510(k) Report

510(k) · Cleared

Cleared

Global Shoulder Steptech Anchor Peg Glenoid

K092122510(k) SpecialCleared Jul 27, 200913-day reviewClass II21 CFR 888.3660

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

DePuy Orthopaedics, Inc. · KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

Cleared on Jul 27, 2009 as substantially equivalent to K052472 (DePuy Orthopaedics, Inc., 2005) after a 13-day review. Its family tree reaches back 4 generations to 1994. 75 later devices descend from it.

Decided
Jul 27, 2009
Review
13 days
Class
Class II
Product code
KWS
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K943462 (1994). 9 ancestors in all, the oldest K914000 (1991). 75 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1998: 1 other ancestor1991: 1 other ancestor1997: 1 other ancestor1996: 1 other ancestor1995: 1 other ancestor2021: 6 descendants cleared2013: 3 descendants cleared2015: 6 descendants cleared2014: 1 descendant cleared2025: 11 descendants cleared2018: 3 descendants cleared2016: 6 descendants cleared2017: 4 descendants cleared2019: 6 descendants cleared2023: 8 descendants cleared2026: 9 descendants cleared2020: 3 descendants cleared2022: 3 descendants cleared2024: 6 descendants clearedK943462 · Darwin Knee System · Johnson & Johnson Professionals, Inc. · 1994 · recalled×2: 2 generations cleared close together, K033272 (2004), K040166 (2004) K052472 · Depuy Global Shoulder Crosslink Glenoid · DePuy Orthopaedics, Inc. · 20051990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 3 other ancestors 1 generation forward: 3 descendants2 generations forward: 7 descendants3 generations forward: 12 descendants4 generations forward: 14 descendants5 generations forward: 10 descendants6 generations forward: 8 descendants7 generations forward: 15 descendants8 generations forward: 5 descendants9 generations forward: 1 descendant K052472 · Depuy Global Shoulder Crosslink Glenoid · DePuy Orthopaedics, Inc. · 2005K040166 · Depuy Sigma XLK Tibial Inserts · DePuy Orthopaedics, Inc. · 2004K033272 · Depuy Sigma Tibial Inserts · DePuy Orthopaedics, Inc. · 2004 · recalledK943462 · Darwin Knee System · Johnson & Johnson Professionals, Inc. · 1994 · recalled0+5
  1. Root · 4 generations backK943462 Darwin Knee SystemJohnson & Johnson Professionals, Inc. · 1994 · Recalled
  2. 3 generations backK033272 Depuy Sigma Tibial InsertsDePuy Orthopaedics, Inc. · 2004 · Recalled
  3. 2 generations backK040166 Depuy Sigma XLK Tibial InsertsDePuy Orthopaedics, Inc. · 2004
  4. 1 generation backK052472 Depuy Global Shoulder Crosslink GlenoidDePuy Orthopaedics, Inc. · 2005
  5. This deviceK092122 Global Shoulder Steptech Anchor Peg GlenoidDePuy Orthopaedics, Inc. · 2009
  6. Descendant · cites it directly · 56 of its ownK123122 Glenoid Intelligent Reusable Instrument System (Glenoid Iris)Cleveland Clinic · 2013
  7. Descendant · 2 generations on · 39 of its ownK131129 Cas Psi ShoulderZimmer Cas · 2013
  8. Descendant · 3 generations on · 31 of its ownK143374 BLUEPRINT Patient Specific InstrumentationTornier S.A.S. · 2015
  9. Descendant · 4 generations on · 25 of its ownK160555 BLUEPRINT Patient Specific InstrumentationTornier S.A.S. · 2016
  10. Descendant · 5 generations on · 23 of its ownK162800 BLUEPRINT Patient Specific InstrumentationTornier S.A.S. · 2017

Showing 5 of 75 descendants. See all descendants

Descendants cleared per year

2013: 32014: 12015: 62016: 62017: 42018: 32019: 62020: 32021: 62022: 32023: 82024: 62025: 112026: 9
2013high 11 (2025) · low 1 (2014)2026

