510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Chem Velocity Automated Urine Chemistry System, Strips and Calchek Kit

K101852510(k) TraditionalCleared Mar 23, 2011265-day reviewClass I21 CFR 862.2900Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Iris International Inc., A Division of Iris Iris I · KQO Automated Urinalysis System

Cleared on Mar 23, 2011 as substantially equivalent to K960054 (Intl. Remote Imaging Systems, 1996) and 1 other predicate after a 265-day review. Its family tree reaches back 3 generations to 1994. 3 later devices descend from it. It has been recalled 11 times.

Decided
Mar 23, 2011
Review
265 days
Class
Class I
Product code
KQO
Recalls
11
Predicates
2

Its longest line of descent runs back 3 generations to K943592 (1994). 5 ancestors in all. 3 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1995: 1 other ancestor2021: 1 descendant cleared2017: 1 descendant cleared2025: 1 descendant clearedK943592 · Chemstrip Mini UA Urine Analyzer · Boehringer Mannheim Corp. · 1994K983510 · Chemstrip 101 Urine Analyzer · Boehringer Mannheim Corp. · 1998K013783 · Aution Max AX-4280 Urinalysis System · International Remote Imaging Systems · 20021990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant K013783 · Aution Max AX-4280 Urinalysis System · International Remote Imaging Systems · 2002K983510 · Chemstrip 101 Urine Analyzer · Boehringer Mannheim Corp. · 1998K943592 · Chemstrip Mini UA Urine Analyzer · Boehringer Mannheim Corp. · 19940
  1. Root · 3 generations backK943592 Chemstrip Mini UA Urine AnalyzerBoehringer Mannheim Corp. · 1994
  2. 2 generations backK983510 Chemstrip 101 Urine AnalyzerBoehringer Mannheim Corp. · 1998
  3. 1 generation backK013783 Aution Max AX-4280 Urinalysis SystemInternational Remote Imaging Systems · 2002
  4. This deviceK101852 Chem Velocity Automated Urine Chemistry System, Strips and Calchek KitIris International Inc., A Division of Iris Iris I · 2011
  5. Descendant · cites it directly · 2 of its ownK171083 iChemVELOCITY Automated Urine Chemistry SystemIris Diagnostics, A Division of Iris International, Inc. · 2017
  6. Descendant · cites it directly · 1 of its ownK210127 iQ200 System, iChemVELOCITY Automated Urine Chemistry SystemBeckman Coulter, Inc. · 2021 · Recalled
  7. Descendant · 2 generations on · 0 of its ownK252580 iQ200 SeriesBeckman Coulter, Inc. · 2025 · Recalled

Most influential descendants

Showing 3 of 3 · See all
  • K171083iChemVELOCITY Automated Urine Chemistry SystemIris Diagnostics, A Division of Iris International, Inc. · 2017 · cites it directly2descendants
  • K210127iQ200 System, iChemVELOCITY Automated Urine Chemistry SystemBeckman Coulter, Inc. · 2021 · cites it directly1descendant
  • K252580iQ200 SeriesBeckman Coulter, Inc. · 2025 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K943592Chemstrip Mini UA Urine AnalyzerBoehringer Mannheim Corp.1994—
−2K983510Chemstrip 101 Urine AnalyzerBoehringer Mannheim Corp.1998—
−1K013783Aution Max AX-4280 Urinalysis SystemInternational Remote Imaging Systems2002—
0 (this device)K101852Chem Velocity Automated Urine Chemistry System, Strips and Calchek KitIris International Inc., A Division of Iris Iris I2011—
+1K171083iChemVELOCITY Automated Urine Chemistry SystemIris Diagnostics, A Division of Iris International, Inc.2017—
+1K210127iQ200 System, iChemVELOCITY Automated Urine Chemistry SystemBeckman Coulter, Inc.20216
+2K252580iQ200 SeriesBeckman Coulter, Inc.20252

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202110
202510
FIG. 1 — Lineage of K101852, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KQO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K960054900UDX Urine Pathology System
Summary page 6
…ssified: Urinary Color, Urinary Clarity (Turbidity) Device to which substantial equivalence is claimed IRISpec Gravity Control 1 and IRISpec Gravity Control 2Th File No. K960054 Arkray Check Strips K01 3783 Description of the proposed device: iChem VELOCITY CalChek Reagents are a set of buffer-based soluti…
Intl. Remote Imaging Systems1996—
*K013783Aution Max AX-4280 Urinalysis System
Summary page 6
…ary Clarity (Turbidity) Device to which substantial equivalence is claimed IRISpec Gravity Control 1 and IRISpec Gravity Control 2Th File No. K960054 Arkray Check Strips K01 3783 Description of the proposed device: iChem VELOCITY CalChek Reagents are a set of buffer-based solutions intended for the in vitro…
International Remote Imaging Systems2002—
*—iChem 100 Urine Chemistry Analyzer and iChem 10SG strips
Named without a submission number
—

What descends from it? (3)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171083iChemVELOCITY Automated Urine Chemistry SystemIris Diagnostics, A Division of Iris International, Inc.20172
K210127iQ200 System, iChemVELOCITY Automated Urine Chemistry SystemBeckman Coulter, Inc.20211

