Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · LKM Counter, Urine Particle

Cleared on Sep 10, 2025 as substantially equivalent to K022774 (Intl. Remote Imaging Systems, 2002) and 2 other predicates after a 26-day review. Its family tree reaches back 6 generations to 1994. It has been recalled 2 times.

Decided
Sep 10, 2025
Review
26 days
Class
Class II
Product code
LKM
Recalls
2
Predicates
3

Its longest line of descent runs back 6 generations to K943592 (1994). 18 ancestors in all, the oldest K914256 (1992). No later device cites it as a predicate yet.

  1. Root · 6 generations backK943592 Chemstrip Mini UA Urine AnalyzerBoehringer Mannheim Corp. · 1994
  2. 5 generations backK983510 Chemstrip 101 Urine AnalyzerBoehringer Mannheim Corp. · 1998
  3. 4 generations backK013783 Aution Max AX-4280 Urinalysis SystemInternational Remote Imaging Systems · 2002
  4. 3 generations backK101852 Chem Velocity Automated Urine Chemistry System, Strips and Calchek KitIris International Inc., A Division of Iris Iris I · 2011 · Recalled
  5. 2 generations backK171083 iChemVELOCITY Automated Urine Chemistry SystemIris Diagnostics, A Division of Iris International, Inc. · 2017
  6. 1 generation backK210127 iQ200 System, iChemVELOCITY Automated Urine Chemistry SystemBeckman Coulter, Inc. · 2021 · Recalled
  7. This deviceK252580 iQ200 SeriesBeckman Coulter, Inc. · 2025
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K943592Chemstrip Mini UA Urine AnalyzerBoehringer Mannheim Corp.1994—
−5K983510Chemstrip 101 Urine AnalyzerBoehringer Mannheim Corp.1998—
−4K013783Aution Max AX-4280 Urinalysis SystemInternational Remote Imaging Systems2002—
−3K101852Chem Velocity Automated Urine Chemistry System, Strips and Calchek KitIris International Inc., A Division of Iris Iris I201111
−2K171083iChemVELOCITY Automated Urine Chemistry SystemIris Diagnostics, A Division of Iris International, Inc.2017—
−1K210127iQ200 System, iChemVELOCITY Automated Urine Chemistry SystemBeckman Coulter, Inc.20216
0 (this device)K252580iQ200 SeriesBeckman Coulter, Inc.2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199210
199410
199510
199620
200210
200410
200510
200710
201030
FIG. 1 — Lineage of K252580, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LKM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LKM device

  1. #1 K232416 AUTION EYE AI-4510 Urine Particle Analysis…score 7.3 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022774IQ 200 System
Summary page 5
…cell counter Classification Name : Counter, Urine Particle Regulation Number : 864.5200 Product Code(s) : LKM, KQO III. LEGALLY MARKETED PREDICATE DEVICE Predicate # K022774, K093861, K210127 Predicate Trade Name : iQ200 System Product Code : LKM, KQO IV. DEVICE DESCRIPTION SUMMARY The iQ200 Series A…
Intl. Remote Imaging Systems200214
*K093861IQ 20 Series Analyzer, IQ 200 Series System, Nad IQ Lamina Cradle
Summary page 5
…nter Classification Name : Counter, Urine Particle Regulation Number : 864.5200 Product Code(s) : LKM, KQO III. LEGALLY MARKETED PREDICATE DEVICE Predicate # K022774, K093861, K210127 Predicate Trade Name : iQ200 System Product Code : LKM, KQO IV. DEVICE DESCRIPTION SUMMARY The iQ200 Series Automated…
Iris International, Inc.20104
*K210127iQ200 System, iChemVELOCITY Automated Urine Chemistry System
Summary page 5
…sification Name : Counter, Urine Particle Regulation Number : 864.5200 Product Code(s) : LKM, KQO III. LEGALLY MARKETED PREDICATE DEVICE Predicate # K022774, K093861, K210127 Predicate Trade Name : iQ200 System Product Code : LKM, KQO IV. DEVICE DESCRIPTION SUMMARY The iQ200 Series Automated Urine Mic…
Beckman Coulter, Inc.20216

Also cleared against K022774: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 12, 2026Class IIZ-0161-2027Component design/selectionThe rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance…
Aug 12, 2026Class IIZ-0160-2027Component design/selectionThe rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replacement was excluded from the annual preventive maintenance…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 5 companies; a new generation was cleared every 4.2 years on median.
  • 5 of the 18 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K022774.

How does its category compare?

All LKM devices

LKM Counter, Urine Particle: FDA decisions, review times, recalls and adverse event reports over 20 years. 2025, this device’s year, is highlighted.

FDA decisions per year · LKM

1in 2025

2007: 12008: 12009: 12010: 12018: 12020: 12021: 12024: 12025: 1
2007high 1 (2007) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LKM

Too few LKM 510(k)s in any year since 2007 to give a median.

Recalls per year · LKM

4in 2023

2013: 12019: 12023: 4
2007high 4 (2023) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LKM

1in 2021

2013: 32014: 12016: 32017: 12019: 22020: 22021: 1
2007high 3 (2013) · low 0 (2007)2025

MAUDE reports naming product code LKM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20081—00
20091—00
20101—00
20110—00
20120—00
20130—13
20140—01
20150—00
20160—03
20170—01
20181—00
20190—12
20201—02
20211—01
20220—00
20230—40
20241—00
20251—00
20260—0—

What is it indicated for?

The iQ200 System is an in-vitro diagnostic device used to automate the complete urinalysis profile, including urine test strip chemistry panel, and microscopic sediment analysis. Optionally, the iQ200 Analyzer can be used as a stand-alone unit, or the results from the iQ200 analyzer can be combined with other urine chemistry results received from an LIS. It produces quantitative or qualitative counts of all formed sediment elements present in urine, including cells, casts, crystals, and organisms. A competent human operator can set criteria for auto-reporting and flagging specimens for review. All instrument analyte image decisions may be reviewed and overridden by a trained technologist.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K252580: iQ200 Series.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K252580

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026