510(k) Report

510(k) · Cleared

Recalled · Class I · 2013

Guardian II Hemostasis Valve

K092711510(k) SpecialCleared Oct 1, 200928-day reviewClass II21 CFR 870.4290Recalls 2 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zerusa Limited · DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass

Cleared on Oct 1, 2009 as substantially equivalent to K052381 (Zerusa Limited, 2005) and 1 other predicate after a 28-day review. Its family tree reaches back 5 generations to 1985. 3 of the 14 devices in its family, counting itself, have a Class I recall. 8 later devices descend from it. It has been recalled 2 times, including a Class I recall.

Decided
Oct 1, 2009
Review
28 days
Class
Class II
Product code
DTL
Recalls
2 (1 Class I)
Predicates
2

Its longest line of descent runs back 5 generations to K854261 (1985). 8 later devices descend from it, 3 citing it directly. 3 of the 14 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2012: 1 descendant cleared, 1 with a Class I recall12010: 2 descendants cleared, 1 with a Class I recall12014: 1 descendant cleared2017: 2 descendants cleared2022: 1 descendant cleared2020: 1 descendant clearedK854261 · Advanced Cardiovascular Systems Hemostatic Valves · Advanced Cardiovascular Systems, Inc. · 1985 · recalledK991102 · Copilot Bleedback Control Valve, Model 1003331 · Guidant Corp. · 1999 · recalledK042060 · Easypass US Y-Connector Hemostatic Valve · Millimed A/S · 2004K052381 · Guardian Hemostasis Valve with Guidewire Introducer · Zerusa Limited · 2005K073620 · Guardian Hemostasis Valve · Zerusa Limited · 20081990200020102020
  1. Root · 5 generations backK854261 Advanced Cardiovascular Systems Hemostatic ValvesAdvanced Cardiovascular Systems, Inc. · 1985 · Recalled
  2. 4 generations backK991102 Copilot Bleedback Control Valve, Model 1003331Guidant Corp. · 1999 · Recalled
  3. 3 generations backK042060 Easypass US Y-Connector Hemostatic ValveMillimed A/S · 2004
  4. 2 generations backK052381 Guardian Hemostasis Valve with Guidewire IntroducerZerusa Limited · 2005
  5. 1 generation backK073620 Guardian Hemostasis ValveZerusa Limited · 2008
  6. This deviceK092711 Guardian II Hemostasis ValveZerusa Limited · 2009
  7. Descendant · cites it directly · 4 of its ownK100093 Vsi Torque DeviceVascular Solutions, Inc. · 2010
  8. Descendant · 2 generations on · 3 of its ownK141054 Distal Access Torque DeviceDistal Access, LLC · 2014
  9. Descendant · 3 generations on · 1 of its ownK170684 Firebow Wire Torque Assist DeviceVesatek, LLC · 2017
  10. Descendant · cites it directly · 1 of its ownK122301 Guardian II Hemostasis Valve Guardian II NC Hemostasis ValveVascular Solutions Zerusa, Ltd. · 2012 · Recalled (Class I)
  11. Descendant · 3 generations on · 0 of its ownK221917 Stent Positioning Assistance System (SPAS)Seven Sons, Ltd. · 2022

