510(k) Report

510(k) · Cleared

Recalled · Class II · 2026

Medivators Advantage Plus Endoscope Reprocessing System, Rapicide PA High Level Disinfectant

K082988510(k) TraditionalCleared Feb 13, 2009129-day reviewClass II21 CFR 876.1500Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medivators Reprocessing Systems · FEB Accessories, Cleaning, For Endoscope

Cleared on Feb 13, 2009 as substantially equivalent to K063876 (Medivators Reprocessing Systems, 2007) and 3 other predicates after a 129-day review. Its family tree reaches back 3 generations to 1994. 7 later devices descend from it. It has been recalled 4 times.

Decided
Feb 13, 2009
Review
129 days
Class
Class II
Product code
FEB
Recalls
4
Predicates
4

Its longest line of descent runs back 3 generations to K914145 (1994). 7 ancestors in all. 7 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor2005: 1 other ancestor1997: 1 other ancestor1999: 1 other ancestor2010: 1 descendant cleared2017: 1 descendant cleared2016: 1 descendant cleared2020: 1 descendant cleared2025: 2 descendants cleared2023: 1 descendant clearedK914145 · DSD-91(TM) Disinfector for Flexible Endoscopes · Medivators, Inc. · 1994 · recalledK040883 · Evotech Integrated Endoscope Disinfection System · Advanced Sterilization Products · 2005 · recalledK061899 · Evotech Integrated Endoscope Disinfection System · Advanced Sterilization Products · 2006 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 3 other ancestors3 generations back: 1 other ancestor 1 generation forward: 4 descendants2 generations forward: 3 descendants K061899 · Evotech Integrated Endoscope Disinfection System · Advanced Sterilization Products · 2006 · recalledK040883 · Evotech Integrated Endoscope Disinfection System · Advanced Sterilization Products · 2005 · recalledK914145 · DSD-91(TM) Disinfector for Flexible Endoscopes · Medivators, Inc. · 1994 · recalled0
  1. Root · 3 generations backK914145 DSD-91(TM) Disinfector for Flexible EndoscopesMedivators, Inc. · 1994 · Recalled
  2. 2 generations backK040883 Evotech Integrated Endoscope Disinfection SystemAdvanced Sterilization Products · 2005 · Recalled
  3. 1 generation backK061899 Evotech Integrated Endoscope Disinfection SystemAdvanced Sterilization Products · 2006 · Recalled
  4. This deviceK082988 Medivators Advantage Plus Endoscope Reprocessing System, Rapicide PA…Medivators Reprocessing Systems · 2009
  5. Descendant · cites it directly · 2 of its ownK152394 Rapicide PA High-Level Disinfectant Test StripsMedivators, Inc. · 2016
  6. Descendant · cites it directly · 1 of its ownK192228 CS Medical TD 200 disinfector with TD-12 high-level disinfectant, CS…Cs Medical, LLC · 2020
  7. Descendant · 2 generations on · 0 of its ownK251035 DISINTEK™ PA Test StripsSerim Research · 2025
  8. Descendant · 2 generations on · 0 of its ownK251048 Rapicide PA High-Level Disinfection Test Strip Model Ref # = ML02-0118STERIS Corporation · 2025
  9. Descendant · 2 generations on · 0 of its ownK233049 TD 200® Automated TEE Probe Disinfector with TD-12® High-Level…Cs Medical, LLC · 2023

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2010: 12016: 12017: 12020: 12023: 12025: 2
2010high 2 (2025) · low 0 (2011)2026

Most influential descendants

Showing 5 of 7 · See all
  • K152394Rapicide PA High-Level Disinfectant Test StripsMedivators, Inc. · 2016 · cites it directly2descendants
  • K192228CS Medical TD 200 disinfector with TD-12 high-level disinfectant, CS Medical TD-12…Cs Medical, LLC · 2020 · cites it directly1descendant
  • K251035DISINTEK™ PA Test StripsSerim Research · 2025 · 2 generations on0descendants
  • K251048Rapicide PA High-Level Disinfection Test Strip Model Ref # = ML02-0118STERIS Corporation · 2025 · 2 generations on0descendants
  • K233049TD 200® Automated TEE Probe Disinfector with TD-12® High-Level Disinfectant with…Cs Medical, LLC · 2023 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K914145DSD-91(TM) Disinfector for Flexible EndoscopesMedivators, Inc.19941
−2K040883Evotech Integrated Endoscope Disinfection SystemAdvanced Sterilization Products20053
−1K061899Evotech Integrated Endoscope Disinfection SystemAdvanced Sterilization Products20062
0 (this device)K082988Medivators Advantage Plus Endoscope Reprocessing System, Rapicide PA High…Medivators Reprocessing Systems2009—
+1K152394Rapicide PA High-Level Disinfectant Test StripsMedivators, Inc.2016—
+1K192228CS Medical TD 200 disinfector with TD-12 high-level disinfectant, CS Medical…Cs Medical, LLC2020—
+2K251035DISINTEK™ PA Test StripsSerim Research2025—
+2K251048Rapicide PA High-Level Disinfection Test Strip Model Ref # = ML02-0118STERIS Corporation2025—
+2K233049TD 200® Automated TEE Probe Disinfector with TD-12® High-Level Disinfectant…Cs Medical, LLC2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199710
199910
200510
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201010
201610
201710
202010
202310
202520
FIG. 1 — Lineage of K082988, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FEB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FEB device

