Single Use Electrosurgical Knife Series, Model KD-610L, KD-611L, KD-620LR, KD-630L, KF-640L

K092309510(k) TraditionalCleared Oct 30, 200992-day reviewClass II21 CFR 876.4300Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Medical Systems Corporation · KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories)

Cleared on Oct 30, 2009 after a 92-day review. The FDA summary names its predicate without a number. 15 later devices descend from it. It has been recalled 9 times.

Decided
Oct 30, 2009
Review
92 days
Class
Class II
Product code
KNS
Recalls
9
Predicates
0

15 later devices descend from it, 7 citing it directly.

Descendants, by year cleared 2025: 1 descendant cleared2020: 4 descendants cleared2017: 3 descendants cleared2016: 3 descendants cleared2022: 2 descendants cleared2019: 1 descendant cleared2021: 1 descendant cleared20052010201520202025
  1. This deviceK092309 Single Use Electrosurgical Knife Series, Model KD-610L, KD-611L…Olympus Medical Systems Corporation · 2009
  2. Descendant · cites it directly · 7 of its ownK171158 Single Use Electrosurgical knifeOlympus Medical Systems Corp. · 2017 · Recalled
  3. Descendant · cites it directly · 3 of its ownK151474 Fujifilm Diathermic Slitter (FlushKnife), Diathermic Slitter…Fujifilm Medical System U.S.A., Inc. · 2016
  4. Descendant · 2 generations on · 2 of its ownK161186 Fujifilm Diathermic Slitter (FlushKnife), Fujifilm Diathermic…Fujifilm Medical Systems U.S.A, Inc. · 2016
  5. Descendant · 2 generations on · 1 of its ownK193601 Single Use Electrosurgical KnifeMicro-Tech (Nanjing) Co., Ltd. · 2020
  6. Descendant · 2 generations on · 1 of its ownK182408 Single Use Electrosurgical Knife KD-625 , Single Use Electrosurgical…Olympus Medical Systems Corp. · 2019 · Recalled

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2016: 32017: 32019: 12020: 42021: 12022: 22025: 1
2016high 4 (2020) · low 0 (2018)2026

Most influential descendants

Showing 5 of 15 · See all
  • K171158Single Use Electrosurgical knifeOlympus Medical Systems Corp. · 2017 · cites it directly7descendants
  • K151474Fujifilm Diathermic Slitter (FlushKnife), Diathermic Slitter (ClutchCutter)Fujifilm Medical System U.S.A., Inc. · 2016 · cites it directly3descendants
  • K161186Fujifilm Diathermic Slitter (FlushKnife), Fujifilm Diathermic Slitter (ClutchCutter)Fujifilm Medical Systems U.S.A, Inc. · 2016 · 2 generations on2descendants
  • K193601Single Use Electrosurgical KnifeMicro-Tech (Nanjing) Co., Ltd. · 2020 · 2 generations on1descendant
  • K182408Single Use Electrosurgical Knife KD-625 , Single Use Electrosurgical Knife KD-645Olympus Medical Systems Corp. · 2019 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K092309Single Use Electrosurgical Knife Series, Model KD-610L, KD-611L, KD-620LR…Olympus Medical Systems Corporation2009—
+1K171158Single Use Electrosurgical knifeOlympus Medical Systems Corp.20172
+1K151474Fujifilm Diathermic Slitter (FlushKnife), Diathermic Slitter (ClutchCutter)Fujifilm Medical System U.S.A., Inc.2016—
+2K161186Fujifilm Diathermic Slitter (FlushKnife), Fujifilm Diathermic Slitter…Fujifilm Medical Systems U.S.A, Inc.2016—
+2K193601Single Use Electrosurgical KnifeMicro-Tech (Nanjing) Co., Ltd.2020—
+2K182408Single Use Electrosurgical Knife KD-625 , Single Use Electrosurgical Knife…Olympus Medical Systems Corp.20193

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201630
201730
201910
202040
202110
202220
202510
FIG. 1 — Lineage of K092309, 2009–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KNS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KNS device

  1. #1 K241271 Aqua Medical RF Vapor Ablation Systemscore 17.2 · 20 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—KD-IL-1
Named without a submission number
—

What descends from it? (15)

See all 15

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171158Single Use Electrosurgical knifeOlympus Medical Systems Corp.20177
K151474Fujifilm Diathermic Slitter (FlushKnife), Diathermic Slitter (ClutchCutter)Fujifilm Medical System U.S.A., Inc.20163
K171983Creo Medical Electrosurgical System including Speedboat RS2 Surgical AccessoryCreo Medical, Ltd.20171
K250351Single Use Electrosurgical Knife (KD-612L, KD-612U); Single Use Electrosurgical Knife…Olympus Medical Systems Corporation20250
K212999Multi-Functional Electrosurgical KnifeAnrei Medical (Hangzhou) Co., Ltd.20220
K192342Electrosurgical System (Electrosurgical Generator with Bipolar polypectomy snare…Hangzhou AGS MedTech Co., Ltd.20200
K152771SB Knife Jr type, SB Knife Short type, SB Knife Standard type, SB Knife Jr type - LongSumitomo Bakelite Co., Ltd.20160

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 17, 2024Class IIZ-1222-2026Device DesignA deterioration of the cutting knife, including overheating and burning, can contribute to tip breakage during use.
Jan 4, 2021Class IIZ-0980-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0977-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0976-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0974-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0975-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0978-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0979-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Jan 4, 2021Class IIZ-0981-2021Under Investigation by firmInspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K171158 (Olympus Medical Systems Corp., 2017) has gone furthest, with 7 descendants of its own.

How does its category compare?

All KNS devices

KNS Unit, Electrosurgical, Endoscopic (With Or Without Accessories): FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · KNS

11in 2025

2007: 12008: 32009: 32010: 42011: 12012: 22013: 22014: 12015: 32016: 32017: 32018: 52019: 72020: 42021: 42022: 32023: 92024: 32025: 112026 (partial year): 2
2007high 11 (2025) · low 1 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KNS

171in 2025

2008: 61 days median2009: 212 days median2010: 168 days median2015: 252 days median2016: 30 days median2017: 48 days median2018: 62 days median2019: 66 days median2020: 212 days median2021: 139 days median2022: 182 days median2023: 83 days median2024: 198 days median2025: 171 days median
2007high 252 (2015) · low 30 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KNS

14in 2024

2007: 32008: 52009: 22011: 22016: 62017: 12021: 412024: 142026 (partial year): 4
2007high 41 (2021) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KNS

1,504in 2025

2007: 1642008: 5872009: 1,0272010: 9082011: 8602012: 8482013: 4492014: 4052015: 4062016: 4572017: 4462018: 5632019: 6682020: 9062021: 8142022: 9132023: 9762024: 1,5332025: 1,504
2007high 1,533 (2024) · low 164 (2007)2025

MAUDE reports naming product code KNS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—3164
20083615587
2009321221,027
201041680908
20111—2860
20122—0848
20132—0449
20141—0405
201532520406
20163306457
20173481446
20185620563
20197660668
202042120906
2021413941814
202231820913
20239830976
20243198141,533
20251117101,504
20262—4—

What is it indicated for?

Singqle Use Electrosurgical Knife KD-610OL/KD-61 IL/KD-620LR/KD-63OL/KD-640L

This instrument has been designed to be used with Olympus endoscopes and electrosurgical units to cut tissue within the digestive tract using high-frequency current.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K092309: Single Use Electrosurgical Knife Series, Model KD-610L, KD-611L, KD-620LR, KD-630L, KF-640L.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K092309

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026