510(k) Report

510(k) · Cleared

Recalled · Class II · 2011

Varian High Energy Linear Accelerator

K100890510(k) SpecialCleared Jul 13, 2010104-day reviewClass II21 CFR 892.5050Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Varian Medical Systems · IYE Accelerator, Linear, Medical

Cleared on Jul 13, 2010 as substantially equivalent to K081188 (Varian Medical Systems, Inc., 2008) and 1 other predicate after a 104-day review. Its family tree reaches back 6 generations to 1991. 4 later devices descend from it. It has been recalled 11 times.

Decided
Jul 13, 2010
Review
104 days
Class
Class II
Product code
IYE
Recalls
11
Predicates
2

Its longest line of descent runs back 6 generations to K913119 (1991). 8 ancestors in all. 4 later devices descend from it, 1 citing it directly.

  1. Root · 6 generations backK913119 Varian C-Series Clinacs (Clinac 2300C/D)Varian Assoc., Inc. · 1991 · Recalled
  2. 5 generations backK033343 Trilogy Radiotherapy Delivery SystemVarian Medical Systems, Inc. · 2003 · Recalled
  3. 4 generations backK061140 Trilogy Radiotherapy Delivery SystemVarian Medical Systems · 2006
  4. 3 generations backK070094 Trilogy Radiotherapy Delivery System and Trilogy TX Delivery SystemVarian Medical Systems, Inc. · 2007 · Recalled
  5. 2 generations backK072916 Trilogy System with RapidarcVarian Medical Systems, Inc. · 2007 · Recalled
  6. 1 generation backK081188 Trilogy with External System Gating InterfaceVarian Medical Systems, Inc. · 2008 · Recalled
  7. This deviceK100890 Varian High Energy Linear AcceleratorVarian Medical Systems · 2010
  8. Descendant · cites it directly · 3 of its ownK112839 Varian High Energy Linear AcceleratorVarian Medical Systems · 2011 · Recalled
  9. Descendant · 2 generations on · 2 of its ownK131807 High Energy Linear AcceleratorVarian Medical Systems, Inc. · 2013 · Recalled
  10. Descendant · 3 generations on · 1 of its ownK162476 Varian High Energy Linear AcceleratorVarian Medical Systems, Inc. · 2016
  11. Descendant · 4 generations on · 0 of its ownK181404 Varian High Energy Linear Accelerator, UNIQUE, Varian Multileaf…Varian Medical Systems, Inc. · 2018 · Recalled

Most influential descendants

Showing 4 of 4 · See all
  • K112839Varian High Energy Linear AcceleratorVarian Medical Systems · 2011 · cites it directly3descendants
  • K131807High Energy Linear AcceleratorVarian Medical Systems, Inc. · 2013 · 2 generations on2descendants
  • K162476Varian High Energy Linear AcceleratorVarian Medical Systems, Inc. · 2016 · 3 generations on1descendant
  • K181404Varian High Energy Linear Accelerator, UNIQUE, Varian Multileaf CollimatorVarian Medical Systems, Inc. · 2018 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K913119Varian C-Series Clinacs (Clinac 2300C/D)Varian Assoc., Inc.19916
−5K033343Trilogy Radiotherapy Delivery SystemVarian Medical Systems, Inc.20033
−4K061140Trilogy Radiotherapy Delivery SystemVarian Medical Systems2006—
−3K070094Trilogy Radiotherapy Delivery System and Trilogy TX Delivery SystemVarian Medical Systems, Inc.20075
−2K072916Trilogy System with RapidarcVarian Medical Systems, Inc.20072
−1K081188Trilogy with External System Gating InterfaceVarian Medical Systems, Inc.20081
0 (this device)K100890Varian High Energy Linear AcceleratorVarian Medical Systems2010—
+1K112839Varian High Energy Linear AcceleratorVarian Medical Systems20111
+2K131807High Energy Linear AcceleratorVarian Medical Systems, Inc.20132
+3K162476Varian High Energy Linear AcceleratorVarian Medical Systems, Inc.2016—
+4K181404Varian High Energy Linear Accelerator, UNIQUE, Varian Multileaf CollimatorVarian Medical Systems, Inc.20181

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201310
201610
201810
FIG. 1 — Lineage of K100890, 1991–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K081188Trilogy with External System Gating Interface
Summary page 1
…Class II Product Code: 90 lYE 3. Predicate Device: Varian Trilogy Radiotherapy System: K081 188, K072916 4. Description of the Device: The Varian High Energy Linear Accelerator models provide various selec…
Varian Medical Systems, Inc.20081
*K072916Trilogy System with Rapidarc
Summary page 1
…Class II Product Code: 90 lYE 3. Predicate Device: Varian Trilogy Radiotherapy System: K081 188, K072916 4. Description of the Device: The Varian High Energy Linear Accelerator models provide various selections amo…
Varian Medical Systems, Inc.20072

Also cleared against K081188: 1 other device →

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K112839Varian High Energy Linear AcceleratorVarian Medical Systems20113

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 9, 2011Not classifiedZ-0181-2012Device DesignVarian has received reports in which a user has remotely rotated the gantry into contact with the couch or with the patient, in both the manual mode and the automate mode.
May 31, 2011Class IIZ-1325-2013Process designUnder certain conditions, photon beams in High Energy Clinacs may experience a gradual change in beam symmetry, potentially reaching an asymmetry of up to about 7% before interlock occurs. At maximum asymmetry, this may result in no more…
May 26, 2011Not classifiedZ-2978-2011Device DesignOn Varian Linear Accelerators where the customer may calibrate the collimator angle position readout in reverse, switching the 90 degree and 270 degree positions.
May 24, 2011Not classifiedZ-2893-2011Device DesignThe throat cover on the High Energy Clinac may detach if not properly installed and possibly strike a patient.
May 18, 2011Not classifiedZ-2764-2011OtherThe Coolant System input water supply manifold may leak in some Clinac Linear Accelerators, posing a risk of electrical shock to any person working within the protective housing.
May 13, 2011Not classifiedZ-2778-2011Nonconforming Material/ComponentSome drive stand bolts may fail to meet mechanical specifications.
Mar 31, 2011Not classifiedZ-2224-2011Process designThe coolant may leak. The resulting coolant leak presents a risk of electrical shock to any person working within the protective housing.
Mar 28, 2011Not classifiedZ-2195-2011Process controlVarian has identified an anomaly whereby, following prolonged use, the screw fastener holding the wedge body to the tray may fail.
Mar 17, 2011Not classifiedZ-3204-2011Component design/selectionAn improperly installed collimator cover may detach and fall from the machine, potentially striking the patient or nearby personnel.
Feb 24, 2011Not classifiedZ-2046-2011Labeling mix-upsThe wedge angle labeling on the wedge tray may not correctly match the wedge body. However, the wedge body is labeled with the correct wedge angle.
Dec 10, 2010Not classifiedZ-1045-2011Process controlAn anomaly was identified whereby the bolts used to fasten the counterweight to the gantry may not, in some cases, be fully tightened to the required torque specification.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 19 years and passes through 3 companies; a new generation was cleared every 2.0 years on median.
  • 6 of the 8 ancestors have a recall of any class.
  • Of its descendants, K112839 (Varian Medical Systems, 2011) has gone furthest, with 3 descendants of its own.
  • 1 other device also cites its predicate K081188.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K100890: Varian High Energy Linear Accelerator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K100890

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.