510(k) Report

510(k) · Cleared

Cleared

Philips NM3 Respiratory Monitor with Ventassist

K103578510(k) TraditionalCleared Jun 2, 2011178-day reviewClass II21 CFR 868.1850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Respironics Novametrix, LLC · BZK Spirometer, Monitoring (W/Wo Alarm)

Cleared on Jun 2, 2011 as substantially equivalent to K091459 (Respironics Novametrix, LLC, 2009) and 3 other predicates after a 178-day review. Its family tree reaches back 8 generations to 1991. 4 of the 45 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it.

Decided
Jun 2, 2011
Review
178 days
Class
Class II
Product code
BZK
Recalls
None
Predicates
4

Its longest line of descent runs back 8 generations to K902859 (1991). 40 ancestors in all, the oldest K900696 (1990). 4 later devices descend from it, 2 citing it directly. 4 of the 45 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor1990: 1 other ancestor2003: 1 other ancestor2002: 2 other ancestors2001: 4 other ancestors1999: 3 other ancestors1998: 5 other ancestors1997: 4 other ancestors2000: 4 other ancestors, 1 with a Class I recall11996: 2 other ancestors1995: 1 other ancestor1994: 1 other ancestor1991: 2 other ancestors1992: 1 other ancestor2023: 1 descendant cleared2019: 1 descendant cleared2025: 2 descendants clearedK902859 · Servo Ventilator 300 · Siemens Life Support Systems · 1991 · recalledK961687 · EVITA 4 · Drager, Inc. · 1996K970460 · Nellcor Puritan Bennett, 840 Ventilator System (Model 840) · Puritan Bennett Corp. · 1998 · Class I recallK984535 · Puritan-Bennett 840 Ventilatory System with Bilevel Option · Puritan Bennett Corp. · 1998 · Class I recallK022679 · Raphael · Hamilton Medical AG · 2003 · recalledK040574 · Galileo Gold · Hamilton Medical, Inc. · 2004K051263 · Evitaxl with Option Smartcare · Drager Medical AG & Co. Kgaa · 2005 · recalledK053388 · Model 840 Ventilator with Proportional Assist Ventilation Plus Software Option · Puritan Bennett Corp. · 2006 · Class I recall1990200020102020
  1. Root · 8 generations backK902859 Servo Ventilator 300Siemens Life Support Systems · 1991 · Recalled
  2. 7 generations backK961687 EVITA 4Drager, Inc. · 1996
  3. 6 generations backK970460 Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · Recalled (Class I)
  4. 5 generations backK984535 Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · Recalled (Class I)
  5. 4 generations backK022679 RaphaelHamilton Medical AG · 2003 · Recalled
  6. 3 generations backK040574 Galileo GoldHamilton Medical, Inc. · 2004
  7. 2 generations backK051263 Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa · 2005 · Recalled
  8. 1 generation backK053388 Model 840 Ventilator with Proportional Assist Ventilation Plus…Puritan Bennett Corp. · 2006 · Recalled (Class I)
  9. This deviceK103578 Philips NM3 Respiratory Monitor with VentassistRespironics Novametrix, LLC · 2011
  10. Descendant · cites it directly · 2 of its ownK222897 Enlight 2100Timpel S.A. · 2023
  11. Descendant · 2 generations on · 0 of its ownK250464 Enlight 2100 (TPL-E2103-0)Timpel S.A. · 2025
  12. Descendant · 2 generations on · 0 of its ownK243765 LuMon (TM) SystemSentec AG · 2025
  13. Descendant · cites it directly · 0 of its ownK192584 BEACON Caresystem Model 00002144Mermaid Care A/S · 2019

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K222897Enlight 2100Timpel S.A. · 2023 · cites it directly2descendants
  • K250464Enlight 2100 (TPL-E2103-0)Timpel S.A. · 2025 · 2 generations on0descendants
  • K243765LuMon (TM) SystemSentec AG · 2025 · 2 generations on0descendants
  • K192584BEACON Caresystem Model 00002144Mermaid Care A/S · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K902859Servo Ventilator 300Siemens Life Support Systems19913
−7K961687EVITA 4Drager, Inc.1996—
−6K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp.1998Class I
−5K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp.1998Class I
−4K022679RaphaelHamilton Medical AG20033
−3K040574Galileo GoldHamilton Medical, Inc.2004—
−2K051263Evitaxl with Option SmartcareDrager Medical AG & Co. Kgaa20051
−1K053388Model 840 Ventilator with Proportional Assist Ventilation Plus Software OptionPuritan Bennett Corp.2006Class I
0 (this device)K103578Philips NM3 Respiratory Monitor with VentassistRespironics Novametrix, LLC2011—
+1K222897Enlight 2100Timpel S.A.2023—
+2K250464Enlight 2100 (TPL-E2103-0)Timpel S.A.2025—
+2K243765LuMon (TM) SystemSentec AG2025—
+1K192584BEACON Caresystem Model 00002144Mermaid Care A/S2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199120
199210
199410
199510
199620
199740
199850
199930
200041
200140
200220
200310
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202310
202520
FIG. 1 — Lineage of K103578, 1991–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BZK device

