510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

Carto 3 EP Navigation System, V1.0 and Accessories

K090017510(k) TraditionalCleared Oct 13, 2009284-day reviewClass II21 CFR 870.1425Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biosense Webster, Inc. · DQK Computer, Diagnostic, Programmable

Cleared on Oct 13, 2009 as substantially equivalent to K072202 (Biosense Webster, Inc., 2007) and 2 other predicates after a 284-day review. Its family tree reaches back 6 generations to 2000. 45 later devices descend from it. It has been recalled 4 times.

Decided
Oct 13, 2009
Review
284 days
Class
Class II
Product code
DQK
Recalls
4
Predicates
3

Its longest line of descent runs back 6 generations to K001437 (2000). 14 ancestors in all, the oldest K982415 (1998). 45 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2006: 2 other ancestors2005: 1 other ancestor2001: 1 other ancestor1998: 1 other ancestor2003: 1 other ancestor2000: 1 other ancestor2010: 1 descendant cleared2009: 1 descendant cleared2011: 2 descendants cleared2020: 2 descendants cleared2014: 2 descendants cleared2017: 2 descendants cleared2022: 3 descendants cleared2023: 5 descendants cleared2025: 7 descendants cleared2021: 2 descendants cleared2012: 1 descendant cleared2018: 5 descendants cleared2024: 4 descendants cleared2016: 1 descendant cleared2019: 2 descendants cleared2026: 5 descendants clearedK001437 · Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2000K012926 · Modification to Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2001K030129 · Modification to Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2003K042493 · Ensite System, Model EE3000 · Endocardial Solutions, Inc. · 2004K060954 · Modification to Ensite System, Model EE3000 · St. Jude Medical, Inc.-Endocardial Solutions · 2006K072202 · Carto 3 Navigation System (Model:FG-5400-00), Carto 3 External Reference Patches (Model:D-1283-01) · Biosense Webster, Inc. · 20071995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 4 other ancestors3 generations back: 2 other ancestors 1 generation forward: 2 descendants2 generations forward: 3 descendants3 generations forward: 7 descendants4 generations forward: 5 descendants5 generations forward: 4 descendants6 generations forward: 5 descendants7 generations forward: 2 descendants8 generations forward: 1 descendant9 generations forward: 2 descendants10 generations forward: 3 descendants11 generations forward: 2 descendants12 generations forward: 5 descendants13 generations forward: 4 descendants K072202 · Carto 3 Navigation System (Model:FG-5400-00), Carto 3 External Reference Patches (Model:D-1283-01) · Biosense Webster, Inc. · 2007K060954 · Modification to Ensite System, Model EE3000 · St. Jude Medical, Inc.-Endocardial Solutions · 2006K042493 · Ensite System, Model EE3000 · Endocardial Solutions, Inc. · 2004K030129 · Modification to Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2003K012926 · Modification to Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2001K001437 · Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2000−50+5+10
  1. Root · 6 generations backK001437 Ensite 3000 System, Model EE3000Endocardial Solutions, Inc. · 2000
  2. 5 generations backK012926 Modification to Ensite 3000 System, Model EE3000Endocardial Solutions, Inc. · 2001
  3. 4 generations backK030129 Modification to Ensite 3000 System, Model EE3000Endocardial Solutions, Inc. · 2003
  4. 3 generations backK042493 Ensite System, Model EE3000Endocardial Solutions, Inc. · 2004
  5. 2 generations backK060954 Modification to Ensite System, Model EE3000St. Jude Medical, Inc.-Endocardial Solutions · 2006
  6. 1 generation backK072202 Carto 3 Navigation System (Model:FG-5400-00), Carto 3 External…Biosense Webster, Inc. · 2007
  7. This deviceK090017 Carto 3 EP Navigation System, V1.0 and AccessoriesBiosense Webster, Inc. · 2009
  8. Descendant · cites it directly · 34 of its ownK093455 Carto 3 Version 1.05 EP Navigation System and AccessoriesBiosense Webster, Inc. · 2009 · Recalled
  9. Descendant · 2 generations on · 33 of its ownK103746 Carto 3 EP Navigation System, Version 2.0 (W/Standard Location…Biosense Webster, Inc. · 2011 · Recalled
  10. Descendant · 3 generations on · 32 of its ownK112007 Carto 3 EP Navigation System, Version 2.2Biosense Webster, Inc. · 2011 · Recalled
  11. Descendant · 4 generations on · 31 of its ownK120550 Carto (R) EP Navigation System, Version 3.0 (with Standard Location…Biosense Webster, Inc. · 2012 · Recalled
  12. Descendant · 5 generations on · 27 of its ownK133916 Carto 3 EP Navigation System, Version 4.2Biosense Webster, Inc. · 2014 · Recalled

