510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Disposable Reflective Marker Spheres

K100038510(k) TraditionalCleared Jul 14, 2010188-day reviewClass II21 CFR 882.4560Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Brainlab AG · HAW Neurological Stereotaxic Instrument

Cleared on Jul 14, 2010 as substantially equivalent to K082060 (Brainlab AG, 2009) after a 188-day review. Its family tree reaches back 4 generations to 1999. 2 later devices descend from it. It has been recalled 3 times.

Decided
Jul 14, 2010
Review
188 days
Class
Class II
Product code
HAW
Recalls
3
Predicates
1

Its longest line of descent runs back 4 generations to K983831 (1999). 8 ancestors in all. 2 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor2003: 1 other ancestor2001: 1 other ancestor2000: 1 other ancestor2011: 2 descendants clearedK983831 · VECTORVISION2 · Brainlab AG · 1999K014256 · Kolibri Igs Systems · Brainlab AG · 2002K042391 · Kolibri Image Guided Surgery System · Brainlab AG · 2004K082060 · Cranial Image Guided Surgery System · Brainlab AG · 20091995200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor3 generations back: 2 other ancestors4 generations back: 1 other ancestor 1 generation forward: 2 descendants K082060 · Cranial Image Guided Surgery System · Brainlab AG · 2009K042391 · Kolibri Image Guided Surgery System · Brainlab AG · 2004K014256 · Kolibri Igs Systems · Brainlab AG · 2002K983831 · VECTORVISION2 · Brainlab AG · 19990
  1. Root · 4 generations backK983831 VECTORVISION2Brainlab AG · 1999
  2. 3 generations backK014256 Kolibri Igs SystemsBrainlab AG · 2002
  3. 2 generations backK042391 Kolibri Image Guided Surgery SystemBrainlab AG · 2004
  4. 1 generation backK082060 Cranial Image Guided Surgery SystemBrainlab AG · 2009
  5. This deviceK100038 Disposable Reflective Marker SpheresBrainlab AG · 2010
  6. Descendant · cites it directly · 1 of its ownK103192 Navigation Marker, Reflective Marker SpheresIlumark GmbH · 2011
  7. Descendant · cites it directly · 0 of its ownK111443 X12 Sterile Reflective Marker SpherePinsco Inc. Dba B & L Engineering · 2011

Most influential descendants

Showing 2 of 2 · See all
  • K103192Navigation Marker, Reflective Marker SpheresIlumark GmbH · 2011 · cites it directly1descendant
  • K111443X12 Sterile Reflective Marker SpherePinsco Inc. Dba B & L Engineering · 2011 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K983831VECTORVISION2Brainlab AG1999—
−3K014256Kolibri Igs SystemsBrainlab AG2002—
−2K042391Kolibri Image Guided Surgery SystemBrainlab AG2004—
−1K082060Cranial Image Guided Surgery SystemBrainlab AG2009—
0 (this device)K100038Disposable Reflective Marker SpheresBrainlab AG2010—
+1K103192Navigation Marker, Reflective Marker SpheresIlumark GmbH2011—
+1K111443X12 Sterile Reflective Marker SpherePinsco Inc. Dba B & L Engineering2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200110
200310
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201120
FIG. 1 — Lineage of K100038, 1999–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HAW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HAW device

  1. #1 K253856 Geniant Cranial (Navigated Neurosurgical…score 15.0 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K082060Cranial Image Guided Surgery System
Summary page 1
…ective Marker Spheres Regulation number: 21 CFR 882.4560 Product codes: HAW, OLO Predicate Device: Cranial Image Guided Surgery System K082060 Device Classification Name: Instrument, Stereotaxic Regulatory Class: Class 1I This submission does not change the indications fo…
Brainlab AG2009—

Also cleared against K082060: 4 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K103192Navigation Marker, Reflective Marker SpheresIlumark GmbH20111
K111443X12 Sterile Reflective Marker SpherePinsco Inc. Dba B & L Engineering20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 18, 2015Class IIZ-2795-2015Process controlDisposable Reflective Marker Spheres (DRMS) for Brainlab Image Guided Surgery (IGS) Systems: DRMS spheres may separate at the mid-point where the two halves of the sphere are sealed together.
Jun 10, 2015Class IIZ-2084-2015Process controlInadequate curing of adhesive, may cause the two halves of the spheres to separate.
May 20, 2015Class IIZ-2093-2015Process controlInadequate curing of adhesive, which may cause the two halves of the spheres to separate.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 1 company; a new generation was cleared every 3.2 years on median.
  • 2 of the 8 ancestors have a recall of any class.
  • Of its descendants, K103192 (Ilumark GmbH, 2011) has gone furthest, with 1 descendant of its own.
  • 4 other devices also cite its predicate K082060.

How does its category compare?

All HAW devices

HAW Neurological Stereotaxic Instrument: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · HAW

9in 2025

2007: 152008: 122009: 82010: 62011: 42012: 82013: 32014: 52015: 102016: 92017: 72018: 62019: 102020: 52021: 92022: 72023: 142024: 72025: 92026 (partial year): 5
2007high 15 (2007) · low 3 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HAW

163in 2025

2007: 107 days median2008: 112 days median2009: 221 days median2010: 138 days median2011: 66 days median2012: 178 days median2013: 172 days median2014: 238 days median2015: 232 days median2016: 175 days median2017: 226 days median2018: 112 days median2019: 147 days median2020: 88 days median2021: 168 days median2022: 128 days median2023: 142 days median2024: 107 days median2025: 163 days median2026 (partial year): 162 days median
2007high 238 (2014) · low 66 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HAW

2in 2025

2007: 22008: 102009: 62010: 282011: 52012: 42013: 6, 1 Class I2014: 52015: 32016: 52017: 22018: 2, 2 Class I2019: 4, 1 Class I2020: 12021: 4, 2 Class I2022: 12023: 4, 1 Class I2024: 2, 2 Class I2025: 22026 (partial year): 6
2007high 28 (2010) · low 1 (2020)2026

Class I (9 since 2007) Class II/III not classified

Adverse event reports per year · HAW

3,425in 2025

2007: 492008: 532009: 962010: 2992011: 1,7252012: 9862013: 1,2902014: 1,6062015: 1,7672016: 1,8282017: 3,5052018: 4,4392019: 6,0622020: 3,4802021: 2,8272022: 1,9292023: 2,5102024: 3,7552025: 3,425
2007high 6,062 (2019) · low 49 (2007)2025

MAUDE reports naming product code HAW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715107249
2008121121053
20098221696
2010613828299
201146651,725
201281784986
201331726 (1)1,290
2014523851,606
20151023231,767
2016917551,828
2017722623,505
201861122 (2)4,439
2019101474 (1)6,062
202058813,480
202191684 (2)2,827
2022712811,929
2023141424 (1)2,510
202471072 (2)3,755
2025916323,425
202651626—

What is it indicated for?

Disposable reflective marker spheres are attached to reference arrays and instruments, thus enabling infrared tracking systems to detect the position of the patient and instruments in the surgical field.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K100038: Disposable Reflective Marker Spheres.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K100038

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026