510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Carto (R) EP Navigation System, Version 3.0 (with Standard Location Pad)

K120550510(k) TraditionalCleared May 7, 201274-day reviewClass II21 CFR 870.1425Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biosense Webster, Inc. · DQK Computer, Diagnostic, Programmable

Cleared on May 7, 2012 as substantially equivalent to K112007 (Biosense Webster, Inc., 2011) and 2 other predicates after a 74-day review. Its family tree reaches back 10 generations to 2000. 31 later devices descend from it. It has been recalled 2 times.

Decided
May 7, 2012
Review
74 days
Class
Class II
Product code
DQK
Recalls
2
Predicates
3

Its longest line of descent runs back 10 generations to K001437 (2000). 21 ancestors in all, the oldest K982415 (1998). 31 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2006: 3 other ancestors2005: 1 other ancestor2010: 2 other ancestors2003: 1 other ancestor2001: 1 other ancestor2000: 1 other ancestor1998: 1 other ancestor2016: 1 descendant cleared2014: 2 descendants cleared2017: 2 descendants cleared2018: 5 descendants cleared2023: 3 descendants cleared2024: 4 descendants cleared2019: 2 descendants cleared2022: 2 descendants cleared2020: 1 descendant cleared2021: 1 descendant cleared2025: 3 descendants cleared2026: 5 descendants clearedK001437 · Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2000K012926 · Modification to Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2001K030129 · Modification to Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2003K042493 · Ensite System, Model EE3000 · Endocardial Solutions, Inc. · 2004K060954 · Modification to Ensite System, Model EE3000 · St. Jude Medical, Inc.-Endocardial Solutions · 2006K072202 · Carto 3 Navigation System (Model:FG-5400-00), Carto 3 External Reference Patches (Model:D-1283-01) · Biosense Webster, Inc. · 2007×2: 2 generations cleared close together, K090017 (2009), K093455 (2009) K103746 · Carto 3 EP Navigation System, Version 2.0 (W/Standard Location Pad),CARTO3 Epnavigation System,Version 2.0 W/RMT Pad) · Biosense Webster, Inc. · 2011 · recalledK112007 · Carto 3 EP Navigation System, Version 2.2 · Biosense Webster, Inc. · 2011 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 2 other ancestors5 generations back: 2 other ancestors6 generations back: 2 other ancestors 1 generation forward: 3 descendants2 generations forward: 7 descendants3 generations forward: 4 descendants4 generations forward: 1 descendant5 generations forward: 2 descendants6 generations forward: 3 descendants7 generations forward: 2 descendants8 generations forward: 5 descendants9 generations forward: 4 descendants K112007 · Carto 3 EP Navigation System, Version 2.2 · Biosense Webster, Inc. · 2011 · recalledK103746 · Carto 3 EP Navigation System, Version 2.0 (W/Standard Location Pad),CARTO3 Epnavigation System,Version 2.0 W/RMT Pad) · Biosense Webster, Inc. · 2011 · recalledK093455 · Carto 3 Version 1.05 EP Navigation System and Accessories · Biosense Webster, Inc. · 2009 · recalledK090017 · Carto 3 EP Navigation System, V1.0 and Accessories · Biosense Webster, Inc. · 2009 · recalledK072202 · Carto 3 Navigation System (Model:FG-5400-00), Carto 3 External Reference Patches (Model:D-1283-01) · Biosense Webster, Inc. · 2007K060954 · Modification to Ensite System, Model EE3000 · St. Jude Medical, Inc.