510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Atlan

K230931510(k) TraditionalCleared Jul 23, 2023111-day reviewClass II21 CFR 868.5160Recalls 9 (8 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Drägerwerk AG Co. Kgaa · BSZ Gas-Machine, Anesthesia

Cleared on Jul 23, 2023 as substantially equivalent to K133886 (Draeger Medical GmbH, 2014) after a 111-day review. Its family tree reaches back 7 generations to 1990. 3 of the 33 devices in its family, counting itself, have a Class I recall. 1 later device descends from it. It has been recalled 9 times, including 8 Class I recalls.

Decided
Jul 23, 2023
Review
111 days
Class
Class II
Product code
BSZ
Recalls
9 (8 Class I)
Predicates
1

Its longest line of descent runs back 7 generations to K901713 (1990). 31 ancestors in all, the oldest K900598 (1990). 1 later device descends from it, 1 citing it directly. 3 of the 33 devices in this family, counting this one, have a Class I recall.

  1. Root · 7 generations backK901713 Narkomed 4 Anesthesia SystemDraeger Medical, Inc. · 1990
  2. 6 generations backK980553 Narkomed 6000 Model Number NM6000Draeger Medical, Inc. · 1998
  3. 5 generations backK993826 Narkomed 6000Draeger Medical, Inc. · 2000
  4. 4 generations backK033498 Narkomed 6400, Model NM6400Draeger Medical, Inc. · 2004 · Recalled
  5. 3 generations backK041622 Modification to Fabius GS Anesthesia System; Fabius Tiro Anesthesia…Draeger Medical, Inc. · 2004
  6. 2 generations backK042607 Primus USDraeger Medical, Inc. · 2004 · Recalled
  7. 1 generation backK133886 Perseus A500Draeger Medical GmbH · 2014 · Recalled (Class I)
  8. This deviceK230931 AtlanDrägerwerk AG Co. Kgaa · 2023
  9. Descendant · cites it directly · 0 of its ownK252746 Atlan (A100); Atlan (A100 XL); Atlan (A350); Atlan (A350 XL)Drägerwerk AG Co. Kgaa · 2026

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K252746Atlan (A100); Atlan (A100 XL); Atlan (A350); Atlan (A350 XL)Drägerwerk AG Co. Kgaa · 2026 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K901713Narkomed 4 Anesthesia SystemDraeger Medical, Inc.1990—
−6K980553Narkomed 6000 Model Number NM6000Draeger Medical, Inc.1998—
−5K993826Narkomed 6000Draeger Medical, Inc.2000—
−4K033498Narkomed 6400, Model NM6400Draeger Medical, Inc.20041
−3K041622Modification to Fabius GS Anesthesia System; Fabius Tiro Anesthesia SystemDraeger Medical, Inc.2004—
−2K042607Primus USDraeger Medical, Inc.20042
−1K133886Perseus A500Draeger Medical GmbH2014Class I
0 (this device)K230931AtlanDrägerwerk AG Co. Kgaa2023—
+1K252746Atlan (A100); Atlan (A100 XL); Atlan (A350); Atlan (A350 XL)Drägerwerk AG Co. Kgaa2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199110
199210
199320
199410
199520
199630
199720
199820
199910
200020
200220
200310
200420
201011

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202610
FIG. 1 — Lineage of K230931, 1990–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BSZ device

  1. #1 K253330 A5 Anesthesia System (A5); A7 Anesthesia…score 16.7 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K133886Perseus A500
Summary page 4
…Product code: BSZ Class: II Predicate Device: Perseus A500, K133886 Reference Devices: Dräger Primus US Apollo, K042607 Maquet Flow-I, K191027…
Draeger Medical GmbH2014Class I
†K042607Primus USDraeger Medical, Inc.20042
†K191027Flow-i Anesthesia System, Flow-c Anesthesia System, Flow-e Anesthesia SystemMaquet Critical Care AB2020Class I
†K170872Aisys CS2Datex-Ohmeda, Inc.20174
†K042419Fabius GS/Fabius Tiro Anesthesia SystemDraeger Medical, Inc.2004Class I
†K011404Fabius GS Anesthesia System, Model Fabius GSDraeger Medical, Inc.2002—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K252746Atlan (A100); Atlan (A100 XL); Atlan (A350); Atlan (A350 XL)Drägerwerk AG Co. Kgaa20260

