510(k) Report

510(k) · Cleared

Recalled · Class I · 2019

Teleflex Isis HVT Tracheal Tube, Cuffed, with Subglottic Secretion Suction Port

K091761510(k) AbbreviatedCleared Oct 29, 2009135-day reviewClass II21 CFR 868.5730Recalls 15 (15 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Teleflexmedical, Inc. · BTR Tube, Tracheal (W/Wo Connector)

Cleared on Oct 29, 2009 as substantially equivalent to K822082 (Sheridan Catheter Corp., 1982) and 2 other predicates after a 135-day review. Its family tree reaches back 5 generations to 1990. 2 of the 15 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 15 times, including 15 Class I recalls.

Decided
Oct 29, 2009
Review
135 days
Class
Class II
Product code
BTR
Recalls
15 (15 Class I)
Predicates
3

Its longest line of descent runs back 5 generations to K903730 (1990). 12 ancestors in all, the oldest K811033 (1981). 2 later devices descend from it, 1 citing it directly. 2 of the 15 devices in this family, counting this one, have a Class I recall.

  1. Root · 5 generations backK903730 D.I.C. Tracheostomy Tubes & Laryngectomy TubesConcord/Portex · 1990
  2. 4 generations backK960429 Per-Fit Percutaneous Tracheostomy Kit W/Portex Tracheostmy Tube &…Smiths Industries Medical Systems, Inc. · 1996
  3. 3 generations backK980466 Per-Fit Percutaneous Dilational Tracheostomy Kit with Specialty…Sims Portex, Inc. · 1998 · Recalled
  4. 2 generations backK030570 Blue Line Ultra Suctionaid Tracheostomy TubePortex, Ltd. · 2003
  5. 1 generation backK081086 Portex Blue Line Sacett Suction Above the Cuff Tracheal Tube, Size…Smiths Medical Asd, Inc. · 2008
  6. This deviceK091761 Teleflex Isis HVT Tracheal Tube, Cuffed, with Subglottic Secretion…Teleflexmedical, Inc. · 2009
  7. Descendant · cites it directly · 1 of its ownK141939 Telefelx Isis HVT Tracheal Tube, Cuffed with Subglottic Secretion…Teleflexmedical, Inc. · 2014
  8. Descendant · 2 generations on · 0 of its ownK190274 Endotracheal TubeTianjin Medis Medical Device Co., Ltd. · 2019

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K141939Telefelx Isis HVT Tracheal Tube, Cuffed with Subglottic Secretion Suction Port; with…Teleflexmedical, Inc. · 2014 · cites it directly1descendant
  • K190274Endotracheal TubeTianjin Medis Medical Device Co., Ltd. · 2019 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K903730D.I.C. Tracheostomy Tubes & Laryngectomy TubesConcord/Portex1990—
−4K960429Per-Fit Percutaneous Tracheostomy Kit W/Portex Tracheostmy Tube & Disp. Inner…Smiths Industries Medical Systems, Inc.1996—
−3K980466Per-Fit Percutaneous Dilational Tracheostomy Kit with Specialty Tracheostomy…Sims Portex, Inc.19981
−2K030570Blue Line Ultra Suctionaid Tracheostomy TubePortex, Ltd.2003—
−1K081086Portex Blue Line Sacett Suction Above the Cuff Tracheal Tube, Size 6.0 and…Smiths Medical Asd, Inc.2008—
0 (this device)K091761Teleflex Isis HVT Tracheal Tube, Cuffed, with Subglottic Secretion Suction PortTeleflexmedical, Inc.2009—
+1K141939Telefelx Isis HVT Tracheal Tube, Cuffed with Subglottic Secretion Suction…Teleflexmedical, Inc.2014—
+2K190274Endotracheal TubeTianjin Medis Medical Device Co., Ltd.2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198211
199120
199310
199410
199710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201910
FIG. 1 — Lineage of K091761, 1990–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BTR device

