510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Ultrasound General Purpose Guidance System

K093713510(k) TraditionalCleared Jan 19, 201049-day reviewClass II21 CFR 892.1570Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Civco Medical Instruments · ITX Transducer, Ultrasonic, Diagnostic

Cleared on Jan 19, 2010 as substantially equivalent to K882383 (CIVCO Medical Instruments Co., Inc., 1988) and 2 other predicates after a 49-day review. Its family tree reaches back 5 generations to 1985. 4 later devices descend from it. It has been recalled 7 times.

Decided
Jan 19, 2010
Review
49 days
Class
Class II
Product code
ITX
Recalls
7
Predicates
3

Its longest line of descent runs back 5 generations to K850478 (1985). 8 ancestors in all. 4 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2003: 1 other ancestor1988: 1 other ancestor1986: 1 other ancestor2017: 1 descendant cleared2012: 1 descendant cleared2026: 2 descendants clearedK850478 · PLUS 1 · Dymax Corp · 1985K993624 · Site-Rite 3 Ultrasound System · Dymax Corp · 1999K032135 · Site-Rite IV Ultrasound System · Dymax Corp · 2003K052517 · Site-Rite 5 Ultrasound System · C.R. Bard, Inc. · 2005K071204 · Site-Rite 6 Ultrasound System, Model Site-Rite 6 · C.R. Bard, Inc. · 20071990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors4 generations back: 1 other ancestor 1 generation forward: 4 descendants K071204 · Site-Rite 6 Ultrasound System, Model Site-Rite 6 · C.R. Bard, Inc. · 2007K052517 · Site-Rite 5 Ultrasound System · C.R. Bard, Inc. · 2005K032135 · Site-Rite IV Ultrasound System · Dymax Corp · 2003K993624 · Site-Rite 3 Ultrasound System · Dymax Corp · 1999K850478 · PLUS 1 · Dymax Corp · 1985−50
  1. Root · 5 generations backK850478 PLUS 1Dymax Corp · 1985
  2. 4 generations backK993624 Site-Rite 3 Ultrasound SystemDymax Corp · 1999
  3. 3 generations backK032135 Site-Rite IV Ultrasound SystemDymax Corp · 2003
  4. 2 generations backK052517 Site-Rite 5 Ultrasound SystemC.R. Bard, Inc. · 2005
  5. 1 generation backK071204 Site-Rite 6 Ultrasound System, Model Site-Rite 6C.R. Bard, Inc. · 2007
  6. This deviceK093713 Ultrasound General Purpose Guidance SystemCivco Medical Instruments · 2010
  7. Descendant · cites it directly · 0 of its ownK260114 Align® Cover and Pad Guidance ToolCIVCO Medical Instruments Co., Inc. · 2026
  8. Descendant · cites it directly · 0 of its ownK262474 ND Needle Guidance SystemAccurate Access, LLC · 2026
  9. Descendant · cites it directly · 0 of its ownK171040 Reusable guides ranges (PERINE range, NAIL range, CON-LEAD range and…Koelis · 2017
  10. Descendant · cites it directly · 0 of its ownK111426 Euclid (TM) Tier 1-GW Access SystemHouston Medical Robotics · 2012

Most influential descendants

Showing 4 of 4 · See all
  • K260114Align® Cover and Pad Guidance ToolCIVCO Medical Instruments Co., Inc. · 2026 · cites it directly0descendants
  • K262474ND Needle Guidance SystemAccurate Access, LLC · 2026 · cites it directly0descendants
  • K171040Reusable guides ranges (PERINE range, NAIL range, CON-LEAD range and LIN-LEAD range) and…Koelis · 2017 · cites it directly0descendants
  • K111426Euclid (TM) Tier 1-GW Access SystemHouston Medical Robotics · 2012 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K850478PLUS 1Dymax Corp1985—
−4K993624Site-Rite 3 Ultrasound SystemDymax Corp1999—
−3K032135Site-Rite IV Ultrasound SystemDymax Corp2003—
−2K052517Site-Rite 5 Ultrasound SystemC.R. Bard, Inc.2005—
−1K071204Site-Rite 6 Ultrasound System, Model Site-Rite 6C.R. Bard, Inc.2007—
0 (this device)K093713Ultrasound General Purpose Guidance SystemCivco Medical Instruments2010—
+1K260114Align® Cover and Pad Guidance ToolCIVCO Medical Instruments Co., Inc.2026—
+1K262474ND Needle Guidance SystemAccurate Access, LLC2026—
+1K171040Reusable guides ranges (PERINE range, NAIL range, CON-LEAD range and LIN-LEAD…Koelis2017—
+1K111426Euclid (TM) Tier 1-GW Access SystemHouston Medical Robotics2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
198810
200310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201210
201710
202620
FIG. 1 — Lineage of K093713, 1985–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in ITX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new ITX device

