510(k) · Cleared
Recalled · Class II · 2023Olympus CHF Type Y0003
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Olympus Medical Systems Corporation · FBN Choledochoscope And Accessories, Flexible/Rigid
Cleared on Mar 10, 2010 as substantially equivalent to K081456 (Olympus Medical Systems Corporation, 2008) after a 117-day review. Its family tree reaches back 2 generations to 1985. It has been recalled once.
- Decided
- Mar 10, 2010
- Review
- 117 days
- Class
- Class II
- Product code
- FBN
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 2 generations to K843084 (1985). No later device cites it as a predicate yet.
View FIG. 1 as a table
Line of descent and top descendants
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FBN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FBN device
- #1 K254153 Single-use Video Scope; PB Digital Controllerscore 14.4 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K081456 | Olympus CHF Type VSummary page 2…Aizuwakamatsu-shi, Fukushima, Japan 965-8520 Establishment Registration No.: 9610595 * 5 10(k) No. K081456 4 Device Description The CIIF-Y0003 choledochoscope is a flexible video endoscope used for endoscopic diagnosis and… | 2008 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 25, 2023 | Class II | Z-0345-2024 | Under Investigation by firm | Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 25 years and passes through 2 companies; a new generation was cleared every 23.8 years on median.
- 1 of the 2 ancestors has a recall of any class.
- 1 other device also cites its predicate K081456.
How does its category compare?
All FBN devicesFBN Choledochoscope And Accessories, Flexible/Rigid: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.
FDA decisions per year · FBN
3in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FBN
150in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FBN
1in 2024
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · FBN
118in 2025
MAUDE reports naming product code FBN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 1 |
| 2008 | 3 | 64 | 0 | 5 |
| 2009 | 0 | — | 0 | 4 |
| 2010 | 1 | — | 0 | 2 |
| 2011 | 0 | — | 0 | 2 |
| 2012 | 1 | — | 0 | 0 |
| 2013 | 0 | — | 0 | 2 |
| 2014 | 0 | — | 0 | 2 |
| 2015 | 1 | — | 0 | 2 |
| 2016 | 1 | — | 0 | 168 |
| 2017 | 0 | — | 0 | 224 |
| 2018 | 1 | — | 0 | 639 |
| 2019 | 1 | — | 0 | 105 |
| 2020 | 1 | — | 1 | 173 |
| 2021 | 0 | — | 0 | 158 |
| 2022 | 0 | — | 0 | 283 |
| 2023 | 4 | 215 | 1 | 105 |
| 2024 | 2 | — | 1 (1) | 107 |
| 2025 | 3 | 150 | 0 | 118 |
| 2026 | 3 | 175 | 1 | — |
What is it indicated for?
This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bile duct through a percutaneous route and/or intraoperatively.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K093507 · printed Nov 6, 2026
Cite this page
510(k) Report. “K093507: Olympus CHF Type Y0003.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K093507
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026