  1. #1 K251794 Tenax® Laser Resistant Endotracheal Tubescore 12.5 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K822082Sheridan/HVT,CF, Uncuffed
Summary page 2
…evices: Predicte~evce ~ ~uatre 1()N. ar Cerd Sheridan HVT, Cuffed, Teleflex Medical, Inc. K822082 08/10/1982 Uncuffed Tracheal Tube __________ _____ Hi-Lo Evac and E…
Sheridan Catheter Corp.1982Class I
*K965132HI-LO Evac and Evac II Endotracheal Tubes
Summary page 2
…10/1982 Uncuffed Tracheal Tube __________ _____ Hi-Lo Evac and Evac II Mallinckrodt Medical K965 132 03/20/1997 Endotracheal Tubes___________ Portex Blue Line Sacett Suction Smiths Medical ASD, K081086…
Mallinckrodt Medical1997—
*K081086Portex Blue Line Sacett Suction Above the Cuff Tracheal Tube, Size 6.0 and 6.5, 7.0 and…
Summary page 2
…Mallinckrodt Medical K965 132 03/20/1997 Endotracheal Tubes___________ Portex Blue Line Sacett Suction Smiths Medical ASD, K081086 08/13/2008 Above the Cuff Tracheal Tube Inc. ______ _____ I. Summary of Test…
Smiths Medical Asd, Inc.2008—

Also cleared against K822082: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K141939Telefelx Isis HVT Tracheal Tube, Cuffed with Subglottic Secretion Suction Port; with…Teleflexmedical, Inc.20141

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 7, 2019Class IZ-1956-2019Device DesignCenturion is conducting a sub-recall for Teleflex Sheridan branded Endotracheal Tracheal Tubes due to increased incidence of Sheridan connector becoming disconnected from Endotracheal Tube.
Jan 6, 2014Class IZ-0950-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0953-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0954-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0957-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0958-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0959-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0961-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0963-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0955-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0956-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0960-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0962-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0951-2014OtherThe tracheal tube can kink during patient use.
Jan 6, 2014Class IZ-0952-2014OtherThe tracheal tube can kink during patient use.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K822082Sheridan/HVT,CF, UncuffedSheridan Catheter Corp. · 1982 · Direct predicate · ◆ in the 1982 ancestors bar of FIG. 117 Class I recalls

What does this lineage say?

  • The line of descent spans 19 years and passes through 6 companies; a new generation was cleared every 4.9 years on median.
  • 2 of the 12 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K141939 (Teleflexmedical, Inc., 2014) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K822082.

How does its category compare?

All BTR devices

BTR Tube, Tracheal (W/Wo Connector): FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · BTR

7in 2025

2007: 12008: 72009: 72010: 32011: 22012: 52013: 12014: 32015: 32016: 12017: 22018: 52019: 52020: 32021: 12022: 12023: 12024: 12025: 72026 (partial year): 2
2007high 7 (2008) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BTR

258in 2025

2008: 61 days median2009: 135 days median2010: 179 days median2012: 261 days median2014: 308 days median2015: 262 days median2018: 242 days median2019: 211 days median2020: 246 days median2025: 258 days median
2007high 308 (2014) · low 61 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BTR

3in 2022

2007: 22009: 212014: 20, 19 Class I2016: 12020: 1, 1 Class I2021: 22022: 32026 (partial year): 1
2007high 21 (2009) · low 0 (2008)2026

Class I (20 since 2007) Class II/III not classified

Adverse event reports per year · BTR

1,799in 2025

2007: 1362008: 1692009: 1552010: 3352011: 4222012: 5112013: 8072014: 9852015: 1,0162016: 8362017: 8122018: 9772019: 1,3792020: 1,0702021: 1,3292022: 1,1322023: 1,3302024: 1,6782025: 1,799
2007high 1,799 (2025) · low 136 (2007)2025

MAUDE reports naming product code BTR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—2136
20087610169
2009713521155
201031790335
20112—0422
201252610511
20131—0807
2014330820 (19)985
2015326201,016
20161—1836
20172—0812
201852420977
2019521101,379
202032461 (1)1,070
20211—21,329
20221—31,132
20231—01,330
20241—01,678
2025725801,799
20262—1—

What is it indicated for?

The TELEFLEX ISIS™ HVT™ Tracheal Tube, Cuffed with Subglottic Secretion Suction Port is indicated for airway management by oral intubation during mechanical ventilation and anesthesia including the capability to aid in the removal of subglottic secretions.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K091761: Teleflex Isis HVT Tracheal Tube, Cuffed, with Subglottic Secretion Suction Port.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K091761

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026