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 7, 2026Class IZ-2208-2026Process change controlThe Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
May 7, 2026Class IZ-2207-2026Process change controlThe Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Sep 9, 2025Class IZ-2617-2025Process controlThe device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Sep 9, 2025Class IZ-2616-2025Process controlThe device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Oct 15, 2024Class IZ-0305-2025Process controlThe Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Oct 15, 2024Class IZ-0304-2025Process controlThe Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Oct 15, 2024Class IZ-0667-2025Process controlThe Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Oct 15, 2024Class IZ-0668-2025Process controlThe Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Nov 17, 2023Class IIZ-0971-2024Use errorInternal backup battery may fail spontaneously while the Atlan is being operated without AC mains supply, resulting in unexpected shutdown of the device while running on batteries.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K133886Perseus A500Draeger Medical GmbH · 2014 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K093633Infinity Acute Care System Workstation Critical CareDraegerwerk AG & CO Kgaa · 2010 · 2 generations back · ◆ in the 2010 ancestors bar of FIG. 14 Class I recalls

What does this lineage say?

  • The line of descent spans 33 years and passes through 3 companies; a new generation was cleared every 4.1 years on median.
  • 6 of the 31 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K252746 (Drägerwerk AG Co. Kgaa, 2026) has gone furthest, with 0 descendants of its own.

How does its category compare?

All BSZ devices

BSZ Gas-Machine, Anesthesia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2023, this device’s year, is highlighted.

FDA decisions per year · BSZ

1in 2025

2007: 32008: 82009: 32010: 12011: 62012: 52013: 52014: 42015: 42016: 12017: 32018: 22020: 12021: 12023: 22025: 12026 (partial year): 2
2007high 8 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BSZ

145in 2017

2007: 175 days median2008: 99 days median2009: 34 days median2011: 218 days median2012: 134 days median2013: 125 days median2014: 235 days median2015: 134 days median2017: 145 days median
2007high 235 (2014) · low 34 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BSZ

5in 2025

2007: 12008: 82009: 32010: 142011: 12012: 7, 6 Class I2013: 5, 5 Class I2015: 7, 2 Class I2016: 12017: 5, 1 Class I2018: 32019: 5, 2 Class I2020: 72021: 2, 1 Class I2022: 4, 3 Class I2023: 62024: 13, 13 Class I2025: 5, 3 Class I2026 (partial year): 7, 2 Class I
2007high 14 (2010) · low 0 (2014)2026

Class I (38 since 2007) Class II/III not classified

Adverse event reports per year · BSZ

9,198in 2025

2007: 1322008: 1042009: 852010: 972011: 872012: 892013: 3082014: 2792015: 5722016: 2,1352017: 2,3552018: 2,5892019: 1,9572020: 2,6872021: 3,1232022: 4,3522023: 5,5952024: 6,0672025: 9,198
2007high 9,198 (2025) · low 85 (2009)2025

MAUDE reports naming product code BSZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200731751132
20088998104
2009334385
20101—1497
20116218187
201251347 (6)89
201351255 (5)308
201442350279
201541347 (2)572
20161—12,135
201731455 (1)2,355
20182—32,589
20190—5 (2)1,957
20201—72,687
20211—2 (1)3,123
20220—4 (3)4,352
20232—65,595
20240—13 (13)6,067
20251—5 (3)9,198
20262—7 (2)—

What is it indicated for?

Intended use This device is intended for use in anesthetizing adults, pediatric patients, and neonates. The device can be used for mechanical ventilation, manual ventilation, pressure-supported spontaneous breathing, and spontaneous breathing. The device is equipped with the following basic functions: — Ventilation monitoring — Inspiratory O2 measurement — Device monitoring — Anesthetic gas receiving system The following options are additionally…

Read the full statement (154 words)

Intended use

This device is intended for use in anesthetizing adults, pediatric patients, and

neonates. The device can be used for mechanical ventilation, manual ventilation,

pressure-supported spontaneous breathing, and spontaneous breathing.

The device is equipped with the following basic functions:

— Ventilation monitoring

— Inspiratory O2 measurement

— Device monitoring

— Anesthetic gas receiving system

The following options are additionally available:

— Patient-gas measurement module for O2, CO2, N20, and anesthetic gases

— 02 insufflation

Anesthesia is achieved through a mixture of pure oxygen and Air (medical

compressed air) or pure oxygen and nitrous oxide, with the addition of volatile

anesthetic agents.

Ventilation is accomplished on the patient through a laryngeal mask, a breathing

mask, or an endotracheal tube.

The integrated breathing system can be used with partial rebreathing (low-flow or

minimum-flow).

Indications

The device is specified for inhalational anesthesia and/or patient ventilation in

accordance with the intended use during surgical or diagnostic interventions.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K230931: Atlan.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K230931

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026