510(k) Report

510(k) · Cleared

Recalled · 2010

Masimo Rainbow Set Pronto -7 Pulse CO- Oximeter and Accessories, Model Pronto-7

K100403510(k) TraditionalCleared Jun 23, 2010127-day reviewClass II21 CFR 870.2700Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Masimo Corporation · DQA Oximeter

Cleared on Jun 23, 2010 as substantially equivalent to K091057 (Masimo Corporation, 2009) after a 127-day review. Its family tree reaches back 9 generations to 2000. 1 later device descends from it. It has been recalled once.

Decided
Jun 23, 2010
Review
127 days
Class
Class II
Product code
DQA
Recalls
1
Predicates
1

Its longest line of descent runs back 9 generations to K000126 (2000). 11 ancestors in all, the oldest K853990 (1985). 1 later device descends from it, 1 citing it directly.

  1. Root · 9 generations backK000126 Masimo Set Radical Pulse Oximeter with Satshare, Masimo…Masimo Corp. · 2000
  2. 5 generations between
  3. 3 generations backK080238 Masimo Rainbow Set Radical 7 Pulse CO-Oximeter, Masimo Rainbow Set…Masimo Corporation · 2008 · Recalled
  4. 2 generations backK082052 Masimo Rainbow Set Radcheck Pulse CO-Oximeter, Radcheck, Rad 57 Spot…Masimo Corporation · 2008
  5. 1 generation backK091057 Masimo Rainbow Set Pronto Pulse CO-Oximeter and AccessoriesMasimo Corporation · 2009
  6. This deviceK100403 Masimo Rainbow Set Pronto -7 Pulse CO- Oximeter and Accessories…Masimo Corporation · 2010
  7. Descendant · cites it directly · 0 of its ownK111403 Masimo Rainbow Set Pronto-7 Pulse CO-Oximeter and AccessoriesMasimo Corporation · 2011
Show all 9 generations
  1. Root · 9 generations backK000126 Masimo Set Radical Pulse Oximeter with Satshare, Masimo…Masimo Corp. · 2000
  2. 8 generations backK031330 Modification to Masimo Set Radical Pulse Oximeter with Satshare and…Masimo Corp. · 2003
  3. 7 generations backK033296 Masimo Set Rad-5 Pulse Oximeter, Model Rad-5; Masimo Set Rad-5V…Masimo Corp. · 2004 · Recalled
  4. 6 generations backK042536 Masimo Set Rad 57 Pulse CO-OximeterMasimo Corporation · 2005 · Recalled
  5. 5 generations backK053477 Masimo Set Rad 57 CM and M Pulse CO-OximetersMasimo Corporation · 2006
  6. 4 generations backK061204 Masimo Set Radical 7 Pulse CO-OximeterMasimo Corporation · 2006
  7. 3 generations backK080238 Masimo Rainbow Set Radical 7 Pulse CO-Oximeter, Masimo Rainbow Set…Masimo Corporation · 2008 · Recalled
  8. 2 generations backK082052 Masimo Rainbow Set Radcheck Pulse CO-Oximeter, Radcheck, Rad 57 Spot…Masimo Corporation · 2008
  9. 1 generation backK091057 Masimo Rainbow Set Pronto Pulse CO-Oximeter and AccessoriesMasimo Corporation · 2009
  10. This deviceK100403 Masimo Rainbow Set Pronto -7 Pulse CO- Oximeter and Accessories…Masimo Corporation · 2010

