Spinning Spiros, Model CH-2000

K082806510(k) SpecialCleared Apr 6, 2009194-day reviewClass II21 CFR 880.5440Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Icu Medical, Inc. · FPA Set, Administration, Intravascular

Cleared on Apr 6, 2009 as substantially equivalent to K070532 (Icu Medical, Inc., 2007) after a 194-day review. Its family tree reaches back 1 generation to 2007. 1 of the 14 devices in its family, counting itself, has a Class I recall. 12 later devices descend from it. It has been recalled 5 times.

Decided
Apr 6, 2009
Review
194 days
Class
Class II
Product code
FPA
Recalls
5
Predicates
1

Its longest line of descent runs back 1 generation to K070532 (2007). 12 later devices descend from it, 2 citing it directly. 1 of the 14 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2011: 1 descendant cleared2013: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared2017: 1 descendant cleared2020: 1 descendant cleared2018: 1 descendant cleared2021: 2 descendants cleared2023: 1 descendant cleared2025: 2 descendants cleared2024: 1 descendant clearedK070532 · Spiros · Icu Medical, Inc. · 2007 · recalled20052010201520202025
Descendants, by generation 1 generation forward: 2 descendants, 1 with a Class I recall12 generations forward: 1 descendant3 generations forward: 4 descendants4 generations forward: 5 descendants K070532 · Spiros · Icu Medical, Inc. · 2007 · recalled0
  1. Root · 1 generation backK070532 SpirosIcu Medical, Inc. · 2007 · Recalled
  2. This deviceK082806 Spinning Spiros, Model CH-2000Icu Medical, Inc. · 2009
  3. Descendant · cites it directly · 10 of its ownK131549 Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · Recalled (Class I)
  4. Descendant · 2 generations on · 9 of its ownK141448 Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015
  5. Descendant · 3 generations on · 3 of its ownK192866 Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · Recalled
  6. Descendant · 3 generations on · 1 of its ownK201422 Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021
  7. Descendant · 3 generations on · 1 of its ownK180489 TEVADAPTOR Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2018

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2011: 12013: 12015: 12017: 12018: 12020: 12021: 22023: 12024: 12025: 2
2011high 2 (2021) · low 0 (2012)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 12 · See all
  • K131549Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · cites it directly10descendants
  • K141448Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015 · 2 generations on9descendants
  • K192866Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · 3 generations on3descendants
  • K201422Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021 · 3 generations on1descendant
  • K180489TEVADAPTOR Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2018 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K070532SpirosIcu Medical, Inc.20071
0 (this device)K082806Spinning Spiros, Model CH-2000Icu Medical, Inc.2009—
+1K131549Chemolock, Chemolock PortIcu Medical, Inc.2013Class I
+2K141448Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant2015—
+3K192866Chemfort CSTDSimplivia Healthcare, Ltd.20201
+3K201422Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC2021—
+3K180489TEVADAPTOR Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201311
201510
201710
201810
202010
202120
202310
202410
202520
FIG. 1 — Lineage of K082806, 2007–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FPA device

  1. #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K070532Spiros
Summary page 1
…I.V. Administration Set, Accessory 21 CFR 880.5440, Class II Device Product Code: FPA .Legally Marketed Predicate Devices for Substantial Equivalence: *'K070532 - Spires - ICU Medical, Inc. Rationale for SE: The Spinning Spiros device is a passively closed - closed male luer or CML. This…
Icu Medical, Inc.20071

Also cleared against K070532: 3 other devices →

What descends from it? (12)

See all 12

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K131549Chemolock, Chemolock Port Class IIcu Medical, Inc.201310
K112636Cytoguard (TM) Closed Luer ConnectorB.Braun Medical, Inc.20110

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 1, 2022Class IIZ-1723-2022Process controlDue to a manufacturing defect which results in a visible gap between the connector's top and bottom housing.
Mar 16, 2021Class IIZ-1567-2021Component design/selectionDue to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Mar 16, 2021Class IIZ-1568-2021Component design/selectionDue to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Mar 16, 2021Class IIZ-1569-2021Component design/selectionDue to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Aug 31, 2020Class IIZ-0702-2021Process controlPotential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K131549Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K131549 (Icu Medical, Inc., 2013) has gone furthest, with 10 descendants of its own.
  • 3 other devices also cite its predicate K070532.

How does its category compare?

All FPA devices

FPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · FPA

11in 2025

2007: 162008: 182009: 102010: 232011: 122012: 212013: 132014: 222015: 172016: 122017: 172018: 42019: 72020: 102021: 132022: 142023: 142024: 72025: 112026 (partial year): 5
2007high 23 (2010) · low 4 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPA

225in 2025

2007: 75 days median2008: 89 days median2009: 166 days median2010: 120 days median2011: 87 days median2012: 96 days median2013: 90 days median2014: 89 days median2015: 123 days median2016: 195 days median2017: 240 days median2018: 177 days median2019: 161 days median2020: 327 days median2021: 186 days median2022: 213 days median2023: 340 days median2024: 244 days median2025: 225 days median2026 (partial year): 266 days median
2007high 340 (2023) · low 75 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPA

98in 2025

2007: 172008: 152009: 1662010: 582011: 282012: 16, 4 Class I2013: 642014: 242015: 62016: 102017: 132018: 42019: 132020: 1072021: 12022: 9, 5 Class I2023: 12, 4 Class I2024: 13, 1 Class I2025: 98, 1 Class I2026 (partial year): 10, 2 Class I
2007high 166 (2009) · low 1 (2021)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · FPA

34,506in 2025

2007: 1,4992008: 1,9412009: 2,5122010: 2,5372011: 4,4072012: 4,9302013: 4,6512014: 4,6952015: 6,6252016: 8,4702017: 6,4622018: 6,5742019: 11,2272020: 9,5632021: 10,4142022: 13,3932023: 10,9162024: 47,2102025: 34,506
2007high 47,210 (2024) · low 1,499 (2007)2025

MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071675171,499
20081889151,941
2009101661662,512
201023120582,537
20111287284,407
2012219616 (4)4,930
20131390644,651
20142289244,695
20151712366,625
201612195108,470
201717240136,462
2018417746,574
201971611311,227
2020103271079,563
202113186110,414
2022142139 (5)13,393
20231434012 (4)10,916
2024724413 (1)47,210
20251122598 (1)34,506
2026526610 (2)—

What is it indicated for?

The Spiros is a single use, sterile, non-pyrogenic, swab-able, bidirectional valve device intended for use as an accessory to an Intravascular Administration Set. The Spiros provides access for the administration needle or catheter (which is inserted into the vein or artery).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K082806: Spinning Spiros, Model CH-2000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K082806

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026