510(k) · Cleared
Recalled · 2011Apex HP M PH.I.S.I.O. Adult Hollow Fiber Membrane Oxygenator
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Sorin Group Italia S.R.L. · DTZ Oxygenator, Cardiopulmonary Bypass
Cleared on Oct 29, 2008 as substantially equivalent to K020997 (Dideco S.P.A., 2002) and 1 other predicate after a 20-day review. Its family tree reaches back 4 generations to 1992. 4 later devices descend from it. It has been recalled 2 times.
- Decided
- Oct 29, 2008
- Review
- 20 days
- Class
- Class II
- Product code
- DTZ
- Recalls
- 2
- Predicates
- 2
Its longest line of descent runs back 4 generations to K922933 (1992). 7 ancestors in all. 4 later devices descend from it, 1 citing it directly.
- Root · 4 generations backK922933 Sorin Monolyrh Integrated Membrane LungSorin Biomedica, Fiat, USA, Inc. · 1992
- 3 generations backK980600 D903 Avant Adult Hollow Fiber OxygenatorDideco S.P.A. · 1998
- 2 generations backK020351 D 903 Avant 2 PH.I.S.I.O. Adult Hollow Fiber OxygenatorDideco S.P.A. · 2002
- 1 generation backK020997 Apex PH.I.S.I.O. Adult Hollow Fiber Membrane OxygenatorDideco S.P.A. · 2002 · Recalled
- This deviceK083021 Apex HP M PH.I.S.I.O. Adult Hollow Fiber Membrane OxygenatorSorin Group Italia S.R.L. · 2008
- Descendant · cites it directly · 3 of its ownK092895 Apex HP M Adult Hollow Fiber Membrane OxygenatorSorin Group Italia S.R.L. · 2009 · Recalled
- Descendant · 2 generations on · 2 of its ownK212688 Quantum PureFlow Standard Heat Exchanger, Quantum PureFlow…Qura S.R.L · 2022
- Descendant · 3 generations on · 1 of its ownK220110 Quantum PureFlow Standard Heat Exchanger and Quantum PureFlow…Qura S.R.L · 2022 · Recalled
- Descendant · 4 generations on · 0 of its ownK223879 Quantum PureFlow Standard Heat Exchangers (HX11W-S2, HX11W-S1M and…Quara S.R.L. · 2023 · Recalled
Most influential descendants
Showing 4 of 4 · See all- K092895Apex HP M Adult Hollow Fiber Membrane OxygenatorSorin Group Italia S.R.L. · 2009 · cites it directly3descendants
- K212688Quantum PureFlow Standard Heat Exchanger, Quantum PureFlow Cardioplegia Heat ExchangerQura S.R.L · 2022 · 2 generations on2descendants
- K220110Quantum PureFlow Standard Heat Exchanger and Quantum PureFlow Cardioplegia Heat ExchangerQura S.R.L · 2022 · 3 generations on1descendant
- K223879Quantum PureFlow Standard Heat Exchangers (HX11W-S2, HX11W-S1M and HX11W-S2M)Quara S.R.L. · 2023 · 4 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K922933 | Sorin Monolyrh Integrated Membrane Lung | Sorin Biomedica, Fiat, USA, Inc. | 1992 | — |
| −3 | K980600 | D903 Avant Adult Hollow Fiber Oxygenator | Dideco S.P.A. | 1998 | — |
| −2 | K020351 | D 903 Avant 2 PH.I.S.I.O. Adult Hollow Fiber Oxygenator | Dideco S.P.A. | 2002 | — |
| −1 | K020997 | Apex PH.I.S.I.O. Adult Hollow Fiber Membrane Oxygenator | Dideco S.P.A. | 2002 | 2 |
| 0 (this device) | K083021 | Apex HP M PH.I.S.I.O. Adult Hollow Fiber Membrane Oxygenator | Sorin Group Italia S.R.L. | 2008 | — |
| +1 | K092895 | Apex HP M Adult Hollow Fiber Membrane Oxygenator | Sorin Group Italia S.R.L. | 2009 | 1 |
| +2 | K212688 | Quantum PureFlow Standard Heat Exchanger, Quantum PureFlow Cardioplegia Heat… | Qura S.R.L | 2022 | — |
| +3 | K220110 | Quantum PureFlow Standard Heat Exchanger and Quantum PureFlow Cardioplegia… | Qura S.R.L | 2022 | 2 |
| +4 | K223879 | Quantum PureFlow Standard Heat Exchangers (HX11W-S2, HX11W-S1M and HX11W-S2M) | Quara S.R.L. | 2023 | 2 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2001 | 2 | 0 |
