510(k) · Cleared
Recalled · Class II · 2022Spinning Spiros, Model CH-2000
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Icu Medical, Inc. · FPA Set, Administration, Intravascular
Cleared on Apr 6, 2009 as substantially equivalent to K070532 (Icu Medical, Inc., 2007) after a 194-day review. Its family tree reaches back 1 generation to 2007. 1 of the 14 devices in its family, counting itself, has a Class I recall. 12 later devices descend from it. It has been recalled 5 times.
- Decided
- Apr 6, 2009
- Review
- 194 days
- Class
- Class II
- Product code
- FPA
- Recalls
- 5
- Predicates
- 1
Its longest line of descent runs back 1 generation to K070532 (2007). 12 later devices descend from it, 2 citing it directly. 1 of the 14 devices in this family, counting this one, has a Class I recall.
- Root · 1 generation backK070532 SpirosIcu Medical, Inc. · 2007 · Recalled
- This deviceK082806 Spinning Spiros, Model CH-2000Icu Medical, Inc. · 2009
- Descendant · cites it directly · 10 of its ownK131549 Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · Recalled (Class I)
- Descendant · 2 generations on · 9 of its ownK141448 Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015
- Descendant · 3 generations on · 3 of its ownK192866 Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · Recalled
- Descendant · 3 generations on · 1 of its ownK201422 Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021
- Descendant · 3 generations on · 1 of its ownK180489 TEVADAPTOR Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2018
Showing 5 of 12 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K131549 Chemolock, Chemolock Portcites it directly · 2013
Most influential descendants
Showing 5 of 12 · See all- K131549Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · cites it directly10descendants
- K141448Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015 · 2 generations on9descendants
- K192866Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · 3 generations on3descendants
- K201422Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021 · 3 generations on1descendant
- K180489TEVADAPTOR Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2018 · 3 generations on1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K070532 | Spiros | Icu Medical, Inc. | 2007 | 1 |
| 0 (this device) | K082806 | Spinning Spiros, Model CH-2000 | Icu Medical, Inc. | 2009 | — |
| +1 | K131549 | Chemolock, Chemolock Port | Icu Medical, Inc. | 2013 | Class I |
| +2 | K141448 | Tevadaptor Closed Drug Reconstitution and Transfer System | Teva Medical Ltd., Migada Plant | 2015 | — |
| +3 | K192866 | Chemfort CSTD | Simplivia Healthcare, Ltd. | 2020 | 1 |
| +3 | K201422 | Arisure Closed System Drug Transfer Device (CSTD) | Yukon Medical, LLC | 2021 | — |
| +3 | K180489 | TEVADAPTOR Closed Drug Reconstitution and Transfer System | Teva Medical Ltd., Migada Plant | 2018 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2011 | 1 | 0 |
| 2013 | 1 | 1 |
| 2015 | 1 | 0 |
| 2017 | 1 | 0 |
| 2018 | 1 | 0 |
| 2020 | 1 | 0 |
| 2021 | 2 | 0 |
| 2023 | 1 | 0 |
| 2024 | 1 | 0 |
| 2025 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FPA device
- #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K070532 | SpirosSummary page 1…I.V. Administration Set, Accessory 21 CFR 880.5440, Class II Device Product Code: FPA .Legally Marketed Predicate Devices for Substantial Equivalence: *'K070532 - Spires - ICU Medical, Inc. Rationale for SE: The Spinning Spiros device is a passively closed - closed male luer or CML. This… | 2007 |
What descends from it? (12)
See all 122 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 1, 2022 | Class II | Z-1723-2022 | Process control | Due to a manufacturing defect which results in a visible gap between the connector's top and bottom housing. |
| Mar 16, 2021 | Class II | Z-1567-2021 | Component design/selection | Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position. |
| Mar 16, 2021 | Class II | Z-1568-2021 | Component design/selection | Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position. |
| Mar 16, 2021 | Class II | Z-1569-2021 | Component design/selection | Due to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position. |
| Aug 31, 2020 | Class II | Z-0702-2021 | Process control | Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to… |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K131549Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · Cites it directly · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls
What does this lineage say?
How does its category compare?
All FPA devicesFPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.
FDA decisions per year · FPA
11in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FPA
225in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FPA
98in 2025
Class I (17 since 2007) Class II/III not classified
Adverse event reports per year · FPA
34,506in 2025
MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 16 | 75 | 17 | 1,499 |
| 2008 | 18 | 89 | 15 | 1,941 |
| 2009 | 10 | 166 | 166 | 2,512 |
| 2010 | 23 | 120 | 58 | 2,537 |
| 2011 | 12 | 87 | 28 | 4,407 |
| 2012 | 21 | 96 | 16 (4) | 4,930 |
| 2013 | 13 | 90 | 64 | 4,651 |
| 2014 | 22 | 89 | 24 | 4,695 |
| 2015 | 17 | 123 | 6 | 6,625 |
| 2016 | 12 | 195 | 10 | 8,470 |
| 2017 | 17 | 240 | 13 | 6,462 |
| 2018 | 4 | 177 | 4 | 6,574 |
| 2019 | 7 | 161 | 13 | 11,227 |
| 2020 | 10 | 327 | 107 | 9,563 |
| 2021 | 13 | 186 | 1 | 10,414 |
| 2022 | 14 | 213 | 9 (5) | 13,393 |
| 2023 | 14 | 340 | 12 (4) | 10,916 |
| 2024 | 7 | 244 | 13 (1) | 47,210 |
| 2025 | 11 | 225 | 98 (1) | 34,506 |
| 2026 | 5 | 266 | 10 (2) | — |
What is it indicated for?
The Spiros is a single use, sterile, non-pyrogenic, swab-able, bidirectional valve device intended for use as an accessory to an Intravascular Administration Set. The Spiros provides access for the administration needle or catheter (which is inserted into the vein or artery).
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K082806 · printed Nov 4, 2026
Cite this page
510(k) Report. “K082806: Spinning Spiros, Model CH-2000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K082806
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026