510(k) · Cleared
Recalled · Class II · 2026Aptus Foot System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Medartis AG · HRS Plate, Fixation, Bone
Cleared on Jul 13, 2009 after a 55-day review. The FDA summary names no predicate by number. 51 later devices descend from it. It has been recalled 2 times.
- Decided
- Jul 13, 2009
- Review
- 55 days
- Class
- Class II
- Product code
- HRS
- Recalls
- 2
- Predicates
- 0
51 later devices descend from it, 9 citing it directly.
- This deviceK091479 Aptus Foot SystemMedartis AG · 2009
- Descendant · cites it directly · 25 of its ownK110908 Aptus Foot 3.5 SystemMedartis AG · 2011
- Descendant · cites it directly · 22 of its ownK142581 APTUS Foot SystemMedartis AG · 2015
- Descendant · cites it directly · 19 of its ownK162674 BioPro Foot Plating SystemsBiopro, Inc. · 2017
- Descendant · cites it directly · 11 of its ownK193633 APTUS® Ankle Trauma System 2.8/3.5Medartis AG · 2020
- Descendant · 2 generations on · 11 of its ownK193222 Tyber Medical Anatomical Plating SystemTyber Medical, LLC · 2020
Showing 5 of 51 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 51 · See all- K110908Aptus Foot 3.5 SystemMedartis AG · 2011 · cites it directly25descendants
- K142581APTUS Foot SystemMedartis AG · 2015 · cites it directly22descendants
- K162674BioPro Foot Plating SystemsBiopro, Inc. · 2017 · cites it directly19descendants
- K193633APTUS® Ankle Trauma System 2.8/3.5Medartis AG · 2020 · cites it directly11descendants
- K193222Tyber Medical Anatomical Plating SystemTyber Medical, LLC · 2020 · 2 generations on11descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K091479 | Aptus Foot System | Medartis AG | 2009 | — |
| +1 | K110908 | Aptus Foot 3.5 System | Medartis AG | 2011 | — |
| +1 | K142581 | APTUS Foot System | Medartis AG | 2015 | — |
| +1 | K162674 | BioPro Foot Plating Systems | Biopro, Inc. | 2017 | — |
| +1 | K193633 | APTUS® Ankle Trauma System 2.8/3.5 | Medartis AG | 2020 | — |
| +2 | K193222 | Tyber Medical Anatomical Plating System | Tyber Medical, LLC | 2020 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2011 | 2 | 0 |
| 2015 | 2 | 0 |
| 2017 | 4 | 0 |
| 2018 | 1 | 0 |
| 2019 | 3 | 0 |
| 2020 | 6 | 0 |
| 2021 | 2 | 0 |
| 2023 | 10 | 0 |
| 2024 | 8 | 0 |
| 2025 | 5 | 0 |
| 2026 | 8 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in HRS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new HRS device
- #1 K222465 Tyber Medical Anatomical Plating Systemscore 27.5 · 16 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
The FDA summary names no predicate by number.
What descends from it? (51)
See all 519 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K110908 | Aptus Foot 3.5 System | 2011 | 25 |
| K142581 | APTUS Foot System | 2015 | 22 |
| K162674 | BioPro Foot Plating Systems | 2017 | 19 |
| K193633 | APTUS® Ankle Trauma System 2.8/3.5 | 2020 | 11 |
| K193639 | APTUS® Foot 2.8-3.5 System | 2020 | 9 |
| K112560 | Aptus (R) Distal Humerus System | 2011 | 8 |
| K161894 | SurgTech Trauma System | 2017 | 3 |
| K243469 | Signex | 2025 | 0 |
| K232105 | APTUS Foot System 2.8-3.5 | 2024 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jan 9, 2026 | Class II | Z-1550-2026 | Labeling mix-ups | Mix up between specified lots of 2.5mm and 2.8mm outer diameter screws. |
| Apr 12, 2024 | Class II | Z-1821-2024 | Process control | Screw is 16mm long instead of 22mm long. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K110908 (Medartis AG, 2011) has gone furthest, with 25 descendants of its own.
How does its category compare?
All HRS devicesHRS Plate, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.
FDA decisions per year · HRS
54in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · HRS
90in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · HRS
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · HRS
2,306in 2025
MAUDE reports naming product code HRS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 35 | 63 | 2 | 339 |
| 2008 | 35 | 87 | 11 | 311 |
| 2009 | 47 | 85 | 3 | 435 |
| 2010 | 49 | 100 | 16 | 491 |
| 2011 | 43 | 88 | 5 | 773 |
| 2012 | 59 | 106 | 4 | 1,663 |
| 2013 | 56 | 89 | 11 | 2,643 |
| 2014 | 76 | 57 | 0 | 2,413 |
| 2015 | 56 | 97 | 7 | 3,095 |
| 2016 | 44 | 109 | 5 | 2,276 |
| 2017 | 57 | 122 | 3 | 2,220 |
| 2018 | 40 | 116 | 2 | 2,663 |
| 2019 | 46 | 89 | 25 | 3,359 |
| 2020 | 48 | 88 | 1 | 2,790 |
| 2021 | 40 | 87 | 9 | 2,442 |
| 2022 | 47 | 144 | 3 | 1,700 |
| 2023 | 40 | 120 | 1 | 2,049 |
| 2024 | 54 | 96 | 0 | 1,849 |
| 2025 | 54 | 90 | 1 | 2,306 |
| 2026 | 51 | 82 | 3 | — |
What is it indicated for?
The APTUS® Foot System is intended for use in small bones, in particular in fractures, osteotomies and arthrodesis of the tarsals, metatarsals and phalanges.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K091479 · printed Nov 4, 2026
Cite this page
510(k) Report. “K091479: Aptus Foot System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K091479
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026