Most influential descendants

Showing 5 of 75 · See all
  • K123122Glenoid Intelligent Reusable Instrument System (Glenoid Iris)Cleveland Clinic · 2013 · cites it directly56descendants
  • K131129Cas Psi ShoulderZimmer Cas · 2013 · 2 generations on39descendants
  • K143374BLUEPRINT Patient Specific InstrumentationTornier S.A.S. · 2015 · 3 generations on31descendants
  • K160555BLUEPRINT Patient Specific InstrumentationTornier S.A.S. · 2016 · 4 generations on25descendants
  • K162800BLUEPRINT Patient Specific InstrumentationTornier S.A.S. · 2017 · 5 generations on23descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K943462Darwin Knee SystemJohnson & Johnson Professionals, Inc.19948
−3K033272Depuy Sigma Tibial InsertsDePuy Orthopaedics, Inc.200451
−2K040166Depuy Sigma XLK Tibial InsertsDePuy Orthopaedics, Inc.2004—
−1K052472Depuy Global Shoulder Crosslink GlenoidDePuy Orthopaedics, Inc.2005—
0 (this device)K092122Global Shoulder Steptech Anchor Peg GlenoidDePuy Orthopaedics, Inc.2009—
+1K123122Glenoid Intelligent Reusable Instrument System (Glenoid Iris)Cleveland Clinic2013—
+2K131129Cas Psi ShoulderZimmer Cas2013—
+3K143374BLUEPRINT Patient Specific InstrumentationTornier S.A.S.2015—
+4K160555BLUEPRINT Patient Specific InstrumentationTornier S.A.S.2016—
+5K162800BLUEPRINT Patient Specific InstrumentationTornier S.A.S.2017—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199510
199610
199710
199810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201330
201410
201560
201660
201740
201830
201960
202030
202160
202230
202380
202460
2025110
202690
FIG. 1 — Lineage of K092122, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWS device

  1. #1 K220847 AETOS Shoulder Systemscore 21.1 · 16 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052472Depuy Global Shoulder Crosslink Glenoid
Summary page 1
…KWS SUBSTANTIALLY EQUIVALENT DePuy Global Shoulder Crosslinked Glenoid, DEVICE: K052472 (October 11, 2005) DEVICE DESCRIPTION: The DePuy Global Shoulder StepTech Anchor Peg Glenoid is designed for use as the glenoid…
DePuy Orthopaedics, Inc.2005—

Also cleared against K052472: 6 other devices →

What descends from it? (75)

See all 75

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123122Glenoid Intelligent Reusable Instrument System (Glenoid Iris)Cleveland Clinic201356
K150583Aequalis PerFORM+ Shoulder SystemTornier, Inc.201516
K203230Global Shoulder and Delta CTA systemsDepuy Lnc.20210

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 2 companies; a new generation was cleared every 3.8 years on median.
  • 4 of the 9 ancestors have a recall of any class.
  • Of its descendants, K123122 (Cleveland Clinic, 2013) has gone furthest, with 56 descendants of its own.
  • 6 other devices also cite its predicate K052472.

How does its category compare?

All KWS devices

KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · KWS

3in 2025

2007: 22008: 42009: 42010: 72011: 112012: 72013: 122014: 72015: 82016: 102017: 72018: 42019: 92020: 32021: 52022: 52023: 42024: 22025: 32026 (partial year): 2
2007high 12 (2013) · low 2 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWS

97in 2025

2008: 37 days median2009: 108 days median2010: 33 days median2011: 90 days median2012: 86 days median2013: 120 days median2014: 89 days median2015: 129 days median2016: 114 days median2017: 120 days median2018: 152 days median2019: 106 days median2020: 98 days median2021: 151 days median2022: 158 days median2023: 250 days median2025: 97 days median
2007high 250 (2023) · low 33 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KWS

2in 2023

2007: 212008: 292010: 102011: 112012: 102014: 12016: 12017: 42018: 12019: 22020: 52023: 22026 (partial year): 1
2007high 29 (2008) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWS

2,281in 2025

2007: 1582008: 2382009: 2942010: 3932011: 4862012: 5422013: 8182014: 1,0662015: 1,1702016: 9702017: 1,6812018: 1,5452019: 1,7282020: 1,5672021: 1,3582022: 1,2682023: 1,7502024: 2,1222025: 2,281
2007high 2,281 (2025) · low 158 (2007)2025

MAUDE reports naming product code KWS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—21158
200843729238
200941080294
201073310393
2011119011486
201278610542
2013121200818
201478911,066
2015812901,170
2016101141970
2017712041,681
2018415211,545
2019910621,728
202039851,567
2021515101,358
2022515801,268
2023425021,750
20242—02,122
202539702,281
20262—1—

What is it indicated for?

The StepTech Anchor Peg Glenoid is intended for use in total shoulder replacement surgery for patients suffering from:

1. A severely painful and/or disabled joint resulting from osteoarthritis, traumatic arthritis or rheumatoid arthritis.

2. Fracture-dislocations of the proximal humerus where the articular surface is severely communited, separated from its blood supply or where the surgeon's experience indicates that alternative methods of treatment are unsatisfactory.

3. Other difficult clinical problems where shoulder arthrodesis or resection arthroplasty are not acceptable (e.g. revision of a failed primary component).

Glenoid components are intended for cemented use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K092122: Global Shoulder Steptech Anchor Peg Glenoid.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K092122

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026