510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Endeavour Monitoring System

K102004510(k) AbbreviatedCleared Sep 10, 201057-day reviewClass II21 CFR 870.1025Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Mindray DS USA, Inc. · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Sep 10, 2010 as substantially equivalent to K062098 (Datascope Corp., 2007) and 1 other predicate after a 57-day review. Its family tree reaches back 7 generations to 1994. 3 later devices descend from it. It has been recalled 3 times.

Decided
Sep 10, 2010
Review
57 days
Class
Class II
Product code
MHX
Recalls
3
Predicates
2

Its longest line of descent runs back 7 generations to K933143 (1994). 34 ancestors in all, the oldest K842616 (1984). 3 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1999: 1 other ancestor2006: 1 other ancestor2003: 2 other ancestors2002: 2 other ancestors1997: 2 other ancestors2000: 2 other ancestors1998: 1 other ancestor1996: 3 other ancestors1995: 3 other ancestors1993: 1 other ancestor1990: 2 other ancestors1984: 1 other ancestor1994: 3 other ancestors1991: 2 other ancestors1989: 1 other ancestor2013: 1 descendant cleared2015: 1 descendant cleared2016: 1 descendant clearedK933143 · Model Eli 300 Electrocardiograph · Mortara Instrument, Inc. · 1994K962970 · Spacelabs Medical 12-Lead Diagnostic Ecg Analysis Option · Spacelabs Medical, Inc. · 1997K972502 · Spacelabs Medical Integrated Multiparameter Module 90496 · Spacelabs Medical, Inc. · 1998 · recalledK993531 · Passport 2 Vital Signs Monitor, Model 0998-00-0170-XX · Datascope Corp. · 2000 · recalledK020550 · Passport 2 with View 12 Ecg Analysis Module Model # 0998-00-0170-XXXX · Datascope Corp. · 2002 · recalledK031849 · Spectrum Patient Monitor, Model 0998-00-0500-Xxxxx · Datascope Corp. · 2003 · recalledK062098 · Spectrum Monitor, Model 0998-00-1000-Xxxxx · Datascope Corp. · 200719801990200020102020
  1. Root · 7 generations backK933143 Model Eli 300 ElectrocardiographMortara Instrument, Inc. · 1994
  2. 6 generations backK962970 Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc. · 1997
  3. 5 generations backK972502 Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc. · 1998 · Recalled
  4. 4 generations backK993531 Passport 2 Vital Signs Monitor, Model 0998-00-0170-XXDatascope Corp. · 2000 · Recalled
  5. 3 generations backK020550 Passport 2 with View 12 Ecg Analysis Module Model # 0998-00-0170-XXXXDatascope Corp. · 2002 · Recalled
  6. 2 generations backK031849 Spectrum Patient Monitor, Model 0998-00-0500-XxxxxDatascope Corp. · 2003 · Recalled
  7. 1 generation backK062098 Spectrum Monitor, Model 0998-00-1000-XxxxxDatascope Corp. · 2007
  8. This deviceK102004 Endeavour Monitoring SystemMindray DS USA, Inc. · 2010
  9. Descendant · cites it directly · 2 of its ownK132026 V Series Monitoring System, 12.1 Dcu/V12 and 21.0 Dcu/V21Mindray DS USA, Inc. · 2013 · Recalled
  10. Descendant · 2 generations on · 1 of its ownK150352 V Series Monitoring System (including V12 and V21 Monitors)Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2015
  11. Descendant · 3 generations on · 0 of its ownK153448 Passport Series Patient Monitors (Including Passport 8, Passport 12)Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2016 · Recalled

