Frankenman Surgical Staplers

K101378510(k) TraditionalCleared Jun 29, 201043-day reviewClass II21 CFR 878.4750

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Suzhou Frankenman Medical Equipment Co., Ltd. · GDW Staple, Implantable

Cleared on Jun 29, 2010 as substantially equivalent to K090821 (Frankenman International Limited, 2009) and 2 other predicates after a 43-day review. Its family tree reaches back 5 generations to 1994. 5 of the 16 devices in its family, counting itself, have a Class I recall. 1 later device descends from it.

Decided
Jun 29, 2010
Review
43 days
Class
Class II
Product code
GDW
Recalls
None
Predicates
4

Its longest line of descent runs back 5 generations to K940967 (1994). 14 ancestors in all, the oldest K801590 (1980). 1 later device descends from it, 1 citing it directly. 5 of the 16 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1992: 1 other ancestor1982: 1 other ancestor2005: 1 other ancestor, 1 with a Class I recall12003: 1 other ancestor2002: 1 other ancestor2000: 1 other ancestor1995: 1 other ancestor1991: 1 other ancestor1980: 1 other ancestor2014: 1 descendant clearedK940967 · Modified Endopath Ils · Ethicon Endo-Surgery, Inc. · 1994 · Class I recall×2: 2 generations cleared close together, K983536 (1998), K991030 (1999) K030925 · Proximate HCS Hemorrhoidal Circular Stapler and Accessories, Model PPH01 · Ethicon Endo-Surgery, Inc. · 2003 · Class I recallK090821 · Chex Single Use Curved Intraluminal Circular Stapler, Model CS 21, CS 25, CS 28, CS 32, CS 21L, CS 25L, CS 28L, CS 32L · Frankenman International Limited · 200919801990200020102020
  1. Root · 5 generations backK940967 Modified Endopath IlsEthicon Endo-Surgery, Inc. · 1994 · Recalled (Class I)
  2. 4 generations backK983536 Proximate Curved and Straight Intraluminal StaplersEthicon Endo-Surgery, Inc. · 1998 · Recalled (Class I)
  3. 3 generations backK991030 Proximate HCS Hemorrhoidal Circular Stapler and AccessoriesEthicon Endo-Surgery, Inc. · 1999 · Recalled (Class I)
  4. 2 generations backK030925 Proximate HCS Hemorrhoidal Circular Stapler and Accessories, Model…Ethicon Endo-Surgery, Inc. · 2003 · Recalled (Class I)
  5. 1 generation backK090821 Chex Single Use Curved Intraluminal Circular Stapler, Model CS 21…Frankenman International Limited · 2009
  6. This deviceK101378 Frankenman Surgical StaplersSuzhou Frankenman Medical Equipment Co., Ltd. · 2010
  7. Descendant · cites it directly · 0 of its ownK133499 Suzhou Beinuo Surgical Staplers - Circular Staplers for Single Use…Suzhou Beinuo Medical Equipment Co., Ltd. · 2014

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K133499Suzhou Beinuo Surgical Staplers - Circular Staplers for Single Use, Hemorrhoidal…Suzhou Beinuo Medical Equipment Co., Ltd. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K940967Modified Endopath IlsEthicon Endo-Surgery, Inc.1994Class I
−4K983536Proximate Curved and Straight Intraluminal StaplersEthicon Endo-Surgery, Inc.1998Class I
−3K991030Proximate HCS Hemorrhoidal Circular Stapler and AccessoriesEthicon Endo-Surgery, Inc.1999Class I
−2K030925Proximate HCS Hemorrhoidal Circular Stapler and Accessories, Model PPH01Ethicon Endo-Surgery, Inc.2003Class I
−1K090821Chex Single Use Curved Intraluminal Circular Stapler, Model CS 21, CS 25, CS…Frankenman International Limited2009—
0 (this device)K101378Frankenman Surgical StaplersSuzhou Frankenman Medical Equipment Co., Ltd.2010—
+1K133499Suzhou Beinuo Surgical Staplers - Circular Staplers for Single Use…Suzhou Beinuo Medical Equipment Co., Ltd.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198010
198210
199110
199210
199510
200010
200210
200310
200511

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
FIG. 1 — Lineage of K101378, 1994–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GDW device

  1. #1 K253657 Tri-staple 2.0™ Reloads; Endo GIA™ Reloads…score 15.0 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
*K090821Chex Single Use Curved Intraluminal Circular Stapler, Model CS 21, CS 25, CS 28, CS 32…
Summary page 3
…Single Use Circular ChexTM Single Use Circular raennTM Stapler for Rectal Stapler for Rectal Prolapse and Intrnational Lmtd K090821 Prolapse andHeorodInentoaLitd HemorrhoidsHeorid Disposable Reloadable ChexT m Single Use Reloadlable Fr…
Frankenman International Limited2009—
*K920752Endopath Ils Endoscopic Circular Stapler
Summary page 4
…aluminal Circular Stapler International Limited K002 Canal Staplers Endlopath ILS Endloscopic Ethicon K920752 Circular Stapler Single Use Circular Ce T igeUeCrua Stapler for Rectal Staper…
ETHICON, Inc.1992—
Kept off the family treeK091002Chex Single Use Reloadable Linear Stapler and Reloads
Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree
*K822345Disposable Linear Stapler (LSF)
Summary page 4
…apler and Linear Stapler and Reloads International Limited K082 Reloads Disposable Linear Stapler (LSF) Ethicon K822345 Disposable Reloadlable ChexTm Linear Cutter Disposable FrankenmanTm Linear Cutter Stapler Reloadabl…
ETHICON, Inc.1982—

