510(k) Report

510(k) · Cleared

Recalled · 2010

Vitrea Endovascular Stent Planning Software

K091498510(k) TraditionalCleared Jun 5, 200916-day reviewClass II21 CFR 892.2050Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Vital Images, Inc. · LLZ System, Image Processing, Radiological

Cleared on Jun 5, 2009 as substantially equivalent to K071331 (Vital Images, Inc., 2007) and 3 other predicates after a 16-day review. Its family tree reaches back 10 generations to 1991. 1 later device descends from it. It has been recalled once.

Decided
Jun 5, 2009
Review
16 days
Class
Class II
Product code
LLZ
Recalls
1
Predicates
4

Its longest line of descent runs back 10 generations to K913770 (1991). 46 ancestors in all, the oldest K871859 (1987). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2004: 3 other ancestors2000: 3 other ancestors2005: 1 other ancestor2003: 2 other ancestors1999: 3 other ancestors1996: 4 other ancestors2002: 3 other ancestors2001: 4 other ancestors1997: 3 other ancestors1995: 2 other ancestors1993: 2 other ancestors1987: 1 other ancestor1998: 2 other ancestors1994: 2 other ancestors2014: 1 descendant clearedK913770 · Advantage V.S. CT/MR Rad Diagnostic Workstation · General Electric Co. · 1991K960613 · GE Advantage Windows Review Workstation · GE Medical Systems · 1996K020483 · Advantage Workstation 4.1 · GE Medical Systems · 2002 · recalledK023003 · Imagechecker-CT Workstation · R2 Technology, Inc. · 2002×2: 2 generations cleared close together, K032748 (2003), K040305 (2004) K043333 · VITREA2, Version 3.7 Medical Image Processing System · Vital Images, Inc. · 2004 · recalledK052632 · VITREA2 Version 3.8 Medical Image Processing Software · Vital Images, Inc. · 2005 · recalledK061624 · Vitrea, Version 3.9 · Vital Images, Inc. · 2006 · recalledK071331 · Vitrea Version 4.0 · Vital Images, Inc. · 2007 · recalled1990200020102020
  1. Root · 10 generations backK913770 Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co. · 1991
  2. 6 generations between
  3. 3 generations backK052632 VITREA2 Version 3.8 Medical Image Processing SoftwareVital Images, Inc. · 2005 · Recalled
  4. 2 generations backK061624 Vitrea, Version 3.9Vital Images, Inc. · 2006 · Recalled
  5. 1 generation backK071331 Vitrea Version 4.0Vital Images, Inc. · 2007 · Recalled
  6. This deviceK091498 Vitrea Endovascular Stent Planning SoftwareVital Images, Inc. · 2009
  7. Descendant · cites it directly · 0 of its ownK143079 Endovascular Stent Planning SoftwareVital Images, Inc. · 2014
Show all 10 generations
  1. Root · 10 generations backK913770 Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co. · 1991
  2. 9 generations backK960613 GE Advantage Windows Review WorkstationGE Medical Systems · 1996
  3. 8 generations backK020483 Advantage Workstation 4.1GE Medical Systems · 2002 · Recalled
  4. 7 generations backK023003 Imagechecker-CT WorkstationR2 Technology, Inc. · 2002
  5. 6 generations backK032748 VITREA2, Version 3.4 Medical Image Processing SoftwareVital Images, Inc. · 2003 · Recalled
  6. 5 generations backK040305 VITREA2 Version 3.5Vital Images, Inc. · 2004
  7. 4 generations backK043333 VITREA2, Version 3.7 Medical Image Processing SystemVital Images, Inc. · 2004 · Recalled
  8. 3 generations backK052632 VITREA2 Version 3.8 Medical Image Processing SoftwareVital Images, Inc. · 2005 · Recalled
  9. 2 generations backK061624 Vitrea, Version 3.9Vital Images, Inc. · 2006 · Recalled
  10. 1 generation backK071331 Vitrea Version 4.0Vital Images, Inc. · 2007 · Recalled
  11. This deviceK091498 Vitrea Endovascular Stent Planning SoftwareVital Images, Inc. · 2009