  1. #1 K260226 ElintMAX Respiratory Monitor (ElintMAX Monitor…score 19.5 · 26 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K091459Philips NM3 Monitor, Model 7900
Summary page 1
…and promotional literature for predicate devices. The predicate devices are as follows: 1. Philips NM3 Respiratory Profile Monitor K091459 2. Bicore CP-100 Cardiopulmonary Monitor K900696 3. Puritan Bennett Model 840 Ventilator…
Respironics Novametrix, LLC20094
*K900696Biore CP-100 Cardiopulmonary Monitor
Summary page 1
…s are as follows: 1. Philips NM3 Respiratory Profile Monitor K091459 2. Bicore CP-100 Cardiopulmonary Monitor K900696 3. Puritan Bennett Model 840 Ventilator with PAV+ K053388 4. Dr~ger EvitaXL with Option SmartCare…
Bicore Monitoring Systems1990—
*K053388Model 840 Ventilator with Proportional Assist Ventilation Plus Software Option
Summary page 1
…K091459 2. Bicore CP-100 Cardiopulmonary Monitor K900696 3. Puritan Bennett Model 840 Ventilator with PAV+ K053388 4. Dr~ger EvitaXL with Option SmartCare K051263 e. Device Description The NMS monitor with…
Puritan Bennett Corp.2006Class I
*K051263Evitaxl with Option Smartcare
Summary page 1
…K900696 3. Puritan Bennett Model 840 Ventilator with PAV+ K053388 4. Dr~ger EvitaXL with Option SmartCare K051263 e. Device Description The NMS monitor with VentAssist is intended for non-invasive monitoring of the inspired and expi…
Drager Medical AG & Co. Kgaa20051

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K222897Enlight 2100Timpel S.A.20232
K192584BEACON Caresystem Model 00002144Mermaid Care A/S20190

Has anything in its family been recalled? (4 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K053388Model 840 Ventilator with Proportional Assist Ventilation Plus Software OptionPuritan Bennett Corp. · 2006 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K001646Puritan-Bennett 840 Ventilator System with Neomode OptionPuritan Bennett Corp. · 2000 · 3 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K984535Puritan-Bennett 840 Ventilatory System with Bilevel OptionPuritan Bennett Corp. · 1998 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K970460Nellcor Puritan Bennett, 840 Ventilator System (Model 840)Puritan Bennett Corp. · 1998 · 4 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 20 years and passes through 7 companies; a new generation was cleared every 1.4 years on median.
  • 17 of the 40 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K222897 (Timpel S.A., 2023) has gone furthest, with 2 descendants of its own.

How does its category compare?

All BZK devices

BZK Spirometer, Monitoring (W/Wo Alarm): FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · BZK

2in 2019

2008: 12009: 32011: 12012: 12013: 12017: 12018: 22019: 22026 (partial year): 1
2007high 3 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BZK

71in 2009

2009: 71 days median
2007high 71 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BZK

2in 2011

2011: 2
2007high 2 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · BZK

2in 2020

2009: 12012: 12013: 732014: 312015: 122016: 12017: 12018: 12020: 2
2007high 73 (2013) · low 0 (2007)2025

MAUDE reports naming product code BZK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20081—00
200937101
20100—00
20111—20
20121—01
20131—073
20140—031
20150—012
20160—01
20171—01
20182—01
20192—00
20200—02
20210—00
20220—00
20230—00
20240—00
20250—00
20261—0—

What is it indicated for?

The intended use of the Philips INMV3 Respiratory Profile Monitor, Model 7900, is to provide: * cardiac output monitoring via the method of partial rebreathing in adult patients receiving mechanical ventilation during general anesthesia and in the intensive care unit (ICU). * spirometric, and carbon dioxide monitoring in neonatal, pediatric and adult patients during general anesthesia and in the intensive care unit (ICU) and the emergency department…

Read the full statement (217 words)

The intended use of the Philips INMV3 Respiratory Profile Monitor, Model 7900, is to provide:

* cardiac output monitoring via the method of partial rebreathing in adult patients receiving mechanical ventilation during general anesthesia and in the intensive care unit (ICU).

* spirometric, and carbon dioxide monitoring in neonatal, pediatric and adult patients during general anesthesia and in the intensive care unit (ICU) and the emergency department (ED). Separate combination CO2Iflow sensors are provided for adult, pediatric and neonatal use.

* continuous, non-invasive monitoring of functional arterial oxygen saturation and pulse rate in neonatal, pediatric and adult patients during both no motion and motion conditions and for patients who are welt or poorly perfused during general anesthesia and in the intensive care unit (ICU) and the emergency department (ED).

The intended use of the VentAssist software option isto provide:

* non-invasive monitoring of work of breathing per minute in adult patients receiving pressure support mechanical ventilation.

* on-demand advice in mechanically ventilated adult patients as prescribed by the caregiver regarding (a) modifications to the current pressure support settings in order to assess the work of breathing and breathing pattern and (b)modifications to ventilation in order to maintain end-tidal C02 in a range determined by the physician. The patients should be hemodynamically stable and must be breathing spontaneously.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K103578: Philips NM3 Respiratory Monitor with Ventassist.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K103578

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026