510(k) Report

510(k) · Cleared

Recalled · Class III · 2022

3M Comply Lead Free Process Indicators for Steam

K101528510(k) TraditionalCleared Oct 1, 2010121-day reviewClass II21 CFR 880.2800Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

3M Company · JOJ Indicator, Physical/Chemical Sterilization Process

Cleared on Oct 1, 2010 as substantially equivalent to K932129 (3M Company, 1994) after a 121-day review. Its family tree reaches back 1 generation to 1994. 5 later devices descend from it. It has been recalled once.

Decided
Oct 1, 2010
Review
121 days
Class
Class II
Product code
JOJ
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K932129 (1994). 5 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2024: 1 descendant cleared2019: 1 descendant cleared2022: 2 descendants cleared2025: 1 descendant clearedReference device: K093199 · 3M Comply Bowie- Dick Type Lead Free Test Pack · 3M Company · 2009K932129 · 3M Autoclave Tapes · 3M Company · 19941990200020102020
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 2 descendants K932129 · 3M Autoclave Tapes · 3M Company · 19940
  1. Root · 1 generation backK932129 3M Autoclave Tapes3M Company · 1994
  2. This deviceK101528 3M Comply Lead Free Process Indicators for Steam3M Company · 2010
  3. Descendant · cites it directly · 1 of its ownK220564 3M Comply Lead Free Steam Indicator Tape 1322, 3M Attest Lead Free…3M Company · 2022
  4. Descendant · cites it directly · 1 of its ownK181788 Lead-free Chemical Indicators for Steam SterilizationKem Medical Products Corp. · 2019
  5. Descendant · 2 generations on · 0 of its ownK251991 SafeSecure Steam Sterilization Process Indicator (4012, 4013, 4014…Shenzhen Safesecure Medical Infection Control Tech Co., Ltd. · 2025
  6. Descendant · cites it directly · 0 of its ownK241836 Kem Medical Lead-free Chemical Indicators for Steam SterilizationKem Medical Products Corporation · 2024
  7. Descendant · 2 generations on · 0 of its ownK220778 Chemical Indicator for SteamTrue Indicating, LLC · 2022

Most influential descendants

Showing 5 of 5 · See all
  • K2205643M Comply Lead Free Steam Indicator Tape 1322, 3M Attest Lead Free Steam Indicator Tape…3M Company · 2022 · cites it directly1descendant
  • K181788Lead-free Chemical Indicators for Steam SterilizationKem Medical Products Corp. · 2019 · cites it directly1descendant
  • K251991SafeSecure Steam Sterilization Process Indicator (4012, 4013, 4014, 4016); SafeSecure…Shenzhen Safesecure Medical Infection Control Tech Co., Ltd. · 2025 · 2 generations on0descendants
  • K241836Kem Medical Lead-free Chemical Indicators for Steam SterilizationKem Medical Products Corporation · 2024 · cites it directly0descendants
  • K220778Chemical Indicator for SteamTrue Indicating, LLC · 2022 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K9321293M Autoclave Tapes3M Company1994—
0 (this device)K1015283M Comply Lead Free Process Indicators for Steam3M Company2010—
+1K2205643M Comply Lead Free Steam Indicator Tape 1322, 3M Attest Lead Free Steam…3M Company2022—
+1K181788Lead-free Chemical Indicators for Steam SterilizationKem Medical Products Corp.2019—
+2K251991SafeSecure Steam Sterilization Process Indicator (4012, 4013, 4014, 4016)…Shenzhen Safesecure Medical Infection Control Tech Co., Ltd.2025—
+1K241836Kem Medical Lead-free Chemical Indicators for Steam SterilizationKem Medical Products Corporation2024—
+2K220778Chemical Indicator for SteamTrue Indicating, LLC2022—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202220
202410
202510
FIG. 1 — Lineage of K101528, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOJ device

  1. #1 K261992 3M™ Attest™ Steam Chemical Integrator (1243A…score 13.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K9321293M Autoclave Tapes
Summary page 2
…bmission is to: * Replace the lead ink in the current 3MTM CoMplyTm Autoclave Tape with Lead Free Ink. The current Comply Autoclave Tape was cleared under K932 129 * Apply the Lead Free Ink to cardstock as a new 3MTM CoMplyTm Record Card * Replace the lead ink in the process indic…
3M Company1994—
†K0931993M Comply Bowie- Dick Type Lead Free Test Pack3M Company2009—

