510(k) Report

510(k) · Cleared

Recalled · Class I · 2022

Infinity Acute Care System Workstation Critical Care

K093633510(k) TraditionalCleared Mar 30, 2010127-day reviewClass II21 CFR 868.5895Recalls 6 (4 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Draegerwerk AG & CO Kgaa · CBK Ventilator, Continuous, Facility Use

Cleared on Mar 30, 2010 after a 127-day review. The FDA summary names no predicate by number. 3 of the 13 devices in its family, counting itself, have a Class I recall. 12 later devices descend from it. It has been recalled 6 times, including 4 Class I recalls.

Decided
Mar 30, 2010
Review
127 days
Class
Class II
Product code
CBK
Recalls
6 (4 Class I)
Predicates
0

12 later devices descend from it, 5 citing it directly. 3 of the 13 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2014: 2 descendants cleared, 1 with a Class I recall12023: 3 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared2018: 1 descendant cleared2020: 1 descendant cleared2026: 3 descendants cleared2024: 1 descendant cleared2010201520202025
  1. This deviceK093633 Infinity Acute Care System Workstation Critical CareDraegerwerk AG & CO Kgaa · 2010
  2. Descendant · cites it directly · 4 of its ownK181695 Nihon Kohden NKV-550 Series Ventilator SystemNihon Kohden Orangemed, Inc. · 2018 · Recalled
  3. Descendant · 2 generations on · 3 of its ownK192307 Nihon Kohden NKV-550 Series Ventilator SystemNihon Kohden Orangemed, Inc. · 2020 · Recalled
  4. Descendant · cites it directly · 2 of its ownK133886 Perseus A500Draeger Medical GmbH · 2014 · Recalled (Class I)
  5. Descendant · 2 generations on · 1 of its ownK230931 AtlanDrägerwerk AG Co. Kgaa · 2023 · Recalled (Class I)
  6. Descendant · 3 generations on · 1 of its ownK222644 Nihon Kohden NKV-440 Ventilator SystemNihon Kohden Orangemed, Inc. · 2023

Showing 5 of 12 descendants. See all descendants

Descendants cleared per year

2014: 22018: 12019: 12020: 12023: 32024: 12026: 3
2014high 3 (2023) · low 0 (2015)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 12 · See all
  • K181695Nihon Kohden NKV-550 Series Ventilator SystemNihon Kohden Orangemed, Inc. · 2018 · cites it directly4descendants
  • K192307Nihon Kohden NKV-550 Series Ventilator SystemNihon Kohden Orangemed, Inc. · 2020 · 2 generations on3descendants
  • K133886Perseus A500Draeger Medical GmbH · 2014 · cites it directly2descendants
  • K230931AtlanDrägerwerk AG Co. Kgaa · 2023 · 2 generations on1descendant
  • K222644Nihon Kohden NKV-440 Ventilator SystemNihon Kohden Orangemed, Inc. · 2023 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K093633Infinity Acute Care System Workstation Critical CareDraegerwerk AG & CO Kgaa2010—
+1K181695Nihon Kohden NKV-550 Series Ventilator SystemNihon Kohden Orangemed, Inc.20181
+2K192307Nihon Kohden NKV-550 Series Ventilator SystemNihon Kohden Orangemed, Inc.20201
+1K133886Perseus A500Draeger Medical GmbH2014Class I
+2K230931AtlanDrägerwerk AG Co. Kgaa2023Class I
+3K222644Nihon Kohden NKV-440 Ventilator SystemNihon Kohden Orangemed, Inc.2023—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201421
201810
201910
202010
202331
202410
202630
FIG. 1 — Lineage of K093633, 2010–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it? (12)

See all 12

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K181695Nihon Kohden NKV-550 Series Ventilator SystemNihon Kohden Orangemed, Inc.20184
K133886Perseus A500 Class IDraeger Medical GmbH20142
K222024Evita V800, Evita V600Draegerwerk AG & CO Kgaa20231
K121886Savina 300 Savina 300, Country VariantsDraeger Medical GmbH20141
K180779Savina 300Dragerwerk AG & CO Kgaa20190

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 14, 2022Class IIZ-0870-2022Nonconforming Material/ComponentCybersecurity Improvement Action-not equipped against potential cyber security threats
Sep 10, 2021Class IIZ-0108-2022Software designSoftware 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds
Feb 4, 2016Class IZ-0827-2016Software designThe battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in…
Dec 1, 2015Class IZ-0436-2016Software designThe firm became aware of cases in which the battery run times of the optional PS 500 power supply unit with the Infinity Workstation Critical Care (Evita Infinity V 500) were unexpectedly short due to the design of the charging algorithm…
Jul 13, 2015Class IZ-2343-2015Nonconforming Material/ComponentThe battery capacity of the optional PS500 of an IACS Workstation Critical Care (Evita Infinity V500) did not last as long as expected.
Feb 7, 2014Class IZ-1298-2014Component design/selectionThe battery capacity of the optional PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500) did not last as long as expected.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K133886Perseus A500Draeger Medical GmbH · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K230931AtlanDrägerwerk AG Co. Kgaa · 2023 · 2 generations on · ◆ in the 2023 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • Of its descendants, K181695 (Nihon Kohden Orangemed, Inc., 2018) has gone furthest, with 4 descendants of its own.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The Infinity Acute Care System Workstations Critical Care consist of monitoring and control displays and additional therapy units. They are intended to be used as integrated, networked, and configurable workstations to provide critical care specific therapy. The Infinity Acute Care System Workstations Critical Care are intended to be used by qualified and trained medical personnel. The Infinity C Series Medical Cockpits, consisting of the C500 and the…

Read the full statement (206 words)

The Infinity Acute Care System Workstations Critical Care consist of monitoring and control displays and additional therapy units. They are intended to be used as integrated, networked, and configurable workstations to provide critical care specific therapy. The Infinity Acute Care System Workstations Critical Care are intended to be used by qualified and trained medical personnel.

The Infinity C Series Medical Cockpits, consisting of the C500 and the C700, are monitoring and control displays for the Infinity Acute Care System (IACS). Medical Cockpits are intended to be used to monitor waveforms, parameter information, and alarms as well as to control settings. The Infinity Series Medical Cockpits are intended to be used in environments where patient care is provided by trained healthcare professionals.

The Evita V500 ventilation unit of the Infinity Acute Care System is intended for the ventilation of adult, pediatric and neonatal patients. Evita V500 offers mandatory ventilation modes and ventilation modes for spontaneous breathing support and airway monitoring. The Evita V500 ventilation unit is used with Infinity C Series Dr~ger Medical Cockpits. The Evita V500 ventilation unit is intended for use in different medical care areas.

Evita V500 is intended for stationary use in hospitals and medical rooms or for patient transportation within the hospital.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K093633: Infinity Acute Care System Workstation Critical Care.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K093633

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026