510(k) · Cleared
Recalled · Class II · 2021ARIES MI System, ARIES HSV 1&2 Assay Cassettes-Carton of 24 (IVD), ARIES HSV 1&2 Assay Protocol File Kit (IVD)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Luminex Corporation · OOI Real Time Nucleic Acid Amplification System
Cleared on Jun 30, 2016 after a 29-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has been recalled 4 times.
- Decided
- Jun 30, 2016
- Review
- 29 days
- Class
- Class II
- Product code
- OOI
- Recalls
- 4
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in OOI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new OOI device
- #1 K243922 Revogenescore 9.6 · 3 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 17, 2021 | Class II | Z-0553-2022 | Nonconforming Material/Component | Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid. |
| Apr 16, 2021 | Class II | Z-0498-2022 | Equipment maintenance | Impacted lot may leak inside the ARIES instrument. |
| Feb 20, 2019 | Class II | Z-2283-2019 | Process control | Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could… |
| Jun 29, 2018 | Class II | Z-2292-2019 | Process control | An incorrect control board and firmware installed in some of the Hand Held Barcode Scanners could result in sample/patient mismatch or duplicate results (from other patients specimens) for a patient order. |
No ancestor or descendant has a Class I recall.
How does its category compare?
All OOI devicesOOI Real Time Nucleic Acid Amplification System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.
FDA decisions per year · OOI
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · OOI
249in 2010
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · OOI
1in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · OOI
524in 2025
MAUDE reports naming product code OOI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 0 |
| 2009 | 0 | — | 0 | 0 |
| 2010 | 3 | 249 | 1 | 0 |
| 2011 | 0 | — | 2 | 5 |
| 2012 | 1 | — | 0 | 3 |
| 2013 | 1 | — | 1 | 1 |
| 2014 | 1 | — | 0 | 5 |
| 2015 | 2 | — | 0 | 1 |
| 2016 | 2 | — | 0 | 3 |
| 2017 | 1 | — | 0 | 0 |
| 2018 | 0 | — | 0 | 0 |
| 2019 | 0 | — | 0 | 2 |
| 2020 | 0 | — | 0 | 134 |
| 2021 | 0 | — | 4 | 324 |
| 2022 | 1 | — | 4 | 874 |
| 2023 | 1 | — | 1 | 220 |
| 2024 | 0 | — | 1 | 327 |
| 2025 | 1 | — | 0 | 524 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Luminex® ARIES® M1 System is an in vitro diagnostic (IVD) platform that performs nucleic acid based tests in clinical laboratories. The Luminex ARIES® M1 System is capable of automated extraction and purification of nucleic acids from multiple sample types as well as the automated amplification and detection of target nucleic acid sequences by fluorescence-based PCR.
Indications for use as stated in the FDA 510(k) statement.
https://510kreport.com/device/K161495 · printed Nov 6, 2026
Cite this page
510(k) Report. “K161495: ARIES MI System, ARIES HSV 1&2 Assay Cassettes-Carton of 24 (IVD), ARIES HSV 1&2 Assay Protocol File Kit (IVD).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K161495
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026