Showing 5 of 45 descendants. See all descendants

Descendants cleared per year

2009: 12010: 12011: 22012: 12014: 22016: 12017: 22018: 52019: 22020: 22021: 22022: 32023: 52024: 42025: 72026: 5
2009high 7 (2025) · low 0 (2013)2026

Most influential descendants

Showing 5 of 45 · See all
  • K093455Carto 3 Version 1.05 EP Navigation System and AccessoriesBiosense Webster, Inc. · 2009 · cites it directly34descendants
  • K103746Carto 3 EP Navigation System, Version 2.0 (W/Standard Location Pad),CARTO3 Epnavigation…Biosense Webster, Inc. · 2011 · 2 generations on33descendants
  • K112007Carto 3 EP Navigation System, Version 2.2Biosense Webster, Inc. · 2011 · 3 generations on32descendants
  • K120550Carto (R) EP Navigation System, Version 3.0 (with Standard Location Pad)Biosense Webster, Inc. · 2012 · 4 generations on31descendants
  • K133916Carto 3 EP Navigation System, Version 4.2Biosense Webster, Inc. · 2014 · 5 generations on27descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K001437Ensite 3000 System, Model EE3000Endocardial Solutions, Inc.2000—
−5K012926Modification to Ensite 3000 System, Model EE3000Endocardial Solutions, Inc.2001—
−4K030129Modification to Ensite 3000 System, Model EE3000Endocardial Solutions, Inc.2003—
−3K042493Ensite System, Model EE3000Endocardial Solutions, Inc.2004—
−2K060954Modification to Ensite System, Model EE3000St. Jude Medical, Inc.-Endocardial Solutions2006—
−1K072202Carto 3 Navigation System (Model:FG-5400-00), Carto 3 External Reference…Biosense Webster, Inc.2007—
0 (this device)K090017Carto 3 EP Navigation System, V1.0 and AccessoriesBiosense Webster, Inc.2009—
+1K093455Carto 3 Version 1.05 EP Navigation System and AccessoriesBiosense Webster, Inc.20091
+2K103746Carto 3 EP Navigation System, Version 2.0 (W/Standard Location Pad),CARTO3…Biosense Webster, Inc.20111
+3K112007Carto 3 EP Navigation System, Version 2.2Biosense Webster, Inc.20112
+4K120550Carto (R) EP Navigation System, Version 3.0 (with Standard Location Pad)Biosense Webster, Inc.20122
+5K133916Carto 3 EP Navigation System, Version 4.2Biosense Webster, Inc.20142

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
200010
200110
200310
200510
200620
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201010
201120
201210
201420
201610
201720
201850
201920
202020
202120
202230
202350
202440
202570
202650
FIG. 1 — Lineage of K090017, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQK device

  1. #1 K213682 HemoSphere Advanced Monitor, HemoSphere…score 26.6 · 28 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K072202Carto 3 Navigation System (Model:FG-5400-00), Carto 3 External Reference Patches…
Summary page 1
…ogrammable diagnostic computer (per 21 CFR 870.1425, Product Code DQK) Predicate Device: CARTO III EP Navigation System (K072202) CARTO RMT V8 EP Navigation System (K060047) CARTO V9 XP EP Nav…
Biosense Webster, Inc.2007—
*K060047Carto RMT EP Navigation System V8
Summary page 1
…70.1425, Product Code DQK) Predicate Device: CARTO III EP Navigation System (K072202) CARTO RMT V8 EP Navigation System (K060047) CARTO V9 XP EP Navigation System (K070240) Manufacturing Facilities: System & System Ca…
Biosense Webster, Inc.2006—
*K070240Carto XP EP Navigation System, Version 9
Summary page 1
…vigation System (K072202) CARTO RMT V8 EP Navigation System (K060047) CARTO V9 XP EP Navigation System (K070240) Manufacturing Facilities: System & System Cables Biosense Webster (Israel) Ltd.…
Biosense Webster, Inc.2007—

Also cleared against K072202: 1 other device →

What descends from it? (45)

See all 45

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093455Carto 3 Version 1.05 EP Navigation System and AccessoriesBiosense Webster, Inc.200934
K093566Carto XP EP Navigation System, Version 10Biosense Webster, Inc.201014

