510(k) · Cleared
Recalled · Class II · 2026Quantum PureFlow Standard Heat Exchanger and Quantum PureFlow Cardioplegia Heat Exchanger
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Qura S.R.L · DTR Heat-Exchanger, Cardiopulmonary Bypass
Cleared on May 6, 2022 as substantially equivalent to K212688 (Qura S.R.L, 2022) after a 113-day review. Its family tree reaches back 7 generations to 1992. 1 later device descends from it. It has been recalled 2 times.
- Decided
- May 6, 2022
- Review
- 113 days
- Class
- Class II
- Product code
- DTR
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 7 generations to K922933 (1992). 12 ancestors in all. 1 later device descends from it, 1 citing it directly.
- Root · 7 generations backK922933 Sorin Monolyrh Integrated Membrane LungSorin Biomedica, Fiat, USA, Inc. · 1992
- 6 generations backK980600 D903 Avant Adult Hollow Fiber OxygenatorDideco S.P.A. · 1998
- 5 generations backK020351 D 903 Avant 2 PH.I.S.I.O. Adult Hollow Fiber OxygenatorDideco S.P.A. · 2002
- 4 generations backK020997 Apex PH.I.S.I.O. Adult Hollow Fiber Membrane OxygenatorDideco S.P.A. · 2002 · Recalled
- 3 generations backK083021 Apex HP M PH.I.S.I.O. Adult Hollow Fiber Membrane OxygenatorSorin Group Italia S.R.L. · 2008 · Recalled
- 2 generations backK092895 Apex HP M Adult Hollow Fiber Membrane OxygenatorSorin Group Italia S.R.L. · 2009 · Recalled
- 1 generation backK212688 Quantum PureFlow Standard Heat Exchanger, Quantum PureFlow…Qura S.R.L · 2022
- This deviceK220110 Quantum PureFlow Standard Heat Exchanger and Quantum PureFlow…Qura S.R.L · 2022
- Descendant · cites it directly · 0 of its ownK223879 Quantum PureFlow Standard Heat Exchangers (HX11W-S2, HX11W-S1M and…Quara S.R.L. · 2023 · Recalled
Most influential descendants
Showing 1 of 1 · See all- K223879Quantum PureFlow Standard Heat Exchangers (HX11W-S2, HX11W-S1M and HX11W-S2M)Quara S.R.L. · 2023 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −7 | K922933 | Sorin Monolyrh Integrated Membrane Lung | Sorin Biomedica, Fiat, USA, Inc. | 1992 | — |
| −6 | K980600 | D903 Avant Adult Hollow Fiber Oxygenator | Dideco S.P.A. | 1998 | — |
| −5 | K020351 | D 903 Avant 2 PH.I.S.I.O. Adult Hollow Fiber Oxygenator | Dideco S.P.A. | 2002 | — |
| −4 | K020997 | Apex PH.I.S.I.O. Adult Hollow Fiber Membrane Oxygenator | Dideco S.P.A. | 2002 | 2 |
| −3 | K083021 | Apex HP M PH.I.S.I.O. Adult Hollow Fiber Membrane Oxygenator | Sorin Group Italia S.R.L. | 2008 | 2 |
| −2 | K092895 | Apex HP M Adult Hollow Fiber Membrane Oxygenator | Sorin Group Italia S.R.L. | 2009 | 1 |
| −1 | K212688 | Quantum PureFlow Standard Heat Exchanger, Quantum PureFlow Cardioplegia Heat… | Qura S.R.L | 2022 | — |
| 0 (this device) | K220110 | Quantum PureFlow Standard Heat Exchanger and Quantum PureFlow Cardioplegia… | Qura S.R.L | 2022 | — |
| +1 | K223879 | Quantum PureFlow Standard Heat Exchangers (HX11W-S2, HX11W-S1M and HX11W-S2M) | Quara S.R.L. | 2023 | 2 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1994 | 1 | 0 |
| 2001 | 3 | 0 |
| 2002 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2023 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DTR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DTR device
- #1 K252541 Quantum PureFlow Standard Heat Exchanger…score 13.4 · 9 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K212688 | Quantum PureFlow Standard Heat Exchanger, Quantum PureFlow Cardioplegia Heat ExchangerSummary page 5…T2) / Division of Health Technology 2 B (Circulatory Support, Structural and Vascular Devices) 510(k) Number: K212688 Page 1… | 2022 |
What descends from it? (1)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K223879 | Quantum PureFlow Standard Heat Exchangers (HX11W-S2, HX11W-S1M and HX11W-S2M) | 2023 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jul 17, 2026 | Class II | Z-3327-2026 | Process design | Due to blood-to-glycol leak in the heat exchanger. |
| Jul 17, 2026 | Class II | Z-3328-2026 | Process design | Due to blood-to-glycol leak in the heat exchanger. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 30 years and passes through 4 companies; a new generation was cleared every 3.5 years on median.
- 5 of the 12 ancestors have a recall of any class.
- Of its descendants, K223879 (Quara S.R.L., 2023) has gone furthest, with 0 descendants of its own.
How does its category compare?
All DTR devicesDTR Heat-Exchanger, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2022, this device’s year, is highlighted.
FDA decisions per year · DTR
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DTR
Too few DTR 510(k)s in any year since 2007 to give a median.
Recalls per year · DTR
1in 2015
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DTR
44in 2025
MAUDE reports naming product code DTR, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 2 |
| 2008 | 1 | — | 0 | 1 |
| 2009 | 2 | — | 0 | 0 |
| 2010 | 0 | — | 24 | 43 |
| 2011 | 0 | — | 3 | 67 |
| 2012 | 0 | — | 0 | 16 |
| 2013 | 0 | — | 0 | 27 |
| 2014 | 0 | — | 0 | 66 |
| 2015 | 0 | — | 1 | 96 |
| 2016 | 0 | — | 0 | 60 |
| 2017 | 2 | — | 0 | 69 |
| 2018 | 1 | — | 0 | 75 |
| 2019 | 0 | — | 0 | 35 |
| 2020 | 2 | — | 0 | 33 |
| 2021 | 0 | — | 0 | 11 |
| 2022 | 2 | — | 0 | 23 |
| 2023 | 1 | — | 0 | 18 |
| 2024 | 1 | — | 0 | 8 |
| 2025 | 2 | — | 0 | 44 |
| 2026 | 1 | — | 2 | — |
What is it indicated for?
The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration.
Devices are intended for adult patients.
The Quantum PureFlow Cardioplegia Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool cardioplegic solution, to remove air and to monitor the temperature during routine cardiopulmonary bypass (CPB) procedures up to 6 hours duration. Any blood to be treated must contain anticoagulant. Contact with cardioplegic solution for longer than this period is not permitted.
Devices are intended for adult patients.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K220110 · printed Nov 6, 2026
Cite this page
510(k) Report. “K220110: Quantum PureFlow Standard Heat Exchanger and Quantum PureFlow Cardioplegia Heat Exchanger.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K220110
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026