510(k) Report

510(k) · Cleared

Cleared

3M Integrated Cycler (110V) & (220V)

K102314510(k) TraditionalCleared Oct 29, 201074-day reviewClass II21 CFR 862.2570

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Focus Diagnostics, Inc. · OOI Real Time Nucleic Acid Amplification System

Cleared on Oct 29, 2010 as substantially equivalent to K062948 (Cepheid, 2006) after a 74-day review. Its family tree reaches back 4 generations to 1994.

Decided
Oct 29, 2010
Review
74 days
Class
Class II
Product code
OOI
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K936112 (1994). No later device cites it as a predicate yet.

  1. Root · 4 generations backK936112 Strep B OiaBiostar, Inc. · 1994
  2. 3 generations backK991828 Strep B OiaBiostar, Inc. · 1999
  3. 2 generations backK022504 Idi-Strep B AssayInfectio Diagnostic, Inc. · 2002
  4. 1 generation backK062948 Smart GBS and Smartcycle DX System and Software, Version 1.7BCepheid · 2006
  5. This deviceK102314 3M Integrated Cycler (110V) & (220V)Focus Diagnostics, Inc. · 2010
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K936112Strep B OiaBiostar, Inc.1994—
−3K991828Strep B OiaBiostar, Inc.1999—
−2K022504Idi-Strep B AssayInfectio Diagnostic, Inc.2002—
−1K062948Smart GBS and Smartcycle DX System and Software, Version 1.7BCepheid2006—
0 (this device)K1023143M Integrated Cycler (110V) & (220V)Focus Diagnostics, Inc.2010—
FIG. 1 — Lineage of K102314, 1994–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OOI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OOI device

  1. #1 K243922 Revogenescore 9.6 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K062948Smart GBS and Smartcycle DX System and Software, Version 1.7B
Summary page 1
…PnlProduct submite numerDate Code(s) SmartCycler Dx system Cepheid K062948 12/08/2006 (83) NJR (assay) I…
Cepheid2006—

Also cleared against K062948: 3 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 4 companies; a new generation was cleared every 4.1 years on median.
  • None of the 4 ancestors has a recall.
  • 3 other devices also cite its predicate K062948.

How does its category compare?

All OOI devices

OOI Real Time Nucleic Acid Amplification System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · OOI

1in 2025

2010: 32012: 12013: 12014: 12015: 22016: 22017: 12022: 12023: 12025: 1
2007high 3 (2010) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OOI

249in 2010

2010: 249 days median
2007high 249 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OOI

1in 2024

2010: 12011: 22013: 12021: 42022: 42023: 12024: 1
2007high 4 (2021) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OOI

524in 2025

2011: 52012: 32013: 12014: 52015: 12016: 32019: 22020: 1342021: 3242022: 8742023: 2202024: 3272025: 524
2007high 874 (2022) · low 0 (2007)2025

MAUDE reports naming product code OOI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
2010324910
20110—25
20121—03
20131—11
20141—05
20152—01
20162—03
20171—00
20180—00
20190—02
20200—0134
20210—4324
20221—4874
20231—1220
20240—1327
20251—0524
20260—0—

What is it indicated for?

The Integrated Cycler and accompanying Studio Software are intended for in vitro diagnostic use in conjunction with legally marketed SimplexaTM reagent kits and assay protocols labeled for in vitro diagnostic use. The Integrated Cycler is a rapid real-time Polymerase Chain Reaction (PCR) thermocycler used for identification of nucleic acid from prepared biological samples. The instrument utilizes disc media to contain and to process samples. The…

Read the full statement (113 words)

The Integrated Cycler and accompanying Studio Software are intended for in vitro diagnostic use

in conjunction with legally marketed SimplexaTM reagent kits and assay protocols labeled for in

vitro diagnostic use.

The Integrated Cycler is a rapid real-time Polymerase Chain Reaction (PCR) thermocycler used

for identification of nucleic acid from prepared biological samples. The instrument utilizes disc

media to contain and to process samples. The instrument uses real-time flourometric detection to

identify targets within the sample wells. The instrument's operation parameters are controlled by

the use of an external personal computer and associated software. This instrument is intended to be

used by laboratory professionals trained in laboratory techniques and in a laboratory environment.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102314: 3M Integrated Cycler (110V) & (220V).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102314

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026