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 30, 2020Class IIZ-1944-2020Nonconforming Material/ComponentA defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal…
Feb 17, 2020Class IIZ-1535-2020Under Investigation by firmThe Leukocyte and Compensation pads were switched on some vials of lot number 7212269M of the iChem VELOCITY Strips.
Nov 7, 2019Class IIZ-0843-2020Error in labelingBeckman Coulter has become aware of an issue affecting Specific Gravity chemistry settings for the iChemVELOCITY Urine Chemistry System that prevents the trigger of a chemistry system flag S3 and may result in erroneous Specific Gravity…
Oct 28, 2019Class IIZ-0913-2020Software in the Use EnvironmentCustomers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics).
Oct 18, 2018Class IIZ-1937-2019Component change controlFalsely elevated (false positive) urobilinogen patient results due to an increase in the sensitivity of the urobilinogen reaction.
Mar 29, 2018Class IIZ-1894-2018Process controlThe Strip Provider Modules (SPM) on some of the instruments had not been inspected during manufacture for correct seal placement and integrity.
Mar 2, 2018Class IIZ-1643-2018Under Investigation by firmBeckman Coulter has determined that there is a potential for incorrect settings to be installed on North American iChemVELOCITY Urine Chemistry Analyzers. 1. The installation of International settings can result in the generation of…
Dec 4, 2017Class IIZ-1893-2018Mixed-up of materials/componentsA subset of Lot 7212154 A of the iChem VELOCITY Urine Chemistry strips has an incorrect pad placed in the location of the leukocytes pad. This will result in control CB failure for leukocytes and may generate false negative erroneous…
Jan 30, 2017Class IIZ-1384-2017Device DesignThe probe misalignment or bending has the potential to lead to delay in reporting results or incorrect patient results.
Jan 13, 2016Class IIZ-1067-2016Equipment maintenanceIris International noted loose or missing analyte pads found in several locations (strip vial, strip provider module, strip conveyor system or waste container) that are undetectable by the iChemVELOCITY System.
Jul 15, 2015Class IIZ-2441-2015Nonconforming Material/ComponentIris International is recalling the iChemVELOCITY systems because they do not maintain the on-board strip stability claim of 5 days at 18-28 degrees celsius at 20-80% relative humidity (RH) as stated in the iChemVELOCITY product labeling.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 3 companies; a new generation was cleared every 4.0 years on median.
  • None of the 5 ancestors has a recall.
  • Of its descendants, K171083 (Iris Diagnostics, A Division of Iris International, Inc., 2017) has gone furthest, with 2 descendants of its own.

How does its category compare?

All KQO devices

KQO Automated Urinalysis System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · KQO

2in 2015

2007: 12009: 12010: 22011: 22013: 12014: 12015: 2
2007high 2 (2010) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KQO

Too few KQO 510(k)s in any year since 2007 to give a median.

Recalls per year · KQO

1in 2021

2010: 72011: 22013: 12016: 12019: 22021: 12026 (partial year): 2
2007high 7 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KQO

12in 2025

2007: 12008: 42009: 92010: 32011: 82012: 152013: 922014: 1062015: 2382016: 3732017: 202018: 252019: 642020: 212021: 132022: 152023: 192024: 202025: 12
2007high 373 (2016) · low 1 (2007)2025

MAUDE reports naming product code KQO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—01
20080—04
20091—09
20102—73
20112—28
20120—015
20131—192
20141—0106
20152—0238
20160—1373
20170—020
20180—025
20190—264
20200—021
20210—113
20220—015
20230—019
20240—020
20250—012
20260—2—

What is it indicated for?

The iChem®VELOCITY™ automated urine chemistry system is an in vitro diagnostic device used to automate the urine chemistry analysis profile using iChem®VELOCITY™ Urine Chemistry Strips. The iChem VELOCITY can be used as a stand alone-system, as well as in an iQ&200 Series system, a configuration given the proprietary name iRICELL™ as it is designed to be hardware and software compatible with iQ200 Series systems. It produces quantitative results for…

Read the full statement (283 words)

The iChem®VELOCITY™ automated urine chemistry system is an in vitro diagnostic device used to automate the urine chemistry analysis profile using iChem®VELOCITY™ Urine Chemistry Strips. The iChem VELOCITY can be used as a stand alone-system, as well as in an iQ&200 Series system, a configuration given the proprietary name iRICELL™ as it is designed to be hardware and software compatible with iQ200 Series systems. It produces quantitative results for specific gravity, semi-quantitative results for glucose, blood, leukocyte esterase, bilirubin, urobilinogen, pH, protein, ketones and ascorbic acid; and qualitative results for nitrites, color and clarity. IChemVELOCITY strips are intended for use only with the iChem VELOCITY analyzer. In particular they are not intended for visual reading. The iChem VELOCITY test strips are not intended for visual reading. The iChemVELOCITY is not intended to be used as a Point of Care (POC) analyzer.

These measurements are used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver function. Tests performed using the iChem VELOCITY are intended for clinical laboratory use and in vitro diagnostic use only.

The iChern®VELOCITY™ CaiChek Kit is a set of assayed and unassayed in vitro diagnostic controls for monitoring performance of the iChem VELOCITY urine chemistry analyzer.

The controls include assayed liquid controls for the monitoring of the specific gravity measurement module, and unassayed liquid color and clarity controls for the monitoring of color and clarity measurements. The kit also includes a set of assayed reflectance strips for the monitoring of reflectance measurements.

These measurements monitored by these controls are part of an automated urine chemistry analyzer used to aid in the diagnosis of metabolic disorders, kidney function anomalies, urinary tract infections, and liver function.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101852: Chem Velocity Automated Urine Chemistry System, Strips and Calchek Kit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K101852

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026