  1. #1 K243433 enspire 3000 Cleaning and Liquid Chemical…score 10.8 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K063876Modular Disinfection System (MDS) for Endoscope Reprocessing
Summary page 3
…on Within the United States The Advantage Plus is equivalent in function and indications to the Medivators MDS Endoscope Reprocessing System (K063876) and the Advanced Sterilization Products' EvoTech TM Integrated Endoscope Disinfection System (K061899). All of…
Medivators Reprocessing Systems2007—
*K061899Evotech Integrated Endoscope Disinfection System
Summary page 3
…Medivators MDS Endoscope Reprocessing System (K063876) and the Advanced Sterilization Products' EvoTech TM Integrated Endoscope Disinfection System (K061899). All of the machines have the same indications for use and the same methods of providing disinfection. Rapici…
Advanced Sterilization Products20062
*K041984Acecide High-Level Disinfectant and Sterilant
Summary page 3
…f providing disinfection. Rapicide PA High Level Disinfectant is equivalent in indications and usage to Acecide High Level Disinfectant and Sterilant (K041984) and Peract 20 Liquid Sterilant (K960513). All of the products are peracetic acid based germicides labeled for…
Best Sanitizers, Inc.2005—
*K960513Peract 20 Liquid Sterilant
Summary page 3
…gh Level Disinfectant is equivalent in indications and usage to Acecide High Level Disinfectant and Sterilant (K041984) and Peract 20 Liquid Sterilant (K960513). All of the products are peracetic acid based germicides labeled for providing high level disinfection to flex…
Minntech Corp.1997—

What descends from it? (7)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K152394Rapicide PA High-Level Disinfectant Test StripsMedivators, Inc.20162
K192228CS Medical TD 200 disinfector with TD-12 high-level disinfectant, CS Medical TD-12…Cs Medical, LLC20201
K172677Advantage Plus Pass-ThruMedivators, Inc.20170
K092387Medivators DSD Edge Endoscope Reprocessing SystemMinntech Corp.20100

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 21, 2026Class IIZ-0156-2027Process change controlIf the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which…
Sep 21, 2026Class IIZ-0154-2027Process change controlIf the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which…
Sep 21, 2026Class IIZ-0155-2027Process change controlIf the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which…
Nov 13, 2023Class IIZ-0617-2024Process controlIdentification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately result in postponement…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 2.3 years on median.
  • 3 of the 7 ancestors have a recall of any class.
  • Of its descendants, K152394 (Medivators, Inc., 2016) has gone furthest, with 2 descendants of its own.

How does its category compare?

All FEB devices

FEB Accessories, Cleaning, For Endoscope: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · FEB

2in 2024

2007: 22008: 12009: 12010: 22011: 32012: 12013: 12014: 12016: 22017: 22018: 12019: 22020: 22022: 12023: 12024: 22026 (partial year): 1
2007high 3 (2011) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FEB

111in 2011

2011: 111 days median
2007high 111 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FEB

15so far in 2026, the most since 2007

2008: 32010: 12016: 22020: 12021: 12023: 12024: 12025: 32026 (partial year): 15
2007high 15 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FEB

356in 2025

2007: 392008: 1522009: 202010: 502011: 322012: 162013: 272014: 482015: 922016: 1032017: 692018: 882019: 622020: 4262021: 5752022: 6072023: 9982024: 8842025: 356
2007high 998 (2023) · low 16 (2012)2025

MAUDE reports naming product code FEB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—039
20081—3152
20091—020
20102—150
20113111032
20121—016
20131—027
20141—048
20150—092
20162—2103
20172—069
20181—088
20192—062
20202—1426
20210—1575
20221—0607
20231—1998
20242—1884
20250—3356
20261—15—

What is it indicated for?

Medivators Advantage Plus Endoscope Reprocessing System tests, washes, disinfects and rinses flexible endoscopes, such as fiberoptic and video endoscopes between patient uses. The Advantage Plus system is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes. Manual cleaning of endoscopes is required prior to placement in the Advantage Plus system. Rapicide PA High Level Disinfectant is intended for use with the…

Read the full statement (141 words)

Medivators Advantage Plus Endoscope Reprocessing System tests, washes, disinfects and rinses flexible endoscopes, such as fiberoptic and video endoscopes between patient uses. The Advantage Plus system is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes. Manual cleaning of endoscopes is required prior to placement in the Advantage Plus system.

Rapicide PA High Level Disinfectant is intended for use with the Advantage Plus Endoscope Reprocessing System to provide high level disinfection of endoscopes when used according to the directions for use. Rapicide PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is above the minimum recommended concentration (850ppm peracetic acid); this ensures that the disinfectant was above MRC during the entire disinfection cycle. Rapicide PA should be used under the following contact conditions:

High Level Disinfection — 5 minutes — 30(°C) — 850ppm

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K082988: Medivators Advantage Plus Endoscope Reprocessing System, Rapicide PA High Level Disinfectant.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K082988

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026