-Endocardial Solutions · 2006K042493 · Ensite System, Model EE3000 · Endocardial Solutions, Inc. · 2004K030129 · Modification to Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2003K012926 · Modification to Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2001K001437 · Ensite 3000 System, Model EE3000 · Endocardial Solutions, Inc. · 2000−10−50+5
  1. Root · 10 generations backK001437 Ensite 3000 System, Model EE3000Endocardial Solutions, Inc. · 2000
  2. 6 generations between
  3. 3 generations backK093455 Carto 3 Version 1.05 EP Navigation System and AccessoriesBiosense Webster, Inc. · 2009 · Recalled
  4. 2 generations backK103746 Carto 3 EP Navigation System, Version 2.0 (W/Standard Location…Biosense Webster, Inc. · 2011 · Recalled
  5. 1 generation backK112007 Carto 3 EP Navigation System, Version 2.2Biosense Webster, Inc. · 2011 · Recalled
  6. This deviceK120550 Carto (R) EP Navigation System, Version 3.0 (with Standard Location…Biosense Webster, Inc. · 2012
  7. Descendant · cites it directly · 27 of its ownK133916 Carto 3 EP Navigation System, Version 4.2Biosense Webster, Inc. · 2014 · Recalled
  8. Descendant · 2 generations on · 20 of its ownK170600 CARTO 3 EP Navigation System, Version 6.0 and AccessoriesBiosense Webster, Inc. · 2017
  9. Descendant · 3 generations on · 19 of its ownK180238 CARTO 3 EP Navigation System, Version 6.0 and Accessories with…Biosense Webster, Inc. · 2018
  10. Descendant · 4 generations on · 16 of its ownK191660 CARTO 3 EP Navigation System, Version 7.1 and AccessoriesBiosense Webster, Inc. · 2019 · Recalled
  11. Descendant · 5 generations on · 15 of its ownK200484 CARTO® 3 EP Navigation System with Signal Processing UnitBiosense Webster, Inc. · 2020
Show all 10 generations
  1. Root · 10 generations backK001437 Ensite 3000 System, Model EE3000Endocardial Solutions, Inc. · 2000
  2. 9 generations backK012926 Modification to Ensite 3000 System, Model EE3000Endocardial Solutions, Inc. · 2001
  3. 8 generations backK030129 Modification to Ensite 3000 System, Model EE3000Endocardial Solutions, Inc. · 2003
  4. 7 generations backK042493 Ensite System, Model EE3000Endocardial Solutions, Inc. · 2004
  5. 6 generations backK060954 Modification to Ensite System, Model EE3000St. Jude Medical, Inc.-Endocardial Solutions · 2006
  6. 5 generations backK072202 Carto 3 Navigation System (Model:FG-5400-00), Carto 3 External…Biosense Webster, Inc. · 2007
  7. 4 generations backK090017 Carto 3 EP Navigation System, V1.0 and AccessoriesBiosense Webster, Inc. · 2009 · Recalled
  8. 3 generations backK093455 Carto 3 Version 1.05 EP Navigation System and AccessoriesBiosense Webster, Inc. · 2009 · Recalled
  9. 2 generations backK103746 Carto 3 EP Navigation System, Version 2.0 (W/Standard Location…Biosense Webster, Inc. · 2011 · Recalled
  10. 1 generation backK112007 Carto 3 EP Navigation System, Version 2.2Biosense Webster, Inc. · 2011 · Recalled
  11. This deviceK120550 Carto (R) EP Navigation System, Version 3.0 (with Standard Location…Biosense Webster, Inc. · 2012