  1. #1 K260114 Align® Cover and Pad Guidance Toolscore 13.6 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K882383Maggi Series Ultra. Needle, Biopsy/Catheter Guides
Summary page 2
…Instruments claims the proposed devices to be substantially equivalent to the devices previously cleared by FDA in K882383,1 K071204 and K(030064. CIVCO claims this equivalence because the proposed…
CIVCO Medical Instruments Co., Inc.1988—
*K071204Site-Rite 6 Ultrasound System, Model Site-Rite 6
Summary page 2
…and repeatability. Predicate Device: K(882383 - MAGGI SERIES ULTRA. NEEDLE, BIOPSY/CATHBET'ER GUIDES K071204 - SITE-RITE 6 ULTRASOUND SYSTEMv4 MODEL SITE-RITE 6 K(030064 - ROOK 25 NEE…
C.R. Bard, Inc.2007—
*K030064Ilook 25 Needle Guide Attachment and Bracket Assembly
Summary page 2
…BIOPSY/CATHBET'ER GUIDES K071204 - SITE-RITE 6 ULTRASOUND SYSTEMv4 MODEL SITE-RITE 6 K(030064 - ROOK 25 NEEDLE GUIDE ATITACHMENT AND BRACKET ASSEMBLY Substantial Equivalence: CIVCO Medical Ins…
Sonosite,Inc.200316

Also cleared against K882383: 3 other devices →

What descends from it? (4)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K260114Align® Cover and Pad Guidance ToolCIVCO Medical Instruments Co., Inc.20260
K262474ND Needle Guidance SystemAccurate Access, LLC20260
K171040Reusable guides ranges (PERINE range, NAIL range, CON-LEAD range and LIN-LEAD range) and…Koelis20170
K111426Euclid (TM) Tier 1-GW Access SystemHouston Medical Robotics20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2025Class IIZ-1892-2025Process designDuring the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
Feb 23, 2018Class IIZ-2888-2018Device DesignA cover failed leak testing during validation testing of new needle guides from a replacement tool.
Oct 21, 2015Class IIZ-0287-2016Process designSterility of the product cannot be assured.
Oct 21, 2015Class IIZ-0286-2016Process designSterility of the product cannot be assured.
Oct 21, 2015Class IIZ-0288-2016Process designSterility of the product cannot be assured.
Oct 21, 2015Class IIZ-0290-2016Process designSterility of the product cannot be assured.
Oct 21, 2015Class IIZ-0289-2016Process designSterility of the product cannot be assured.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 25 years and passes through 3 companies; a new generation was cleared every 2.7 years on median.
  • 1 of the 8 ancestors has a recall of any class.
  • Of its descendants, K260114 (CIVCO Medical Instruments Co., Inc., 2026) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K882383.

How does its category compare?

All ITX devices

ITX Transducer, Ultrasonic, Diagnostic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · ITX

4in 2025

2007: 102008: 122009: 172010: 102011: 102012: 72013: 32014: 42015: 12016: 42017: 62018: 22019: 12020: 32021: 32022: 22023: 62024: 32025: 42026 (partial year): 6
2007high 17 (2009) · low 1 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · ITX

126in 2025

2007: 26 days median2008: 73 days median2009: 28 days median2010: 42 days median2011: 16 days median2012: 25 days median2013: 105 days median2014: 41 days median2016: 151 days median2017: 110 days median2020: 120 days median2021: 139 days median2023: 213 days median2024: 81 days median2025: 126 days median2026 (partial year): 158 days median
2007high 213 (2023) · low 16 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · ITX

2in 2025

2008: 42009: 42010: 42011: 12012: 12014: 12015: 52017: 12018: 152020: 32021: 10, 10 Class I2022: 62023: 12024: 12025: 2
2007high 15 (2018) · low 0 (2007)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · ITX

571in 2025

2007: 162008: 392009: 232010: 302011: 342012: 292013: 382014: 432015: 1832016: 2692017: 4122018: 1302019: 1192020: 1162021: 1462022: 872023: 1382024: 9882025: 571
2007high 988 (2024) · low 16 (2007)2025

MAUDE reports naming product code ITX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071026016
20081273439
20091728423
20101042430
20111016134
2012725129
20133105038
2014441143
20151—5183
201641510269
201761101412
20182—15130
20191—0119
202031203116
2021313910 (10)146
20222—687
202362131138
20243811988
202541262571
202661580—

What is it indicated for?

The guidance system is intended for directing instruments such as catheters, electrodes and needles into a targeted anatomical location of a patient relative to the imaging instrument for percutaneous procedures. Prior to use healthcare workers should be trained in ultrasonography. The following models are included in this submission:

MODEL — INTENDED USE

Ultra-Pro WIIT Needle Guides

Ultra-Pro 3"h Needle Guides

AccuSITE Transverse Needle Guides

Infiniti Needle Guides

Multi-Pro 2000 Needle Guides

MAGGI HI Plus Needle Guides — The bracket and disposable guide provide physicians with a tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K093713: Ultrasound General Purpose Guidance System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K093713

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026