Most influential descendants

Showing 1 of 1 · See all
  • K143079Endovascular Stent Planning SoftwareVital Images, Inc. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K913770Advantage V.S. CT/MR Rad Diagnostic WorkstationGeneral Electric Co.1991—
−9K960613GE Advantage Windows Review WorkstationGE Medical Systems1996—
−8K020483Advantage Workstation 4.1GE Medical Systems20022
−7K023003Imagechecker-CT WorkstationR2 Technology, Inc.2002—
−6K032748VITREA2, Version 3.4 Medical Image Processing SoftwareVital Images, Inc.20032
−5K040305VITREA2 Version 3.5Vital Images, Inc.2004—
−4K043333VITREA2, Version 3.7 Medical Image Processing SystemVital Images, Inc.20041
−3K052632VITREA2 Version 3.8 Medical Image Processing SoftwareVital Images, Inc.20052
−2K061624Vitrea, Version 3.9Vital Images, Inc.20061
−1K071331Vitrea Version 4.0Vital Images, Inc.20071
0 (this device)K091498Vitrea Endovascular Stent Planning SoftwareVital Images, Inc.2009—
+1K143079Endovascular Stent Planning SoftwareVital Images, Inc.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199320
199420
199520
199640
199730
199820
199930
200030
200140
200230
200320
200430
200510
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
FIG. 1 — Lineage of K091498, 1991–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K071331Vitrea Version 4.0
Summary page 1
…ate devices. Predicate Device Manufacturer FDA 510(k) Vitreag, Version 4.0 Medical Image Vital Images, Inc K071331 Processing Software Vitrea2, version 2.1 Vital Images, Inc K002519 Preview(® Treatment Planni…
Vital Images, Inc.20071
*K002519Vitrea 2, Version 2.1
Summary page 1
…eag, Version 4.0 Medical Image Vital Images, Inc K071331 Processing Software Vitrea2, version 2.1 Vital Images, Inc K002519 Preview(® Treatment Planning MEDICAL MEDIA SYSTEMS, K040852 Software INC GE A…
Vital Images, Inc.20001
*K040852Preview Treatment Planning Software
Summary page 1
…essing Software Vitrea2, version 2.1 Vital Images, Inc K002519 Preview(® Treatment Planning MEDICAL MEDIA SYSTEMS, K040852 Software INC GE AVA 11/Vessel Xpress GE HealthCare K060779 (AVA…
Medical Media Systems, Inc.2004—
*K060779Advance Vessel Analysis II
Summary page 1
…ng MEDICAL MEDIA SYSTEMS, K040852 Software INC GE AVA 11/Vessel Xpress GE HealthCare K060779 (AVA II/Vessel Xpress) Waukesha, WI 2.1. Device Description The Endovascular Stent Planning So…
GE Healthcare2006—

Also cleared against K071331: 14 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K143079Endovascular Stent Planning SoftwareVital Images, Inc.20140

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 23, 2010Not classifiedZ-1600-2010Software designVital images, Inc is issuing a recall notice on their medical device software in the use of CT Lung and CT Colon in Vitrea 5.1, Vitrea fX 3.0 or Vitre Enterprise Suite 1.2. because incorrect measurements may result. Questions or comments…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 4 companies; a new generation was cleared every 11 months on median.
  • 9 of the 46 ancestors have a recall of any class.
  • Of its descendants, K143079 (Vital Images, Inc., 2014) has gone furthest, with 0 descendants of its own.
  • 14 other devices also cite its predicate K071331.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

Vitrea® is a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea is not meant for primary image interpretation in mammography. Vitrea has the following additional indication: The Endovascular Stent Planning Software application is intended for use with CT (computed tomography) images to assist medical…

Read the full statement (126 words)

Vitrea® is a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea is not meant for primary image interpretation in mammography. Vitrea has the following additional indication:

The Endovascular Stent Planning Software application is intended for use with CT (computed tomography) images to assist medical professionals in the analysis, treatment and follow-up of aortic vascular disorders that may require a stent procedure. The software provides 3D segmentation of the aorta and initializes stent measurements based on a template provided by the stent manufacturers. The user can review the 2D and 3D images, verify and correct the results of the segmentation and initialization, and generate a report with the stent measurements.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091498: Vitrea Endovascular Stent Planning Software.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K091498

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026