Also cleared against K090821: 2 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K133499Suzhou Beinuo Surgical Staplers - Circular Staplers for Single Use, Hemorrhoidal…Suzhou Beinuo Medical Equipment Co., Ltd.20140

Has anything in its family been recalled? (5 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K051301Proximate PPH Hemorrhoidal Circular Stapler and AccessoriesEthicon Endo-Surgery, Inc. · 2005 · 2 generations back · ◆ in the 2005 ancestors bar of FIG. 11 Class I recall
  • K030925Proximate HCS Hemorrhoidal Circular Stapler and Accessories, Model PPH01Ethicon Endo-Surgery, Inc. · 2003 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K991030Proximate HCS Hemorrhoidal Circular Stapler and AccessoriesEthicon Endo-Surgery, Inc. · 1999 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K983536Proximate Curved and Straight Intraluminal StaplersEthicon Endo-Surgery, Inc. · 1998 · 3 generations back · ◆ on the line of descent in FIG. 112 Class I recalls
  • K940967Modified Endopath IlsEthicon Endo-Surgery, Inc. · 1994 · 4 generations back · ◆ on the line of descent in FIG. 14 Class I recalls

What does this lineage say?

  • The line of descent spans 16 years and passes through 3 companies; a new generation was cleared every 4.0 years on median.
  • 5 of the 14 ancestors have a recall of any class; 5 have a Class I recall.
  • Of its descendants, K133499 (Suzhou Beinuo Medical Equipment Co., Ltd., 2014) has gone furthest, with 0 descendants of its own.
  • 2 other devices also cite its predicate K090821.

How does its category compare?

All GDW devices

GDW Staple, Implantable: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · GDW

3in 2025

2007: 82008: 72009: 132010: 102011: 112012: 92013: 132014: 152015: 82016: 112017: 132018: 72019: 122020: 102021: 92022: 72023: 42024: 22025: 32026 (partial year): 3
2007high 15 (2014) · low 2 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GDW

111in 2025

2007: 56 days median2008: 57 days median2009: 27 days median2010: 39 days median2011: 49 days median2012: 45 days median2013: 49 days median2014: 63 days median2015: 111 days median2016: 96 days median2017: 103 days median2018: 90 days median2019: 205 days median2020: 88 days median2021: 233 days median2022: 89 days median2023: 267 days median2025: 111 days median2026 (partial year): 141 days median
2007high 267 (2023) · low 27 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GDW

3in 2025

2007: 92009: 222010: 22011: 92012: 11, 1 Class I2013: 12015: 62016: 22017: 32018: 32019: 4, 4 Class I2020: 112023: 12025: 3
2007high 22 (2009) · low 0 (2008)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · GDW

11,905in 2025

2007: 2,4012008: 2,3802009: 4,0142010: 3,7682011: 3,8262012: 3,7172013: 3,7772014: 4,7842015: 4,8992016: 5,1062017: 11,6432018: 14,9142019: 16,6002020: 12,3322021: 13,0152022: 13,6052023: 14,6232024: 13,2802025: 11,905
2007high 16,600 (2019) · low 2,380 (2008)2025

MAUDE reports naming product code GDW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200785692,401
200875702,380
20091327224,014
2010103923,768
2011114993,826
201294511 (1)3,717
2013134913,777
2014156304,784
2015811164,899
2016119625,106
201713103311,643
2018790314,914
2019122054 (4)16,600
202010881112,332
20219233013,015
2022789013,605
20234267114,623
20242—013,280
20253111311,905
202631410—

What is it indicated for?

Frankenman Staplers are indicated as follows: Disposable Alimentary Canal Stapler The Frankenman Disposable Alimentary Canal Stapler is used throughout the alimentary tract for the creation of end-to-end, end-to-side and side-to-side anastomoses in both open and laparoscopic techniques. Single Use Circular Stapler for Rectal Prolapse and Hemorrhoid The Frankenman Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids is a Circular Stapler…

Read the full statement (167 words)

Frankenman Staplers are indicated as follows:

Disposable Alimentary Canal Stapler

The Frankenman Disposable Alimentary Canal Stapler is used throughout the alimentary tract for the creation of end-to-end, end-to-side and side-to-side anastomoses in both open and laparoscopic techniques.

Single Use Circular Stapler for Rectal Prolapse and Hemorrhoid

The Frankenman Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids is a Circular Stapler product, with accessories, that has application for general Surgical treatment of haemorrhoids and anorectal wall defects by means of transanal stapling and resection of mucosal and musculo-mucosal tissue resulting in occlusion of haemorrhoidal inflow, restoring the haemorrhoidal tissue to its normal physiological position.

Disposable Reloadable Linear Stapler and Reloads

The Frankenman Disposable Reloadable Linear Stapler (and Reloads) is used in the resection or transaction of tissue for abdominal, gynecological, pediatric and thoracic surgical procedures.

Disposable Reloadable Linear Cutter Stapler and Reloads

The Frankenman Disposable Reloadlable Linear Cutter Stapler (and Reloads) has application in abdominal, gynecological, thoracic and pediatric surgery transaction, resection, and the creation of anastomoses.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101378: Frankenman Surgical Staplers.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101378

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026