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 9, 2012Not classifiedZ-1230-2012Nonconforming Material/ComponentSome CARTO 3 Systems may get radio frequency interference through certain interconnected cable configurations resulting in jumping or disappearing catheter icon and/or error messages during ablation.
Feb 15, 2011Not classifiedZ-1839-2011OtherThe recall was initiated because Biosense Webster has confirmed that there is an inaccessibility of electronic version of the instruction for use (e-IFU) for the "10 ft. Interface Cable, 10 Pin-Shielded Tip Pins to 12 Pin-Blue", catalog…
Oct 5, 2010Not classifiedZ-0884-2011Device DesignWhen connecting this cable to the mapping system, loss of Body Surface ECG signals may occur. This may result in inaccurate locations appearing for catheters displayed 3D on the CARTO monitor.
Nov 17, 2009Class IIZ-2369-2012Labeling mix-upsThe recall was initiated because Biosense Webster has become aware of information regarding a potential hazard that may occur during the operation of a Bloom EP Stimulator device in conjunction with RF generators and other…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 3 companies; a new generation was cleared every 1.6 years on median.
  • None of the 14 ancestors has a recall.
  • Of its descendants, K093455 (Biosense Webster, Inc., 2009) has gone furthest, with 34 descendants of its own.
  • 1 other device also cites its predicate K072202.

How does its category compare?

All DQK devices

DQK Computer, Diagnostic, Programmable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · DQK

23so far in 2026, the most since 2007

2007: 132008: 42009: 122010: 92011: 122012: 72013: 132014: 92015: 72016: 152017: 72018: 142019: 152020: 162021: 162022: 192023: 212024: 132025: 222026 (partial year): 23
2007high 23 (2026) · low 4 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQK

95in 2025

2007: 83 days median2008: 70 days median2009: 93 days median2010: 64 days median2011: 69 days median2012: 74 days median2013: 68 days median2014: 90 days median2015: 125 days median2016: 190 days median2017: 146 days median2018: 150 days median2019: 113 days median2020: 123 days median2021: 132 days median2022: 136 days median2023: 133 days median2024: 100 days median2025: 95 days median2026 (partial year): 151 days median
2007high 190 (2016) · low 64 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQK

1in 2025

2007: 42008: 12009: 22010: 32011: 22012: 32013: 12014: 42015: 22017: 22018: 62019: 12020: 12021: 12022: 22023: 22024: 242025: 12026 (partial year): 2
2007high 24 (2024) · low 0 (2016)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQK

433in 2025

2007: 182008: 342009: 402010: 492011: 552012: 1292013: 1802014: 1152015: 1312016: 3402017: 2522018: 2612019: 2182020: 2152021: 3512022: 3192023: 3402024: 5572025: 433
2007high 557 (2024) · low 18 (2007)2025

MAUDE reports naming product code DQK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071383418
2008470134
20091293240
2010964349
20111269255
20127743129
201313681180
20149904115
201571252131
2016151900340
201771462252
2018141506261
2019151131218
2020161231215
2021161321351
2022191362319
2023211332340
20241310024557
202522951433
2026231512—

What is it indicated for?

The CARTO® 3 V1.0 System is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as anatomical maps, cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps, cardiac chamber geometry maps and ECG fragmentation maps. The acquired patient signals, including body surface…

Read the full statement (226 words)

The CARTO® 3 V1.0 System is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as anatomical maps, cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps, cardiac chamber geometry maps and ECG fragmentation maps. The acquired patient signals, including body surface ECG and intracardiac electrograms may also be displayed in real time on the System's display screen. The CARTO® 3 V1.0 System is also intended to support EP procedures, maintaining CARTO® System capabilities, in the presence of a high metallic environment and magnetic field strengths upto 0.1 T and provide a data communication channel to the Stereotaxis Niobe® Catheter Navigation System. The CARTO® 3 V1.0 System includes CARTOMERGE® PLUS functionality to import, register and merge CT or MRI structural images with CARTO® map's physiological information and real time catheter navigation. The system includes the Fast Anatomical Mapping (FAM) functionality that allows for the quick creation of cardiac anatomical volumes using catheters with magnetic location sensors. The system's CARTOSOUND image integration functionality enables integration of intracardiac echo (ICE) to enable visualization of 3D combined maps. In addition to the use of specialized navigation catheters with magnetic location sensors, the system is also intended for use with conventional, non-navigational, electrophysiology catheters without magnetic location sensors.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K090017: Carto 3 EP Navigation System, V1.0 and Accessories.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K090017

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026