510(k) Report

510(k) · Cleared

Recalled · 2010

Engage TR Introducer

K093130510(k) SpecialCleared Nov 4, 200930-day reviewClass II21 CFR 870.1340Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

St Jude Medical · DYB Introducer, Catheter

Cleared on Nov 4, 2009 as substantially equivalent to K091137 (St Jude Medical, 2009) and 3 other predicates after a 30-day review. Its family tree reaches back 5 generations to 1995. 1 of the 17 devices in its family, counting itself, has a Class I recall. It has been recalled 5 times.

Decided
Nov 4, 2009
Review
30 days
Class
Class II
Product code
DYB
Recalls
5
Predicates
4

Its longest line of descent runs back 5 generations to K954234 (1995). 16 ancestors in all, the oldest K810675 (1981). No later device cites it as a predicate yet. 1 of the 17 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year cleared 2007: 1 other ancestor2000: 1 other ancestor1991: 2 other ancestors2005: 1 other ancestor2001: 1 other ancestor1996: 1 other ancestor1995: 1 other ancestor1993: 2 other ancestors1981: 1 other ancestor, 1 with a Class I recall1K954234 · Radifocus Introducer II Kit · Terumo Medical Corp. · 1995K033681 · Glidesheath · Terumo Corp. · 2003K062858 · Modification to Glidesheath · Terumo Medical Corp. · 2006K082644 · Glidesheath · Terumo Medical Corp. · 2008K091137 · Ultimum/Engage/Engage TR · St Jude Medical · 2009 · recalled1980198519901995200020052010
Other ancestors, by generation 1 generation back: 3 other ancestors2 generations back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 5 other ancestors, 1 with a Class I recall1 K091137 · Ultimum/Engage/Engage TR · St Jude Medical · 2009 · recalledK082644 · Glidesheath · Terumo Medical Corp. · 2008K062858 · Modification to Glidesheath · Terumo Medical Corp. · 2006K033681 · Glidesheath · Terumo Corp. · 2003K954234 · Radifocus Introducer II Kit · Terumo Medical Corp. · 1995−50
  1. Root · 5 generations backK954234 Radifocus Introducer II KitTerumo Medical Corp. · 1995
  2. 4 generations backK033681 GlidesheathTerumo Corp. · 2003
  3. 3 generations backK062858 Modification to GlidesheathTerumo Medical Corp. · 2006
  4. 2 generations backK082644 GlidesheathTerumo Medical Corp. · 2008
  5. 1 generation backK091137 Ultimum/Engage/Engage TRSt Jude Medical · 2009 · Recalled
  6. This deviceK093130 Engage TR IntroducerSt Jude Medical · 2009

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K954234Radifocus Introducer II KitTerumo Medical Corp.1995—
−4K033681GlidesheathTerumo Corp.2003—
−3K062858Modification to GlidesheathTerumo Medical Corp.2006—
−2K082644GlidesheathTerumo Medical Corp.2008—
−1K091137Ultimum/Engage/Engage TRSt Jude Medical20095
0 (this device)K093130Engage TR IntroducerSt Jude Medical2009—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198111
199120
199320
199510
199610
200010
200110
200510
200710
FIG. 1 — Lineage of K093130, 1995–2009Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DYB device

  1. #1 K251051 AuST Steerable Sheathscore 18.0 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K091137Ultimum/Engage/Engage TR
Summary page 1
…DYB 4.3 Device Description The modified Engage TR Introducer is essentially a modification of the previously cleared Engage/Engage TR/Ultimum introducer (K09 1137). The modified Engage TR Introducer is intended to provide easy access to the vascular system, while providing convenient tempo…
St Jude Medical20095
*K001346Ultimum Hemostasis Introducer, Model 4076XX
Summary page 2
…overed by this submission, is substantially equivalent to the St. Jude Medical Engage/Engage TR Introducer (K091137, April, 22, 2009), Ultimum Hemostasis Introducer (K001346, May 24, 2000), Strada Carotid Guiding Sheath, (K070166, April 06, 2007), and Fast Cath Hemostasis Introducer (K914090, October…
St. Jude Medical, Inc.2000—
*K070166Strada Carotid Guiding Sheath
Summary page 2
…o the St. Jude Medical Engage/Engage TR Introducer (K091137, April, 22, 2009), Ultimum Hemostasis Introducer (K001346, May 24, 2000), Strada Carotid Guiding Sheath, (K070166, April 06, 2007), and Fast Cath Hemostasis Introducer (K914090, October 28, 1991). St. Jude Medical…
St Jude Medical2007—
*K914090Daig Fast Cath (TM) Introducer, Modification
Summary page 2
…ril, 22, 2009), Ultimum Hemostasis Introducer (K001346, May 24, 2000), Strada Carotid Guiding Sheath, (K070166, April 06, 2007), and Fast Cath Hemostasis Introducer (K914090, October 28, 1991). St. Jude Medical 4-2 Engage TR Introducer…
Daig Corp.19911

