510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Acecide-C High Level Disinfectant and Sterilant

K091210510(k) TraditionalCleared Feb 5, 2010287-day reviewClass II21 CFR 880.6885Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Best Sanitizers, Inc. · MED Sterilant, Medical Devices

Cleared on Feb 5, 2010 as substantially equivalent to K041984 (Best Sanitizers, Inc., 2005) after a 287-day review. Its family tree reaches back 1 generation to 2005. It has been recalled once.

Decided
Feb 5, 2010
Review
287 days
Class
Class II
Product code
MED
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K041984 (2005). No later device cites it as a predicate yet.

K041984 · Acecide High-Level Disinfectant and Sterilant · Best Sanitizers, Inc. · 20052006200820102012
  1. Root · 1 generation backK041984 Acecide High-Level Disinfectant and SterilantBest Sanitizers, Inc. · 2005
  2. This deviceK091210 Acecide-C High Level Disinfectant and SterilantBest Sanitizers, Inc. · 2010
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K041984Acecide High-Level Disinfectant and SterilantBest Sanitizers, Inc.2005—
0 (this device)K091210Acecide-C High Level Disinfectant and SterilantBest Sanitizers, Inc.2010—
FIG. 1 — Lineage of K091210, 2005–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MED? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MED device

  1. #1 K251666 SYSTEM 1E Liquid Chemical Sterilant Processing…score 13.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K041984Acecide High-Level Disinfectant and Sterilant
Summary page 1
…High-Level Disinfectant and Sterilant Classification Name: Not Classified d. Predicate Name Acecide High Level Disinfectant and Sterilant (K04 1984) e. Summary of the substantial equivalence (SE) of ACECIDE-C High-Level Disinfectant and Sterilant to the pre…
Best Sanitizers, Inc.2005—

Also cleared against K041984: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 2, 2016Class IIZ-2714-2016Nonconforming Material/ComponentCertain lots of the acid solution in Acecide-C are degrading faster than normal.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • 1 other device also cites its predicate K041984.

How does its category compare?

All MED devices

MED Sterilant, Medical Devices: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · MED

2in 2023

2007: 22008: 12010: 62012: 32013: 12014: 12016: 12017: 12018: 22019: 42021: 22022: 12023: 22026 (partial year): 2
2007high 6 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MED

84in 2019

2010: 281 days median2012: 111 days median2019: 84 days median
2007high 281 (2010) · low 84 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MED

2in 2011

2007: 32008: 22011: 2
2007high 3 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MED

9in 2025

2007: 342008: 792009: 1582010: 892011: 1062012: 3372013: 1402014: 1052015: 1102016: 1352017: 742018: 422019: 352020: 322021: 312022: 162023: 92024: 82025: 9
2007high 337 (2012) · low 8 (2024)2025

MAUDE reports naming product code MED, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—334
20081—279
20090—0158
20106281089
20110—2106
201231110337
20131—0140
20141—0105
20150—0110
20161—0135
20171—074
20182—042
2019484035
20200—032
20212—031
20221—016
20232—09
20240—08
20250—09
20262—0—

What is it indicated for?

ACECIDE-C is a high level disinfectant intended to disinfect reusable clean heat-sensitive semi-critical medical devices (such as flexible endloscopes) which contact mucous membranes when used at or above its minimum recommended concentration of 2000 ppm peracetic acid for 7.0 minutes at 200C. ACECIDE-C is a sterilant intended to sterilize reusable clean heat-sensitive critical and semi-critical medical devices (such as flexible endloscopes) which…

Read the full statement (131 words)

ACECIDE-C is a high level disinfectant intended to disinfect reusable clean heat-sensitive semi-critical medical devices (such as flexible endloscopes) which contact mucous membranes when used at or above its minimum recommended concentration of 2000 ppm peracetic acid for 7.0 minutes at 200C.

ACECIDE-C is a sterilant intended to sterilize reusable clean heat-sensitive critical and semi-critical medical devices (such as flexible endloscopes) which contact and potentially penetrate into sterile body areas, for which there is no other practical method of sterilization, when used at or above its minimum recbmmended concentration of 2000 ppm peracetic acid for 2.0 hours at 200 C.

ACECIDE Test Strips are used with ACECIE-C High Level Disinfectant and Sterilant use solution to ensure that the level of peracetic acid is above the minimum recommended concentration of 2000 ppm.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091210: Acecide-C High Level Disinfectant and Sterilant.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K091210

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026