Most influential descendants

Showing 3 of 3 · See all
  • K132026V Series Monitoring System, 12.1 Dcu/V12 and 21.0 Dcu/V21Mindray DS USA, Inc. · 2013 · cites it directly2descendants
  • K150352V Series Monitoring System (including V12 and V21 Monitors)Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2015 · 2 generations on1descendant
  • K153448Passport Series Patient Monitors (Including Passport 8, Passport 12)Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2016 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K933143Model Eli 300 ElectrocardiographMortara Instrument, Inc.1994—
−6K962970Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc.1997—
−5K972502Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc.19983
−4K993531Passport 2 Vital Signs Monitor, Model 0998-00-0170-XXDatascope Corp.20001
−3K020550Passport 2 with View 12 Ecg Analysis Module Model # 0998-00-0170-XXXXDatascope Corp.20021
−2K031849Spectrum Patient Monitor, Model 0998-00-0500-XxxxxDatascope Corp.20031
−1K062098Spectrum Monitor, Model 0998-00-1000-XxxxxDatascope Corp.2007—
0 (this device)K102004Endeavour Monitoring SystemMindray DS USA, Inc.2010—
+1K132026V Series Monitoring System, 12.1 Dcu/V12 and 21.0 Dcu/V21Mindray DS USA, Inc.20131
+2K150352V Series Monitoring System (including V12 and V21 Monitors)Shenzhen Mindray Bio-Medical Electronics Co., Ltd.2015—
+3K153448Passport Series Patient Monitors (Including Passport 8, Passport 12)Shenzhen Mindray Bio-Medical Electronics Co., Ltd.20162

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198410
198910
199020
199120
199310
199430
199530
199630
199720
199810
199910
200020
200220
200320
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201510
201610
FIG. 1 — Lineage of K102004, 1994–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062098Spectrum Monitor, Model 0998-00-1000-Xxxxx
Summary page 1
…§870.1435- DXG - Single-function, pre-programmed diagnostic computer Predicate Devices: Spectrum Monitor - K062098 Accutorr Plus Monitor -K983575 Device description: The Endeavour Monitoring System is…
Datascope Corp.2007—
*K983575Accutorr Plus Non Invasive Blood Pressure Monitor, Model # 0998-00-0117-XX
Summary page 1
…computer Predicate Devices: Spectrum Monitor - K062098 Accutorr Plus Monitor -K983575 Device description: The Endeavour Monitoring System is a multi physiological paramete…
Datascope Corp.1999—

Also cleared against K062098: 1 other device →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K132026V Series Monitoring System, 12.1 Dcu/V12 and 21.0 Dcu/V21Mindray DS USA, Inc.20132

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 7, 2013Class IIZ-0389-2014Environmental controlMindray has identified an issue with V Series Monitor where the V Dock power pins connecting the V Series Monitor to the V Dock may lose spring tension, potentially causing the screen to temporarily go blank, lose communication with the…
Sep 24, 2012Class IIZ-0528-2013Nonconforming Material/ComponentMindray has identified an issue with the V Series Monitor where the monitors touch screen may stop responding to touch.
May 7, 2012Class IIZ-2279-2012Software designMindray has identified two software anomalies contained in the V Series Monitoring System. There have been no reports of injuries associated with these anomalies. These anomalies affect V Series Systems distributed between September 31…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 4 companies; a new generation was cleared every 2.6 years on median.
  • 10 of the 34 ancestors have a recall of any class.
  • Of its descendants, K132026 (Mindray DS USA, Inc., 2013) has gone furthest, with 2 descendants of its own.
  • 1 other device also cites its predicate K062098.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

The Endeavour Monitoring System is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The Indications for Use for the Endeavour include the monitoring of the following human physiological parameters: * EGG waveform derived from 3, 5, 6 and 12 lead measurements * Heart Rate * Pulse Oximetry (SPO2) * ST Segment Analysis * Arrhythmnia Detection * Non Invasive Blood Pressure (NIBP) * Invasive Blood Pressure…

Read the full statement (157 words)

The Endeavour Monitoring System is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The Indications for Use for the Endeavour include the monitoring of the following human physiological parameters:

* EGG waveform derived from 3, 5, 6 and 12 lead measurements

* Heart Rate

* Pulse Oximetry (SPO2)

* ST Segment Analysis

* Arrhythmnia Detection

* Non Invasive Blood Pressure (NIBP)

* Invasive Blood Pressure (1BP)

* Cardiac Output (GO)

* Respiratory Gasses

* Respiration Rate

* Temperature

The Endeavour Monitor has the capability of performing rV Drug and Hemodynamnic Calculations and interfacing with network devices.

The target populations are adult, pediatric and neonate with the exception of:

* Arrhythmia detection and ST Segment Analysis, for which the target populations are adult

and pediatric only,

* IV Drug Calculations for which the target population is adult only, and

a Cardiac Output for which the target population is adult and pediatric only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102004: Endeavour Monitoring System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K102004

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026