Showing 5 of 31 descendants. See all descendants

Descendants cleared per year

2014: 22016: 12017: 22018: 52019: 22020: 12021: 12022: 22023: 32024: 42025: 32026: 5
2014high 5 (2018) · low 0 (2015)2026

Most influential descendants

Showing 5 of 31 · See all
  • K133916Carto 3 EP Navigation System, Version 4.2Biosense Webster, Inc. · 2014 · cites it directly27descendants
  • K170600CARTO 3 EP Navigation System, Version 6.0 and AccessoriesBiosense Webster, Inc. · 2017 · 2 generations on20descendants
  • K180238CARTO 3 EP Navigation System, Version 6.0 and Accessories with VISITAG SURPOINTBiosense Webster, Inc. · 2018 · 3 generations on19descendants
  • K191660CARTO 3 EP Navigation System, Version 7.1 and AccessoriesBiosense Webster, Inc. · 2019 · 4 generations on16descendants
  • K200484CARTO® 3 EP Navigation System with Signal Processing UnitBiosense Webster, Inc. · 2020 · 5 generations on15descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K001437Ensite 3000 System, Model EE3000Endocardial Solutions, Inc.2000—
−9K012926Modification to Ensite 3000 System, Model EE3000Endocardial Solutions, Inc.2001—
−8K030129Modification to Ensite 3000 System, Model EE3000Endocardial Solutions, Inc.2003—
−7K042493Ensite System, Model EE3000Endocardial Solutions, Inc.2004—
−6K060954Modification to Ensite System, Model EE3000St. Jude Medical, Inc.-Endocardial Solutions2006—
−5K072202Carto 3 Navigation System (Model:FG-5400-00), Carto 3 External Reference…Biosense Webster, Inc.2007—
−4K090017Carto 3 EP Navigation System, V1.0 and AccessoriesBiosense Webster, Inc.20094
−3K093455Carto 3 Version 1.05 EP Navigation System and AccessoriesBiosense Webster, Inc.20091
−2K103746Carto 3 EP Navigation System, Version 2.0 (W/Standard Location Pad),CARTO3…Biosense Webster, Inc.20111
−1K112007Carto 3 EP Navigation System, Version 2.2Biosense Webster, Inc.20112
0 (this device)K120550Carto (R) EP Navigation System, Version 3.0 (with Standard Location Pad)Biosense Webster, Inc.2012—
+1K133916Carto 3 EP Navigation System, Version 4.2Biosense Webster, Inc.20142
+2K170600CARTO 3 EP Navigation System, Version 6.0 and AccessoriesBiosense Webster, Inc.2017—
+3K180238CARTO 3 EP Navigation System, Version 6.0 and Accessories with VISITAG SURPOINTBiosense Webster, Inc.2018—
+4K191660CARTO 3 EP Navigation System, Version 7.1 and AccessoriesBiosense Webster, Inc.20191
+5K200484CARTO® 3 EP Navigation System with Signal Processing UnitBiosense Webster, Inc.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
200010
200110
200310
200510
200630
200710
201020

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201420
201610
201720
201850
201920
202010
202110
202220
202330
202440
202530
202650
FIG. 1 — Lineage of K120550, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQK device

  1. #1 K213682 HemoSphere Advanced Monitor, HemoSphere…score 26.6 · 28 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K112007Carto 3 EP Navigation System, Version 2.2
Summary page 1
…Class 11, 21 CFR 870.1425 Product Code DQK Predicate Device: CARTo® 3 SYSTEM V2.2 and Accessories 510(k) KI 112007 (October 5, 2011) Reference Devices: Reference Device 1: CARTO® XP EP Navig…
Biosense Webster, Inc.20112
*K070240Carto XP EP Navigation System, Version 9
Summary page 1
…ence Devices: Reference Device 1: CARTO® XP EP Navigation System 2007, Version 9, 510(k) K070240 (May 4, 2007) Reference Device 2: Stereotaxis NIOBE® Mag…
Biosense Webster, Inc.2007—
*K060967Niobe Magnetic Navigation System W/Navigant Workstation
Summary page 1
…Reference Device 2: Stereotaxis NIOBE® Magnetic Navigation System 5 10(k) K060967 (October 19, 2006) Manufacturing Facilities: Biosense Webster (Israel), Ltd. a Joh…
Stereotaxis, Inc.2006—

What descends from it? (31)

See all 31

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K133916Carto 3 EP Navigation System, Version 4.2Biosense Webster, Inc.201427
K132782CARTO3 EP Navigation System, Version 3.2Biosense Webster, Inc.20144
K152160Navik 3DApn Health, LLC20160

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 12, 2014Class IIZ-1045-2014OtherBiosense Webster has initiated a recall of the CartoUnivu Module within the Carto 3 System (V3.2.2 and V3.2.3) when used with Siemens Axiom Artis VB Fluoroscopy systems in particular procedural configurations only. If the table rotation…
Mar 5, 2013Class IIZ-1023-2013Employee errorBiosense is recalling the 20-Pole Eco Cable because it was incorrectly calibrated. When a catheter is connected to an affected to an affected 20-Pole Eco Cable, the catheter will not be recognized by the impacted 54 cable serial numbers…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 1.5 years on median.
  • 4 of the 21 ancestors have a recall of any class.
  • Of its descendants, K133916 (Biosense Webster, Inc., 2014) has gone furthest, with 27 descendants of its own.