| 2002 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2009 | 1 | 0 |
| 2022 | 2 | 0 |
| 2023 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DTZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DTZ device
- #1 K240934 Quantum Perfusion Blood Oxygenator ECC…score 14.3 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K020997 | Apex PH.I.S.I.O. Adult Hollow Fiber Membrane OxygenatorSummary page 1…onary Bypass Oxygenator LEGALLY UNMODIFIED DEVICE: Apex M Ph.I.S.I.O. Adult Hollow Fiber Membrane Oxygenator (K020997) PREDICATE DEVICE: Dideco D 903 Avant 2 Ph.I.S.1.0 Adult Hollow… | 2002 |
| * | K020351 | D 903 Avant 2 PH.I.S.I.O. Adult Hollow Fiber OxygenatorSummary page 1…xygenator (K020997) PREDICATE DEVICE: Dideco D 903 Avant 2 Ph.I.S.1.0 Adult Hollow Fiber Oxygenator (K020351) DEVICE DESCRIPTION: The Apex HP M Ph.I.S.1.0 Adult Hollow Fiber Membrane Oxygenator is a cardiopulmonary bypass blood oxygenat… | 2002 |
What descends from it? (4)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K092895 | Apex HP M Adult Hollow Fiber Membrane Oxygenator | 2009 | 3 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Dec 2, 2011 | Not classified | Z-0764-2012 | Device Design | Customer complaints of a leak between the blood and water compartments of the heat exchanger used in the oxygenators due to prolonged contact of the heat exchangers with sodium chloride causing micro-holes. |
| Dec 2, 2011 | Not classified | Z-0765-2012 | Device Design | Customer complaints of a leak between the blood and water compartments of the heat exchanger used in the oxygenators due to prolonged contact of the heat exchangers with sodium chloride causing micro-holes. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 5.7 years on median.
- 2 of the 7 ancestors have a recall of any class.
- Of its descendants, K092895 (Sorin Group Italia S.R.L., 2009) has gone furthest, with 3 descendants of its own.
How does its category compare?
All DTZ devicesDTZ Oxygenator, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2008, this device’s year, is highlighted.
FDA decisions per year · DTZ
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DTZ
124in 2024
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DTZ
1in 2025
Class I (3 since 2007) Class II/III not classified
Adverse event reports per year · DTZ
976in 2025
MAUDE reports naming product code DTZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 4 | 55 | 7 | 134 |
| 2008 | 3 | 88 | 5 | 105 |
| 2009 | 9 | 210 | 0 | 52 |
| 2010 | 2 | — | 1 | 119 |
| 2011 | 4 | 160 | 13 | 190 |
| 2012 | 9 | 90 | 14 | 215 |
| 2013 | 9 | 28 | 0 | 346 |
| 2014 | 5 | 28 | 2 | 764 |
| 2015 | 9 | 134 | 0 | 658 |
| 2016 | 3 | 62 | 1 | 443 |
| 2017 | 2 | — | 0 | 388 |
| 2018 | 3 | 72 | 2 (2) | 320 |
| 2019 | 5 | 90 | 1 | 489 |
| 2020 | 5 | 93 | 3 | 442 |
| 2021 | 5 | 336 | 1 (1) | 595 |
| 2022 | 2 | — | 6 | 559 |
| 2023 | 5 | 145 | 0 | 738 |
| 2024 | 3 | 124 | 0 | 535 |
| 2025 | 2 | — | 1 | 976 |
| 2026 | 3 | 137 | 0 | — |
What is it indicated for?
The Apex HP MPh.I.S.I.O. Adult Hollow Fiber Membrane Oxygenator is intended to be used in adult surgical procedures requiring extracorporeal gas exchange support and blood temperature control for periods of up to 6 hours.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K083021 · printed Nov 6, 2026
Cite this page
510(k) Report. “K083021: Apex HP M PH.I.S.I.O. Adult Hollow Fiber Membrane Oxygenator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K083021
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026