Ultimum/Engage/Engage TR

K091137510(k) TraditionalCleared Apr 22, 20092-day reviewClass II21 CFR 870.1340Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

St Jude Medical · DYB Introducer, Catheter

Cleared on Apr 22, 2009 as substantially equivalent to K001346 (St. Jude Medical, Inc., 2000) and 1 other predicate after a 2-day review. Its family tree reaches back 4 generations to 1995. 1 of the 13 devices in its family, counting itself, has a Class I recall. 2 later devices descend from it. It has been recalled 5 times.

Decided
Apr 22, 2009
Review
2 days
Class
Class II
Product code
DYB
Recalls
5
Predicates
2

Its longest line of descent runs back 4 generations to K954234 (1995). 10 ancestors in all, the oldest K810675 (1981). 2 later devices descend from it, 2 citing it directly. 1 of the 13 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor1996: 1 other ancestor1995: 1 other ancestor1993: 1 other ancestor1991: 1 other ancestor1981: 1 other ancestor, 1 with a Class I recall12009: 2 descendants clearedK954234 · Radifocus Introducer II Kit · Terumo Medical Corp. · 1995K033681 · Glidesheath · Terumo Corp. · 2003K062858 · Modification to Glidesheath · Terumo Medical Corp. · 2006K082644 · Glidesheath · Terumo Medical Corp. · 200819801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 4 other ancestors, 1 with a Class I recall1 1 generation forward: 2 descendants K082644 · Glidesheath · Terumo Medical Corp. · 2008K062858 · Modification to Glidesheath · Terumo Medical Corp. · 2006K033681 · Glidesheath · Terumo Corp. · 2003K954234 · Radifocus Introducer II Kit · Terumo Medical Corp. · 19950
  1. Root · 4 generations backK954234 Radifocus Introducer II KitTerumo Medical Corp. · 1995
  2. 3 generations backK033681 GlidesheathTerumo Corp. · 2003
  3. 2 generations backK062858 Modification to GlidesheathTerumo Medical Corp. · 2006
  4. 1 generation backK082644 GlidesheathTerumo Medical Corp. · 2008
  5. This deviceK091137 Ultimum/Engage/Engage TRSt Jude Medical · 2009
  6. Descendant · cites it directly · 0 of its ownK093130 Engage TR IntroducerSt Jude Medical · 2009 · Recalled
  7. Descendant · cites it directly · 0 of its ownK092575 Engage/Engage TRSt Jude Medical · 2009 · Recalled

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K093130Engage TR IntroducerSt Jude Medical · 2009 · cites it directly0descendants
  • K092575Engage/Engage TRSt Jude Medical · 2009 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K954234Radifocus Introducer II KitTerumo Medical Corp.1995—
−3K033681GlidesheathTerumo Corp.2003—
−2K062858Modification to GlidesheathTerumo Medical Corp.2006—
−1K082644GlidesheathTerumo Medical Corp.2008—
0 (this device)K091137Ultimum/Engage/Engage TRSt Jude Medical2009—
+1K093130Engage TR IntroducerSt Jude Medical20095
+1K092575Engage/Engage TRSt Jude Medical20095

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198111
199110
199310
199510
199610
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200920
FIG. 1 — Lineage of K091137, 1995–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DYB device

  1. #1 K251051 AuST Steerable Sheathscore 18.0 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001346Ultimum Hemostasis Introducer, Model 4076XX
Summary page 2
…onal device testing was required. 5.6. PREDICATE DEVICE The predicate devices are provided below: Product Name 510k # Ultimum Hemostasis K001346 (cleared 05/24/2000) Introducer (St. Jude Medical) GlideSheath (Terumo K082644 (cleared 09/18/2008) Corporation)…
St. Jude Medical, Inc.2000—
*K082644Glidesheath
Summary page 2
…ded below: Product Name 510k # Ultimum Hemostasis K001346 (cleared 05/24/2000) Introducer (St. Jude Medical) GlideSheath (Terumo K082644 (cleared 09/18/2008) Corporation) 7 Ultimum/Engage Introduce…
Terumo Medical Corp.2008—

Also cleared against K001346: 2 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K093130Engage TR IntroducerSt Jude Medical20090
K092575Engage/Engage TRSt Jude Medical20090

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 28, 2010Not classifiedZ-2179-2010Process designPotential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak…
Jun 28, 2010Not classifiedZ-2181-2010Process designPotential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak…
Jun 28, 2010Not classifiedZ-2180-2010Process designPotential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak…
Jun 28, 2010Not classifiedZ-2178-2010Process designPotential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak…
Jun 28, 2010Not classifiedZ-2182-2010Process designPotential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K810675Arrow Radial Artery Catheterization SetArrow Intl., Inc. · 1981 · 3 generations back · ◆ in the 1981 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 2.9 years on median.
  • 2 of the 10 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K093130 (St Jude Medical, 2009) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K001346.

How does its category compare?

All DYB devices

DYB Introducer, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · DYB

16in 2025

2007: 122008: 172009: 142010: 182011: 172012: 222013: 262014: 302015: 172016: 192017: 242018: 192019: 182020: 92021: 122022: 132023: 152024: 182025: 162026 (partial year): 17
2007high 30 (2014) · low 9 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DYB

63in 2025

2007: 95 days median2008: 37 days median2009: 52 days median2010: 123 days median2011: 100 days median2012: 128 days median2013: 118 days median2014: 122 days median2015: 129 days median2016: 95 days median2017: 70 days median2018: 156 days median2019: 47 days median2020: 49 days median2021: 78 days median2022: 143 days median2023: 55 days median2024: 56 days median2025: 63 days median2026 (partial year): 59 days median
2007high 156 (2018) · low 37 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DYB

3in 2025

2007: 152008: 92009: 252010: 182011: 42012: 32013: 5, 1 Class I2014: 22015: 2, 1 Class I2016: 45, 3 Class I2017: 22, 1 Class I2018: 12019: 22020: 21, 21 Class I2021: 62022: 22023: 102024: 22025: 32026 (partial year): 23, 21 Class I
2007high 45 (2016) · low 1 (2018)2026

Class I (48 since 2007) Class II/III not classified

Adverse event reports per year · DYB

4,539in 2025

2007: 3162008: 4232009: 3992010: 4452011: 6282012: 8982013: 7432014: 7802015: 9132016: 1,2342017: 1,6722018: 1,9032019: 2,4502020: 2,4652021: 2,8352022: 4,0842023: 3,7752024: 3,9312025: 4,539
2007high 4,539 (2025) · low 316 (2007)2025

MAUDE reports naming product code DYB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007129515316
200817379423
2009145225399
20101812318445
2011171004628
2012221283898
2013261185 (1)743
2014301222780
2015171292 (1)913
2016199545 (3)1,234
2017247022 (1)1,672
20181915611,903
2019184722,450
202094921 (21)2,465
2021127862,835
20221314324,084
20231555103,775
2024185623,931
2025166334,539
2026175923 (21)—

What is it indicated for?

The introduction of angiographic catheters, closed end catheters, balloon catheters, and electrodes into a blood vessel (including but not limited to femoral, radial, and brachial access) where minimizing blood loss is essential.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091137: Ultimum/Engage/Engage TR.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K091137

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026