How does its category compare?

All DQK devices

DQK Computer, Diagnostic, Programmable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · DQK

23so far in 2026, the most since 2007

2007: 132008: 42009: 122010: 92011: 122012: 72013: 132014: 92015: 72016: 152017: 72018: 142019: 152020: 162021: 162022: 192023: 212024: 132025: 222026 (partial year): 23
2007high 23 (2026) · low 4 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQK

95in 2025

2007: 83 days median2008: 70 days median2009: 93 days median2010: 64 days median2011: 69 days median2012: 74 days median2013: 68 days median2014: 90 days median2015: 125 days median2016: 190 days median2017: 146 days median2018: 150 days median2019: 113 days median2020: 123 days median2021: 132 days median2022: 136 days median2023: 133 days median2024: 100 days median2025: 95 days median2026 (partial year): 151 days median
2007high 190 (2016) · low 64 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQK

1in 2025

2007: 42008: 12009: 22010: 32011: 22012: 32013: 12014: 42015: 22017: 22018: 62019: 12020: 12021: 12022: 22023: 22024: 242025: 12026 (partial year): 2
2007high 24 (2024) · low 0 (2016)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQK

433in 2025

2007: 182008: 342009: 402010: 492011: 552012: 1292013: 1802014: 1152015: 1312016: 3402017: 2522018: 2612019: 2182020: 2152021: 3512022: 3192023: 3402024: 5572025: 433
2007high 557 (2024) · low 18 (2007)2025

MAUDE reports naming product code DQK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071383418
2008470134
20091293240
2010964349
20111269255
20127743129
201313681180
20149904115
201571252131
2016151900340
201771462252
2018141506261
2019151131218
2020161231215
2021161321351
2022191362319
2023211332340
20241310024557
202522951433
2026231512—

What is it indicated for?

The CARTO® 3 System V3.0 is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as anatomical maps, cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps, cardiac chamber geometry maps and ECG fragmentation maps. The acquired patient signals, including body surface…

Read the full statement (246 words)

The CARTO® 3 System V3.0 is intended for catheter-based atrial and ventricular mapping. The mapping system allows real-time display of cardiac maps in a number of different formats. Maps may be displayed as anatomical maps, cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, impedance maps, cardiac chamber geometry maps and ECG fragmentation maps. The acquired patient signals, including body surface ECG and intracardiac electrograms, may also be displayed in real time on the system's display screen. The CARTO® 3 System V3.0 is also intended to support EP procedures, maintaining CARTO® System capabilities in the presence of a high metallic environment and magnetic field strengths up to 0.1 T and provide a data communication channel to the Stereotaxis Niobe® Catheter Navigation System. The CARTO® 3 System V3.0 includes CARTOMERGE® PLUS Module functionality to import, register and merge CT or MRI structural images with CARTO® Map's physiological information and real time catheter navigation. The system includes the Fast Anatomical Mapping (FAM) functionality that allows for the quick creation of cardiac anatomical volumes using catheters with magnetic location sensors. The system's CARTOSOUND® image integration functionality enables integration of intracardiac echo (ICE) to enable visualization of 3D combined maps. A fluoro background capability enables visualization of X-ray images as reference to the CARTO® Maps images and catheters. In addition to the use of specialized navigation catheters with magnetic location sensors, the system is also intended for use with conventional, non-navigational, electrophysiology catheters without magnetic location sensors.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K120550: Carto (R) EP Navigation System, Version 3.0 (with Standard Location Pad).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K120550

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026