Also cleared against K932129: 2 other devices →

What descends from it? (5)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K2205643M Comply Lead Free Steam Indicator Tape 1322, 3M Attest Lead Free Steam Indicator Tape…3M Company20221
K181788Lead-free Chemical Indicators for Steam SterilizationKem Medical Products Corp.20191
K241836Kem Medical Lead-free Chemical Indicators for Steam SterilizationKem Medical Products Corporation20240

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 25, 2022Class IIIZ-1794-2022Environmental controlFacility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K220564 (3M Company, 2022) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K932129.

How does its category compare?

All JOJ devices

JOJ Indicator, Physical/Chemical Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · JOJ

6in 2025

2007: 82008: 92009: 72010: 62011: 132012: 72013: 62014: 62015: 82016: 42017: 72018: 52019: 122020: 72021: 42022: 82023: 42024: 122025: 62026 (partial year): 3
2007high 13 (2011) · low 4 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOJ

90in 2025

2007: 140 days median2008: 57 days median2009: 188 days median2010: 115 days median2011: 151 days median2012: 224 days median2013: 212 days median2014: 198 days median2015: 188 days median2016: 150 days median2017: 172 days median2018: 150 days median2019: 38 days median2020: 108 days median2021: 87 days median2022: 90 days median2023: 53 days median2024: 71 days median2025: 90 days median2026 (partial year): 30 days median
2007high 224 (2012) · low 38 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JOJ

1in 2023

2008: 22009: 22011: 22021: 12023: 1
2007high 2 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOJ

12in 2025

2007: 12008: 12009: 92010: 32011: 12012: 82013: 62014: 112015: 72016: 112017: 352018: 122019: 82020: 62021: 292022: 232023: 392024: 252025: 12
2007high 39 (2023) · low 1 (2007)2025

MAUDE reports naming product code JOJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007814001
200895721
2009718829
2010611503
20111315121
2012722408
2013621206
20146198011
2015818807
20164150011
20177172035
20185150012
2019123808
2020710806
2021487129
2022890023
2023453139
20241271025
2025690012
20263300—

What is it indicated for?

The 3M Comply Lead Free Process Indicators for Steam are designed to demonstrate that the unit or load has been exposed to a steam sterilization process and to distinguish between processed and unprocessed units or loads. 3M Comply Lead Free Process Indicators for Steam include: * Comply 68200 Lead Free Record Card * Comply 1322 Lead Free Indicator Tape * Lead Free Test Pack Labels to be used on test packs including the following - o Comply 41380…

Read the full statement (217 words)

The 3M Comply Lead Free Process Indicators for Steam are designed to demonstrate that the unit or load has been exposed to a steam sterilization process and to distinguish between processed and unprocessed units or loads.

3M Comply Lead Free Process Indicators for Steam include:

* Comply 68200 Lead Free Record Card

* Comply 1322 Lead Free Indicator Tape

* Lead Free Test Pack Labels to be used on test packs including the following -

o Comply 41380 Steam-Plus Test Pack

o Comply 41360 Steam Test Pack

o Attest 1276 Steam Test Pack

o Attest 1296 Steam Test Pack

o Attest 41382 Steam-Plus Test Pack

o Comply 1233LF, 00132LF, 00135LF Bowie Dick Test Packs

Use the 3M Comply Lead Free Process Indicators for Steam in steam sterilization processes described below.

Cycle Type — Temperature — Exposure Time

Gravity — 250 °F/121 °C — > 30 minutes (wrapped)

Gravity — 270 °F/132 °C — > 3 minutes (unwrapped); > 15 minutes (wrapped)

Gravity — 275 °F/135 °C — > 3 minutes (unwrapped); > 11 minutes (wrapped)

Vacuum-assisted (prevacuum) — 270 °F/132 °C — > 3 minutes (unwrapped); > 4 minutes (wrapped)

Vacuum-assisted (prevacuum) — 273 °F/134 °C — > 3.5 minutes (unwrapped); 4 minutes (wrapped)

Vacuum-assisted (prevacuum) — 275 °F/135 °C — > 3 minutes (wrapped or unwrapped)

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101528: 3M Comply Lead Free Process Indicators for Steam.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K101528

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026