510(k) · Cleared
ClearedEra 3000; PA-2 IS-1 Adapter; EK-4-N Pacemaker Test Cable; Era 3000 Charger; Era 3000 Battery (Type 2); NK-11 Power Supp.
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Biotronik, Inc. · DTA Tester, Pacemaker Electrode Function
Cleared on Jan 27, 2003 as substantially equivalent to K964190 (Biotronik, Inc., 1997) after a 192-day review. Its family tree reaches back 2 generations to 1988. 4 later devices descend from it.
- Decided
- Jan 27, 2003
- Review
- 192 days
- Class
- Class II
- Product code
- DTA
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 2 generations to K883930 (1988). 3 ancestors in all. 4 later devices descend from it, 3 citing it directly.
- Root · 2 generations backK883930 Seamed Model 3300 Dual Chamber Pacing System AnalySeamed Corp. · 1988
- 1 generation backK964190 Era 300 Dual Chamber Pacing System AnalyzerBiotronik, Inc. · 1997
- This deviceK022360 Era 3000; PA-2 IS-1 Adapter; EK-4-N Pacemaker Test Cable; Era 3000…Biotronik, Inc. · 2003
- Descendant · cites it directly · 1 of its ownK083674 PK-141 Patient CableBiotronik, Inc. · 2009
- Descendant · 2 generations on · 0 of its ownK170815 PSA CableBoston Scientific Corporation · 2017
- Descendant · cites it directly · 0 of its ownK101982 Merlin Psa Patient Cable EX 3150, Cable Adapter EX3170, M Adapter…St Jude Medical, Crmd · 2010
- Descendant · cites it directly · 0 of its ownK042708 Era 3000 Pacing System AnalyzerBiotronik, Inc. · 2004
Most influential descendants
Showing 4 of 4 · See all- K083674PK-141 Patient CableBiotronik, Inc. · 2009 · cites it directly1descendant
- K170815PSA CableBoston Scientific Corporation · 2017 · 2 generations on0descendants
- K101982Merlin Psa Patient Cable EX 3150, Cable Adapter EX3170, M Adapter EX3180St Jude Medical, Crmd · 2010 · cites it directly0descendants
- K042708Era 3000 Pacing System AnalyzerBiotronik, Inc. · 2004 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K883930 | Seamed Model 3300 Dual Chamber Pacing System Analy | Seamed Corp. | 1988 | — |
| −1 | K964190 | Era 300 Dual Chamber Pacing System Analyzer | Biotronik, Inc. | 1997 | — |
| 0 (this device) | K022360 | Era 3000; PA-2 IS-1 Adapter; EK-4-N Pacemaker Test Cable; Era 3000 Charger… | Biotronik, Inc. | 2003 | — |
| +1 | K083674 | PK-141 Patient Cable | Biotronik, Inc. | 2009 | — |
| +2 | K170815 | PSA Cable | Boston Scientific Corporation | 2017 | — |
| +1 | K101982 | Merlin Psa Patient Cable EX 3150, Cable Adapter EX3170, M Adapter EX3180 | St Jude Medical, Crmd | 2010 | — |
| +1 | K042708 | Era 3000 Pacing System Analyzer | Biotronik, Inc. | 2004 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1989 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2004 | 1 | 0 |
| 2009 | 1 | 0 |
| 2010 | 1 | 0 |
| 2017 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DTA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K964190 | Era 300 Dual Chamber Pacing System AnalyzerSummary page 1…stem Analyzer cleared through 5101kl notification as a predicate device for the ERA 3000 Pacing System Analyzer: BIOTRONIKs ERA 300 Pacing System Analyzer l#K964190. cleared 07-10-971 Indications for Use: The ERA 3000 is intended for use during invasive pacemaker procedures in the following ac… | 1997 |
What descends from it? (4)
3 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 15 years and passes through 2 companies; a new generation was cleared every 8.7 years on median.
- None of the 3 ancestors has a recall.
- Of its descendants, K083674 (Biotronik, Inc., 2009) has gone furthest, with 1 descendant of its own.
- 4 other devices also cite its predicate K964190.
How does its category compare?
All DTA devicesDTA Tester, Pacemaker Electrode Function: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DTA
1in 2017
2026 to Oct 25, outlined.
Median 510(k) review, days · DTA
Too few DTA 510(k)s in any year since 2007 to give a median.
Recalls per year · DTA
No recall names DTA since 2007.
Adverse event reports per year · DTA
1in 2023
MAUDE reports naming product code DTA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 1 | — | 0 | 1 |
| 2011 | 0 | — | 0 | 0 |
| 2012 | 0 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 2 |
| 2014 | 0 | — | 0 | 0 |
| 2015 | 0 | — | 0 | 1 |
| 2016 | 0 | — | 0 | 0 |
| 2017 | 1 | — | 0 | 0 |
| 2018 | 0 | — | 0 | 0 |
| 2019 | 0 | — | 0 | 0 |
| 2020 | 0 | — | 0 | 0 |
| 2021 | 0 | — | 0 | 0 |
| 2022 | 0 | — | 0 | 0 |
| 2023 | 0 | — | 0 | 1 |
| 2024 | 0 | — | 0 | 0 |
| 2025 | 0 | — | 0 | 0 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The ERA 3000 is intended for use during invasive pacemaker procedures in the following activities:
Temporary External Pacing
Provides temporary stimulation under DDD, DDI, DOO, WI, VDD, VOO, MI, AOO, or OD0 modalities during implantable pacemaker procedures or physician evaluations.
Lead Threshold Determination
Determines in situ lead characteristics of impedance, capture threshold, P/R wave amplitude and P/R wave slew rate. Determines the in vivo retrograde conduction time.
Pacemaker Function Test
Tests and analyzes the in vitro operation of external or implantable pulse generators. Determines the following parameters: pulse amplitude and width, AN delay, and rate/interval.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K022360 · printed Nov 6, 2026
Cite this page
510(k) Report. “K022360: Era 3000; PA-2 IS-1 Adapter; EK-4-N Pacemaker Test Cable; Era 3000 Charger; Era 3000 Battery (Type 2); NK-11 Power Supp..” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K022360
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026