Philips NM3 Monitor, Model 7900

K091459510(k) SpecialCleared Jul 28, 200971-day reviewClass II21 CFR 868.1850Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Respironics Novametrix, LLC · BZK Spirometer, Monitoring (W/Wo Alarm)

Cleared on Jul 28, 2009 as substantially equivalent to K030886 (Respironics Novametrix, Inc., 2003) after a 71-day review. Its family tree reaches back 6 generations to 1997. 5 later devices descend from it. It has been recalled 4 times.

Decided
Jul 28, 2009
Review
71 days
Class
Class II
Product code
BZK
Recalls
4
Predicates
1

Its longest line of descent runs back 6 generations to K962603 (1997). 10 ancestors in all, the oldest K913516 (1991). 5 later devices descend from it, 1 citing it directly.

  1. Root · 6 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 5 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 4 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 3 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. 2 generations backK993979 MARSPO2, Model 2001Novametrix Medical Systems, Inc. · 2000 · Recalled
  6. 1 generation backK030886 Nico with Mars, Model 7300Respironics Novametrix, Inc. · 2003 · Recalled
  7. This deviceK091459 Philips NM3 Monitor, Model 7900Respironics Novametrix, LLC · 2009
  8. Descendant · cites it directly · 4 of its ownK103578 Philips NM3 Respiratory Monitor with VentassistRespironics Novametrix, LLC · 2011
  9. Descendant · 2 generations on · 2 of its ownK222897 Enlight 2100Timpel S.A. · 2023
  10. Descendant · 3 generations on · 0 of its ownK250464 Enlight 2100 (TPL-E2103-0)Timpel S.A. · 2025
  11. Descendant · 3 generations on · 0 of its ownK243765 LuMon (TM) SystemSentec AG · 2025
  12. Descendant · 2 generations on · 0 of its ownK192584 BEACON Caresystem Model 00002144Mermaid Care A/S · 2019

Most influential descendants

Showing 5 of 5 · See all
  • K103578Philips NM3 Respiratory Monitor with VentassistRespironics Novametrix, LLC · 2011 · cites it directly4descendants
  • K222897Enlight 2100Timpel S.A. · 2023 · 2 generations on2descendants
  • K250464Enlight 2100 (TPL-E2103-0)Timpel S.A. · 2025 · 3 generations on0descendants
  • K243765LuMon (TM) SystemSentec AG · 2025 · 3 generations on0descendants
  • K192584BEACON Caresystem Model 00002144Mermaid Care A/S · 2019 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−5K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−4K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−3K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
−2K993979MARSPO2, Model 2001Novametrix Medical Systems, Inc.20001
−1K030886Nico with Mars, Model 7300Respironics Novametrix, Inc.20038
0 (this device)K091459Philips NM3 Monitor, Model 7900Respironics Novametrix, LLC2009—
+1K103578Philips NM3 Respiratory Monitor with VentassistRespironics Novametrix, LLC2011—
+2K222897Enlight 2100Timpel S.A.2023—
+3K250464Enlight 2100 (TPL-E2103-0)Timpel S.A.2025—
+3K243765LuMon (TM) SystemSentec AG2025—
+2K192584BEACON Caresystem Model 00002144Mermaid Care A/S2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199610
199810
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201910
202310
202520
FIG. 1 — Lineage of K091459, 1997–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in BZK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BZK device

  1. #1 K260226 ElintMAX Respiratory Monitor (ElintMAX Monitor…score 19.5 · 26 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K030886Nico with Mars, Model 7300
Summary page 1
…alve). Classification Name: Class II,21 CFR 868.1850, 868.1400, 868.5675 and 870.2700 Predicate Device: NICO with MARS monitor, Model 7300 [510(k) K030886] Description of Device: The Philips NM3 monitor is intended for non-invasive monitoring of the inspired and expired airflow and…
Respironics Novametrix, Inc.20038

Also cleared against K030886: 5 other devices →

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103578Philips NM3 Respiratory Monitor with VentassistRespironics Novametrix, LLC20114

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 29, 2011Not classifiedZ-3288-2011Manufacturing material removalPressure lines maybe occluded and display incorrect tidal volumes.
Jul 29, 2011Not classifiedZ-3287-2011Manufacturing material removalPressure lines maybe occluded and display incorrect tidal volumes.
Jan 4, 2011Not classifiedZ-2168-2011Software designSoftware: two displayed respiratory parameters are incorrect due to errors in the system software: MValv (alveolar minute ventilation) and Vt/kg (tidal volume/patient weight). In both instances, the displayed parameters are higher than…
Jan 4, 2011Not classifiedZ-2167-2011Software designSoftware: two displayed respiratory parameters are incorrect due to errors in the system software: MValv (alveolar minute ventilation) and Vt/kg (tidal volume/patient weight). In both instances, the displayed parameters are higher than…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 6 companies; a new generation was cleared every 11 months on median.
  • 5 of the 10 ancestors have a recall of any class.
  • Of its descendants, K103578 (Respironics Novametrix, LLC, 2011) has gone furthest, with 4 descendants of its own.
  • 5 other devices also cite its predicate K030886.

How does its category compare?

All BZK devices

BZK Spirometer, Monitoring (W/Wo Alarm): FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · BZK

2in 2019

2008: 12009: 32011: 12012: 12013: 12017: 12018: 22019: 22026 (partial year): 1
2007high 3 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BZK

71in 2009

2009: 71 days median
2007high 71 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BZK

2in 2011

2011: 2
2007high 2 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · BZK

2in 2020

2009: 12012: 12013: 732014: 312015: 122016: 12017: 12018: 12020: 2
2007high 73 (2013) · low 0 (2007)2025

MAUDE reports naming product code BZK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20081—00
200937101
20100—00
20111—20
20121—01
20131—073
20140—031
20150—012
20160—01
20171—01
20182—01
20192—00
20200—02
20210—00
20220—00
20230—00
20240—00
20250—00
20261—0—

What is it indicated for?

The intended use of the Philips NM3 monitor, Model 7900 is to provide: cardiac output monitoring via the method of partial rebreathing in adult patients receiving mechanical ventilation during general anesthesia and in the intensive care unit (ICU). spirometric, and carbon dioxide monitoring in neonatal, pediatric and adult patients during general anesthesia and in the intensive care unit (ICU) and the emergency department (ED). Separate combination…

Read the full statement (191 words)

The intended use of the Philips NM3 monitor, Model 7900 is to provide:

cardiac output monitoring via the method of partial rebreathing in adult patients receiving mechanical ventilation during general anesthesia and in the intensive care unit (ICU).

spirometric, and carbon dioxide monitoring in neonatal, pediatric and adult patients during general anesthesia and in the intensive care unit (ICU) and the emergency department (ED). Separate combination C02/flow sensors are provided for adult, pediatric and neonatal use.

continuous, non-invasive monitoring of functional arterial oxygen saturation and pulse rate in neonatal, pediatric and adult patients during both no motion and motion conditions and for patients who are well or poorly perfused during general anesthesia and in the intensive care unit (ICU) and the emergency department (ED).

The NM3 monitor Model 7900 and its sensors are intended to be used by trained operators when spirometric, capnographic, pulse oximetry, or cardiac output monitoring is indicated in the judgement of a physician.

The use of the NM3 monitor Model 7900 for cardiac output monitoring is contraindicated in patients in which a small rise (3-5 mmHg) in their arterial partial pressure of C02 level cannot be tolerated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091459: Philips NM3 Monitor, Model 7900.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K091459

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026