510(k) Report

510(k) · Cleared

Cleared

Hotdog Multi Function Controller

K094056510(k) TraditionalCleared Jun 3, 2010154-day reviewClass II21 CFR 870.5900

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Augustine Biomedical & Design, LLC · DWJ System, Thermal Regulating

Cleared on Jun 3, 2010 as substantially equivalent to K052392 (Augustine Biomedical & Design, LLC, 2006) and 1 other predicate after a 154-day review. Its family tree reaches back 4 generations to 2002. 4 later devices descend from it.

Decided
Jun 3, 2010
Review
154 days
Class
Class II
Product code
DWJ
Recalls
None
Predicates
2

Its longest line of descent runs back 4 generations to K011859 (2002). 9 ancestors in all, the oldest K960167 (1996). 4 later devices descend from it, 1 citing it directly.

Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 3 other ancestors3 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 1 descendant4 generations forward: 1 descendant K092807 · Hot Dog Patient Warming Mattress System, Models Wcub, M101, M102 and M103 · Augustine Biomedical & Design, LLC · 2009K051419 · Inditherm Patient Warming System, Models MECU1; OTM1; OTM2; GTM1; PTM1; Otb; RB1 · Inditherm Medical · 2006K031728 · Klimamed Thermal Blankets, Including Arm Shoulder West Thermal Blanket and Leg Warmer Thermal Blanket · Klimamed Technologie Medizingerate GmbH · 2004K011859 · Klimamed Thermal Mat & Controller 95 and 55 Watts · Klimamed Technologie Medizingerate GmbH · 20020
  1. Root · 4 generations backK011859 Klimamed Thermal Mat & Controller 95 and 55 WattsKlimamed Technologie Medizingerate GmbH · 2002
  2. 3 generations backK031728 Klimamed Thermal Blankets, Including Arm Shoulder West Thermal…Klimamed Technologie Medizingerate GmbH · 2004
  3. 2 generations backK051419 Inditherm Patient Warming System, Models MECU1; OTM1; OTM2; GTM1…Inditherm Medical · 2006
  4. 1 generation backK092807 Hot Dog Patient Warming Mattress System, Models Wcub, M101, M102 and…Augustine Biomedical & Design, LLC · 2009
  5. This deviceK094056 Hotdog Multi Function ControllerAugustine Biomedical & Design, LLC · 2010
  6. Descendant · cites it directly · 3 of its ownK201779 HotDog Temperature Management ControllerAugustine Temperature Management · 2020 · Recalled
  7. Descendant · 2 generations on · 2 of its ownK220941 HotDog Temperature Management SystemAugustine Temperature Management · 2022 · Recalled
  8. Descendant · 3 generations on · 1 of its ownK230866 HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC · 2023 · Recalled
  9. Descendant · 4 generations on · 0 of its ownK232627 HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC · 2024

Most influential descendants

Showing 4 of 4 · See all
  • K201779HotDog Temperature Management ControllerAugustine Temperature Management · 2020 · cites it directly3descendants
  • K220941HotDog Temperature Management SystemAugustine Temperature Management · 2022 · 2 generations on2descendants
  • K230866HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC · 2023 · 3 generations on1descendant
  • K232627HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC · 2024 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K011859Klimamed Thermal Mat & Controller 95 and 55 WattsKlimamed Technologie Medizingerate GmbH2002—
−3K031728Klimamed Thermal Blankets, Including Arm Shoulder West Thermal Blanket and Leg…Klimamed Technologie Medizingerate GmbH2004—
−2K051419Inditherm Patient Warming System, Models MECU1; OTM1; OTM2; GTM1; PTM1; Otb; RB1Inditherm Medical2006—
−1K092807Hot Dog Patient Warming Mattress System, Models Wcub, M101, M102 and M103Augustine Biomedical & Design, LLC2009—
0 (this device)K094056Hotdog Multi Function ControllerAugustine Biomedical & Design, LLC2010—
+1K201779HotDog Temperature Management ControllerAugustine Temperature Management20204
+2K220941HotDog Temperature Management SystemAugustine Temperature Management20224
+3K230866HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC20234
+4K232627HotDog Warming Mattress + Return ElectrodeAugustine Temperature Management, LLC2024—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
200020
200210
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
202210
202310
202410
FIG. 1 — Lineage of K094056, 2002–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DWJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWJ device

  1. #1 K260278 ThermAffyx (TM) Patient Safety Systemscore 12.8 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K052392Hot Dog Patient Warming System
Summary page 1
…er: WC52 Product Code DWJ (21 CFR Part 870.5900) Common Thermal Regulating System Name Predicate Hot Dog Patient Warming System K052392 Device Hot Dog Patient Warming Mattress System K092807 Device The Hot Dog Multifunction Controll…
Augustine Biomedical & Design, LLC2006—
*K092807Hot Dog Patient Warming Mattress System, Models Wcub, M101, M102 and M103
Summary page 1
…Thermal Regulating System Name Predicate Hot Dog Patient Warming System K052392 Device Hot Dog Patient Warming Mattress System K092807 Device The Hot Dog Multifunction Controller consists of a temperature control Description unit that m…
Augustine Biomedical & Design, LLC2009—

Also cleared against K052392: 4 other devices →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K201779HotDog Temperature Management ControllerAugustine Temperature Management20203

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 3 companies; a new generation was cleared every 2.5 years on median.
  • 1 of the 9 ancestors has a recall of any class.
  • Of its descendants, K201779 (Augustine Temperature Management, 2020) has gone furthest, with 3 descendants of its own.
  • 4 other devices also cite its predicate K052392.

How does its category compare?

All DWJ devices

DWJ System, Thermal Regulating: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · DWJ

3in 2025

2007: 42008: 32009: 32010: 82011: 42012: 52013: 52014: 42016: 22017: 52018: 32019: 22020: 52021: 22022: 22023: 62024: 22025: 32026 (partial year): 5
2007high 8 (2010) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWJ

131in 2025

2007: 106 days median2008: 73 days median2009: 106 days median2010: 134 days median2011: 180 days median2012: 69 days median2013: 77 days median2014: 131 days median2017: 153 days median2018: 95 days median2020: 30 days median2023: 42 days median2025: 131 days median2026 (partial year): 99 days median
2007high 180 (2011) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DWJ

3in 2024

2009: 62010: 22011: 72012: 12013: 22015: 12017: 32018: 52019: 32020: 22021: 12022: 52024: 32026 (partial year): 3
2007high 7 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DWJ

2,384in 2025

2007: 522008: 402009: 712010: 1042011: 1362012: 1062013: 732014: 1162015: 2452016: 1,4002017: 1,4732018: 1,5172019: 2,2942020: 2,5462021: 2,6892022: 3,2262023: 3,4552024: 3,3292025: 2,384
2007high 3,455 (2023) · low 40 (2008)2025

MAUDE reports naming product code DWJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074106052
2008373040
20093106671
201081342104
201141807136
20125691106
2013577273
201441310116
20150—1245
20162—01,400
2017515331,473
201839551,517
20192—32,294
202053022,546
20212—12,689
20222—53,226
202364203,455
20242—33,329
2025313102,384
20265993—

What is it indicated for?

The Hot Dog Patient Warming System is intended to prevent or treat hypothermia and to provide warmth to patients. The Hot Dog Patient Multifunction Controller should be used in circumstances in which patients may not maintain a state of normothermia.

The System is intended primarily for use in hospitals and surgical centers including without limitation operating, recovery and emergency rooms and on medical/surgical floors.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K094056: Hotdog Multi Function Controller.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K094056

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026