510(k) Report

510(k) · Cleared

Cleared

PK-141 Patient Cable

K083674510(k) TraditionalCleared Apr 1, 2009111-day reviewClass II21 CFR 870.3630

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biotronik, Inc. · DTC Analyzer, Pacemaker Generator Function

Cleared on Apr 1, 2009 as substantially equivalent to K022360 (Biotronik, Inc., 2003) after a 111-day review. Its family tree reaches back 3 generations to 1988. 1 later device descends from it.

Decided
Apr 1, 2009
Review
111 days
Class
Class II
Product code
DTC
Recalls
None
Predicates
1

Its longest line of descent runs back 3 generations to K883930 (1988). 4 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 3 generations backK883930 Seamed Model 3300 Dual Chamber Pacing System AnalySeamed Corp. · 1988
  2. 2 generations backK964190 Era 300 Dual Chamber Pacing System AnalyzerBiotronik, Inc. · 1997
  3. 1 generation backK022360 Era 3000; PA-2 IS-1 Adapter; EK-4-N Pacemaker Test Cable; Era 3000…Biotronik, Inc. · 2003
  4. This deviceK083674 PK-141 Patient CableBiotronik, Inc. · 2009
  5. Descendant · cites it directly · 0 of its ownK170815 PSA CableBoston Scientific Corporation · 2017

Most influential descendants

Showing 1 of 1 · See all
  • K170815PSA CableBoston Scientific Corporation · 2017 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K883930Seamed Model 3300 Dual Chamber Pacing System AnalySeamed Corp.1988—
−2K964190Era 300 Dual Chamber Pacing System AnalyzerBiotronik, Inc.1997—
−1K022360Era 3000; PA-2 IS-1 Adapter; EK-4-N Pacemaker Test Cable; Era 3000 Charger…Biotronik, Inc.2003—
0 (this device)K083674PK-141 Patient CableBiotronik, Inc.2009—
+1K170815PSA CableBoston Scientific Corporation2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
FIG. 1 — Lineage of K083674, 1988–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DTC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022360Era 3000; PA-2 IS-1 Adapter; EK-4-N Pacemaker Test Cable; Era 3000 Charger; Era 3000…
Summary page 1
…ame: PK-141 Common Name: Programmer, Pacemaker (Accessories) Classification: Class Il (21 CFR 870.3700) Classification Name: Cabling Product Code: KRG §10(k) Number: K022360 Date Prepared: December 10, 2008 General Description: The PK-141 patient cable is compatible with the following BIOTRONIK extern…
Biotronik, Inc.2003—

Also cleared against K022360: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K170815PSA CableBoston Scientific Corporation20170

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 2 companies; a new generation was cleared every 6.2 years on median.
  • None of the 4 ancestors has a recall.
  • Of its descendants, K170815 (Boston Scientific Corporation, 2017) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K022360.

How does its category compare?

All DTC devices

DTC Analyzer, Pacemaker Generator Function: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · DTC

1in 2017

2009: 12017: 1
2007high 1 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTC

Too few DTC 510(k)s in any year since 2007 to give a median.

Recalls per year · DTC

No recall names DTC since 2007.

Adverse event reports per year · DTC

1,159in 2025

2010: 22011: 12014: 12017: 342018: 222019: 1382020: 1892021: 6362022: 2182023: 2152024: 1,0762025: 1,159
2007high 1,159 (2025) · low 0 (2007)2025

MAUDE reports naming product code DTC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20091—00
20100—02
20110—01
20120—00
20130—00
20140—01
20150—00
20160—00
20171—034
20180—022
20190—0138
20200—0189
20210—0636
20220—0218
20230—0215
20240—01,076
20250—01,159
20260—0—

What is it indicated for?

The PK-141 patient cable is an accessory indicated for use with external equipment from BIOTRONIK and is used for the transmission of sensing signals and pacing pulses for diagnostics and therapy in the context of intracardiac examinations including the following activities: Temporary External Pacing Provides temporary stimulation during implantable pacemaker procedures or physician evaluations. Lead Threshold Determination Determines in situ lead…

Read the full statement (108 words)

The PK-141 patient cable is an accessory indicated for use with external equipment from BIOTRONIK and is used for the transmission of sensing signals and pacing pulses for diagnostics and therapy in the context of intracardiac examinations including the following activities:

Temporary External Pacing

Provides temporary stimulation during implantable pacemaker procedures or physician evaluations.

Lead Threshold Determination

Determines in situ lead characteristics of impedance, capture threshold, P/R wave amplitude and P/R wave slew rate. Determines the in vivo retrograde conduction time.

Pacemaker Function Test

Tests and analyzes the.in vitro operation of external or implantable pulse generators. Determines the following parameters: pulse amplitude arid width, A/V delay, and rate/interval.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K083674: PK-141 Patient Cable.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K083674

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026