Also cleared against K091137: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 28, 2010Not classifiedZ-2179-2010Process designPotential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak…
Jun 28, 2010Not classifiedZ-2181-2010Process designPotential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak…
Jun 28, 2010Not classifiedZ-2180-2010Process designPotential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak…
Jun 28, 2010Not classifiedZ-2178-2010Process designPotential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak…
Jun 28, 2010Not classifiedZ-2182-2010Process designPotential for a partial or complete separation of the Shaft (sheath) from the hub or a material break in the hub assembly just distal to the hemostasis seal. If either of these were to occur during use, it is likely that fluids would leak…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K810675Arrow Radial Artery Catheterization SetArrow Intl., Inc. · 1981 · 4 generations back · ◆ in the 1981 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 14 years and passes through 3 companies; a new generation was cleared every 1.9 years on median.
  • 7 of the 16 ancestors have a recall of any class; 1 has a Class I recall.
  • 1 other device also cites its predicate K091137.

How does its category compare?

All DYB devices

DYB Introducer, Catheter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · DYB

16in 2025

2007: 122008: 172009: 142010: 182011: 172012: 222013: 262014: 302015: 172016: 192017: 242018: 192019: 182020: 92021: 122022: 132023: 152024: 182025: 162026 (partial year): 17
2007high 30 (2014) · low 9 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DYB

63in 2025

2007: 95 days median2008: 37 days median2009: 52 days median2010: 123 days median2011: 100 days median2012: 128 days median2013: 118 days median2014: 122 days median2015: 129 days median2016: 95 days median2017: 70 days median2018: 156 days median2019: 47 days median2020: 49 days median2021: 78 days median2022: 143 days median2023: 55 days median2024: 56 days median2025: 63 days median2026 (partial year): 59 days median
2007high 156 (2018) · low 37 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DYB

3in 2025

2007: 152008: 92009: 252010: 182011: 42012: 32013: 5, 1 Class I2014: 22015: 2, 1 Class I2016: 45, 3 Class I2017: 22, 1 Class I2018: 12019: 22020: 21, 21 Class I2021: 62022: 22023: 102024: 22025: 32026 (partial year): 23, 21 Class I
2007high 45 (2016) · low 1 (2018)2026

Class I (48 since 2007) Class II/III not classified

Adverse event reports per year · DYB

4,539in 2025

2007: 3162008: 4232009: 3992010: 4452011: 6282012: 8982013: 7432014: 7802015: 9132016: 1,2342017: 1,6722018: 1,9032019: 2,4502020: 2,4652021: 2,8352022: 4,0842023: 3,7752024: 3,9312025: 4,539
2007high 4,539 (2025) · low 316 (2007)2025

MAUDE reports naming product code DYB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007129515316
200817379423
2009145225399
20101812318445
2011171004628
2012221283898
2013261185 (1)743
2014301222780
2015171292 (1)913
2016199545 (3)1,234
2017247022 (1)1,672
20181915611,903
2019184722,450
202094921 (21)2,465
2021127862,835
20221314324,084
20231555103,775
2024185623,931
2025166334,539
2026175923 (21)—

What is it indicated for?

The introduction of angiographic catheters, closed end catheters, balloon catheters, and electrodes into a blood vessel (including but not limited to femoral, radial, and brachial access) where minimizing blood loss is essential.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K093130: Engage TR Introducer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K093130

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026