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2010: 22012: 12014: 12017: 22020: 12022: 1
2010high 2 (2010) · low 0 (2011)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 8 · See all
  • K100093Vsi Torque DeviceVascular Solutions, Inc. · 2010 · cites it directly4descendants
  • K141054Distal Access Torque DeviceDistal Access, LLC · 2014 · 2 generations on3descendants
  • K170684Firebow Wire Torque Assist DeviceVesatek, LLC · 2017 · 3 generations on1descendant
  • K122301Guardian II Hemostasis Valve Guardian II NC Hemostasis ValveVascular Solutions Zerusa, Ltd. · 2012 · cites it directly1descendant
  • K221917Stent Positioning Assistance System (SPAS)Seven Sons, Ltd. · 2022 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K854261Advanced Cardiovascular Systems Hemostatic ValvesAdvanced Cardiovascular Systems, Inc.19852
−4K991102Copilot Bleedback Control Valve, Model 1003331Guidant Corp.19991
−3K042060Easypass US Y-Connector Hemostatic ValveMillimed A/S2004—
−2K052381Guardian Hemostasis Valve with Guidewire IntroducerZerusa Limited2005—
−1K073620Guardian Hemostasis ValveZerusa Limited2008—
0 (this device)K092711Guardian II Hemostasis ValveZerusa Limited2009—
+1K100093Vsi Torque DeviceVascular Solutions, Inc.2010—
+2K141054Distal Access Torque DeviceDistal Access, LLC2014—
+3K170684Firebow Wire Torque Assist DeviceVesatek, LLC2017—
+1K122301Guardian II Hemostasis Valve Guardian II NC Hemostasis ValveVascular Solutions Zerusa, Ltd.2012Class I
+3K221917Stent Positioning Assistance System (SPAS)Seven Sons, Ltd.2022—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201021
201211
201410
201720
202010
202210
FIG. 1 — Lineage of K092711, 1985–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052381Guardian Hemostasis Valve with Guidewire Introducer
Summary page 1
…The predicate device used to determnine substantial equivalence for this device was the Zerusa Limited's currently marketed GuardianTm Hemostasis Valve (HK052381 and itK073620). 4. DEVICE DESCRIPTION: The Zerusa GuardianTM II Hemostasis Valve is designed to be used a…
Zerusa Limited2005—
*K073620Guardian Hemostasis Valve
Summary page 1
…e device used to determnine substantial equivalence for this device was the Zerusa Limited's currently marketed GuardianTm Hemostasis Valve (HK052381 and itK073620). 4. DEVICE DESCRIPTION: The Zerusa GuardianTM II Hemostasis Valve is designed to be used as a conduit…
Zerusa Limited2008—

Also cleared against K052381: 1 other device →

What descends from it? (8)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K100093Vsi Torque DeviceVascular Solutions, Inc.20104
K122301Guardian II Hemostasis Valve Guardian II NC Hemostasis Valve Class IVascular Solutions Zerusa, Ltd.20121
K101113Guardian II NC Hemostasis Valve Class IZerusa Limited20100

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 27, 2013Class IZ-0955-2013Vendor change controlRisk that air may be introduced into the device which may lead to an air embolism.
Aug 30, 2012Class IIZ-2400-2012Process controlVascular Solutions, Inc., became aware of a potential problem involving pouches missing a seal from Guardian II (Model 8210) and Guardian II NC (Model FH102) Hemostasis Valve which poses a risk of product contamination.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K122301Guardian II Hemostasis Valve Guardian II NC Hemostasis ValveVascular Solutions Zerusa, Ltd. · 2012 · Cites it directly · ◆ in the 2012 descendants bar of FIG. 12 Class I recalls
  • K101113Guardian II NC Hemostasis ValveZerusa Limited · 2010 · Cites it directly · ◆ in the 2010 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 24 years and passes through 4 companies; a new generation was cleared every 2.3 years on median.
  • 2 of the 5 ancestors have a recall of any class.
  • Of its descendants, K100093 (Vascular Solutions, Inc., 2010) has gone furthest, with 4 descendants of its own.
  • 1 other device also cites its predicate K052381.

How does its category compare?

All DTL devices

DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2009, this device’s year, is highlighted.

FDA decisions per year · DTL

1in 2015

2007: 12008: 52009: 22010: 42012: 82013: 12014: 22015: 1
2007high 8 (2012) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTL

36in 2012

2008: 85 days median2010: 115 days median2012: 36 days median
2007high 115 (2010) · low 36 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTL

1in 2025

2008: 12010: 32011: 12012: 22013: 1, 1 Class I2014: 12017: 22022: 22025: 12026 (partial year): 3, 3 Class I
2007high 3 (2010) · low 0 (2007)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · DTL

110in 2025

2007: 372008: 162009: 622010: 1482011: 1602012: 532013: 672014: 802015: 892016: 672017: 1062018: 842019: 762020: 532021: 552022: 482023: 462024: 4522025: 110
2007high 452 (2024) · low 16 (2008)2025

MAUDE reports naming product code DTL, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—037
2008585116
20092—062
201041153148
20110—1160
2012836253
20131—1 (1)67
20142—180
20151—089
20160—067
20170—2106
20180—084
20190—076
20200—053
20210—055
20220—248
20230—046
20240—0452
20250—1110
20260—3 (3)—

What is it indicated for?

The Guardian TM II HV is intended to maintain hemostasis during the use of diagnostic/interventional devices. The device is indicated for maintaining a seal around diagnostic/interventional devices with outside diameters up to 8.OF (2.67 mm or 0.105") during diagnostic/interventional procedures.

The guidewire introducer is included to facilitate the guidewire's passage through the Guardian TM II HV.

The Torque Device is included to manipulate the steering of the guidewire within the vascular regions.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K092711: Guardian II Hemostasis Valve.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K092711

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026