Most influential descendants

Showing 1 of 1 · See all
  • K111403Masimo Rainbow Set Pronto-7 Pulse CO-Oximeter and AccessoriesMasimo Corporation · 2011 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K000126Masimo Set Radical Pulse Oximeter with Satshare, Masimo…Masimo Corp.2000—
−8K031330Modification to Masimo Set Radical Pulse Oximeter with Satshare and Lnop…Masimo Corp.2003—
−7K033296Masimo Set Rad-5 Pulse Oximeter, Model Rad-5; Masimo Set Rad-5V Pulse…Masimo Corp.20041
−6K042536Masimo Set Rad 57 Pulse CO-OximeterMasimo Corporation20051
−5K053477Masimo Set Rad 57 CM and M Pulse CO-OximetersMasimo Corporation2006—
−4K061204Masimo Set Radical 7 Pulse CO-OximeterMasimo Corporation2006—
−3K080238Masimo Rainbow Set Radical 7 Pulse CO-Oximeter, Masimo Rainbow Set Rad 87…Masimo Corporation20081
−2K082052Masimo Rainbow Set Radcheck Pulse CO-Oximeter, Radcheck, Rad 57 Spot CheckMasimo Corporation2008—
−1K091057Masimo Rainbow Set Pronto Pulse CO-Oximeter and AccessoriesMasimo Corporation2009—
0 (this device)K100403Masimo Rainbow Set Pronto -7 Pulse CO- Oximeter and Accessories, Model Pronto-7Masimo Corporation2010—
+1K111403Masimo Rainbow Set Pronto-7 Pulse CO-Oximeter and AccessoriesMasimo Corporation2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198510
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
FIG. 1 — Lineage of K100403, 2000–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQA device

  1. #1 K221282 Empatica Health Monitoring Platformscore 20.1 · 38 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K091057Masimo Rainbow Set Pronto Pulse CO-Oximeter and Accessories
Summary page 1
…A Predicate Device: Masimo Rainbow SEr1 Pronto Pulse CO-Oximeter and Accessories, 510O(k) Number - K091 057 Device Description The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories (Pronto-?) include the noninvasive technol…
Masimo Corporation2009—

Also cleared against K091057: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K111403Masimo Rainbow Set Pronto-7 Pulse CO-Oximeter and AccessoriesMasimo Corporation20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 23, 2010Not classifiedZ-1387-2011Process designMasimo Corporation is initiating a voluntary recall on the Pronto-7 rainbow 4D Resuable Sensor because when performing an SpHb measurement, the sensor may incorrectly read actual finger temperature. As a result, Pronto-7 Check Pulse…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 11 months on median.
  • 3 of the 11 ancestors have a recall of any class.
  • Of its descendants, K111403 (Masimo Corporation, 2011) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K091057.

How does its category compare?

All DQA devices

DQA Oximeter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · DQA

15in 2025

2007: 202008: 222009: 352010: 292011: 152012: 142013: 152014: 112015: 132016: 142017: 92018: 182019: 192020: 152021: 102022: 202023: 132024: 142025: 152026 (partial year): 8
2007high 35 (2009) · low 9 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQA

182in 2025

2007: 122 days median2008: 97 days median2009: 117 days median2010: 127 days median2011: 195 days median2012: 120 days median2013: 182 days median2014: 175 days median2015: 158 days median2016: 246 days median2017: 200 days median2018: 260 days median2019: 158 days median2020: 126 days median2021: 174 days median2022: 265 days median2023: 317 days median2024: 285 days median2025: 182 days median2026 (partial year): 187 days median
2007high 317 (2023) · low 97 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQA

3in 2024

2007: 62008: 52009: 292010: 42011: 12014: 22015: 12018: 322024: 3
2007high 32 (2018) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DQA

399in 2025

2007: 3252008: 3122009: 2812010: 1852011: 1642012: 3222013: 4682014: 1,1242015: 2,0822016: 1,7752017: 1,1902018: 6942019: 5482020: 4262021: 5532022: 4782023: 3922024: 5452025: 399
2007high 2,082 (2015) · low 164 (2011)2025

MAUDE reports naming product code DQA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007201226325
200822975312
20093511729281
2010291274185
2011151951164
2012141200322
2013151820468
20141117521,124
20151315812,082
20161424601,775
2017920001,190
20181826032694
2019191580548
2020151260426
2021101740553
2022202650478
2023133170392
2024142853545
2025151820399
202681870—

What is it indicated for?

The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for noninvasive spot checking of functional saturation of arterial oxygen hemoglobin (SP02), pulse rate, and total hemoglobin concentration (SpHlb). The Masimo Rainbow SET Pronto-7 Pulse CO-Oximeter and Accessories are indicated for use by trained personnel, with adult and pediatric individuals, in clinical and non-clinical settings (e.g., hospitals, hospital-type facilities, home, clinics, physician offices. blood donation facilities, and ambulatory surgery centers).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K100403: Masimo Rainbow Set Pronto -7 Pulse CO- Oximeter and Accessories, Model Pronto-7.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K100403

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026