015: 12018: 12019: 12021: 112022: 12026 (partial year): 1
2007high 11 (2021) · low 0 (2011)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYO

485in 2025

2007: 392008: 202009: 272010: 202011: 582012: 782013: 4362014: 3042015: 2682016: 2542017: 2382018: 3682019: 5372020: 4982021: 6172022: 5112023: 5272024: 6402025: 485
2007high 640 (2024) · low 20 (2008)2025

MAUDE reports naming product code IYO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072017439
20081436520
20091379927
201015113820
20111169058
20121568278
201317503436
201414632304
201519461268
201610360254
201719720238
201814561368
201913591537
202011570498
2021156411617
202251901511
2023111770527
2024102200640
202571710485
2026101391—

What is it indicated for?

The Volcano s5/s5i Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures. ChromaFlo® is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow…

Read the full statement (261 words)

The Volcano s5/s5i Series Intravascular Imaging and Pressure System is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures.

ChromaFlo® is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.

VH IVUS is intended to be used in conjunction with imaging catheters during diagnostic ultrasound imaging of the peripheral and coronary vasculature. The Volcano VH IVUS System is intended to semi-automatically visualize boundary features and perform spectral analysis of RF ultrasound signals of vascular features that the user may wish to examine more closely during routine diagnostic ultrasound imaging examinations.

The pressure feature is intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure during diagnostic angiography and/or interventional procedures.

Rotational 45MHz feature is intended for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vasculature. As an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures. The pullback feature of the PIMr withdraws the imaging core within the protective sheath for a maximum of 15 cm.

ComboMapTM Pressure and Flow System is a multi-mode system intended for use in all blood vessels, including coronary and peripheral arteries, to measure intravascular blood pressure and/or blood flow velocities during diagnostic angiography and/or interventional procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K082229: S5/S5I Series Intravascular Imaging and Pressure System, Combomap Pressure and Flow System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K082229

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026