Quadrox-I Neonatal Microporous Membrane Oxygenator with Integrated Arterial Filter with Softline/Bioline Coating

K102464510(k) TraditionalCleared Mar 4, 2011189-day reviewClass II21 CFR 870.4350Recalls 5 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Maquet Cardiopulmonary, AG · DTZ Oxygenator, Cardiopulmonary Bypass

Cleared on Mar 4, 2011 as substantially equivalent to K022115 (Terumo Corp., 2002) and 1 other predicate after a 189-day review. Its family tree reaches back 2 generations to 1985. 1 of the 11 devices in its family, counting itself, has a Class I recall. 6 later devices descend from it. It has been recalled 5 times, including 2 Class I recalls.

Decided
Mar 4, 2011
Review
189 days
Class
Class II
Product code
DTZ
Recalls
5 (2 Class I)
Predicates
2

Its longest line of descent runs back 2 generations to K842908 (1985). 4 ancestors in all. 6 later devices descend from it, 2 citing it directly. 1 of the 11 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor1994: 1 other ancestor2011: 1 descendant cleared2013: 2 descendants cleared2015: 1 descendant cleared2021: 1 descendant cleared2026: 1 descendant clearedK842908 · Variable Prime Cobe Membrane Lung · Cobe Laboratories, Inc. · 1985K022115 · Capiox RX05 Hollow Fiber Oxygenator W/Without Hardshell Reservoir · Terumo Corp. · 20021990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant K022115 · Capiox RX05 Hollow Fiber Oxygenator W/Without Hardshell Reservoir · Terumo Corp. · 2002K842908 · Variable Prime Cobe Membrane Lung · Cobe Laboratories, Inc. · 19850
  1. Root · 2 generations backK842908 Variable Prime Cobe Membrane LungCobe Laboratories, Inc. · 1985
  2. 1 generation backK022115 Capiox RX05 Hollow Fiber Oxygenator W/Without Hardshell ReservoirTerumo Corp. · 2002
  3. This deviceK102464 Quadrox-I Neonatal Microporous Membrane Oxygenator with Integrated…Maquet Cardiopulmonary, AG · 2011
  4. Descendant · cites it directly · 4 of its ownK132829 Quadrox-I, Quadrox-ID, Quadrox-IR OxygenatorsMaquet Cardiopulmonary, AG · 2013 · Recalled
  5. Descendant · 2 generations on · 2 of its ownK150267 QUADROX-i Adult/Small Adult Oxygenators;QUADROX-iD Adult OxygenatorsMaquet Cardiopulmonary, AG · 2015 · Recalled
  6. Descendant · 3 generations on · 1 of its ownK203424 Quantum Perfusion Blood Oxygenator ECC VT200-E1, Quantum Perfusion…Qura S.R.L · 2021
  7. Descendant · 4 generations on · 0 of its ownK240934 Quantum Perfusion Blood Oxygenator ECC VT160-E3 (VT160-E3); Quantum…Spectrum Medical S.R.L. · 2026
  8. Descendant · 2 generations on · 0 of its ownK133265 Quadrox-IR Adult and Small AdultMaquet Cardiopulmonary, AG · 2013

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2011: 12013: 22015: 12021: 12026: 1
2011high 2 (2013) · low 0 (2012)2026

Most influential descendants

Showing 5 of 6 · See all
  • K132829Quadrox-I, Quadrox-ID, Quadrox-IR OxygenatorsMaquet Cardiopulmonary, AG · 2013 · cites it directly4descendants
  • K150267QUADROX-i Adult/Small Adult Oxygenators;QUADROX-iD Adult OxygenatorsMaquet Cardiopulmonary, AG · 2015 · 2 generations on2descendants
  • K203424Quantum Perfusion Blood Oxygenator ECC VT200-E1, Quantum Perfusion Blood Oxygenator ECC…Qura S.R.L · 2021 · 3 generations on1descendant
  • K240934Quantum Perfusion Blood Oxygenator ECC VT160-E3 (VT160-E3); Quantum Perfusion Blood…Spectrum Medical S.R.L. · 2026 · 4 generations on0descendants
  • K133265Quadrox-IR Adult and Small AdultMaquet Cardiopulmonary, AG · 2013 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K842908Variable Prime Cobe Membrane LungCobe Laboratories, Inc.1985—
−1K022115Capiox RX05 Hollow Fiber Oxygenator W/Without Hardshell ReservoirTerumo Corp.2002—
0 (this device)K102464Quadrox-I Neonatal Microporous Membrane Oxygenator with Integrated Arterial…Maquet Cardiopulmonary, AG2011—
+1K132829Quadrox-I, Quadrox-ID, Quadrox-IR OxygenatorsMaquet Cardiopulmonary, AG20131
+2K150267QUADROX-i Adult/Small Adult Oxygenators;QUADROX-iD Adult OxygenatorsMaquet Cardiopulmonary, AG20152
+3K203424Quantum Perfusion Blood Oxygenator ECC VT200-E1, Quantum Perfusion Blood…Qura S.R.L2021—
+4K240934Quantum Perfusion Blood Oxygenator ECC VT160-E3 (VT160-E3); Quantum Perfusion…Spectrum Medical S.R.L.2026—
+2K133265Quadrox-IR Adult and Small AdultMaquet Cardiopulmonary, AG2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201320
201510
202110
202610
FIG. 1 — Lineage of K102464, 1985–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTZ device

  1. #1 K240934 Quantum Perfusion Blood Oxygenator ECC…score 14.3 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K022115Capiox RX05 Hollow Fiber Oxygenator W/Without Hardshell ReservoirTerumo Corp.2002—
*K943917Capiox Arterial Filter
Summary page 1
…low Fiber Oxygenator, Terumo Cardiovascular Systems Corp., cleared under K0221,15, Capiox Arterial Filtet, CX *AFO2, Terurno Cardiovascular Systems Corp*., cleared Under K943917. QUJADROX-i Adult microporous membrane oxygenator with and without integrated arterial filter with SOETLINE coating, MAQUET Cardi…
Terumo Medical Corp.1995—

Also cleared against K022115: 7 other devices →

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K132829Quadrox-I, Quadrox-ID, Quadrox-IR OxygenatorsMaquet Cardiopulmonary, AG20134
K112360Quadrox-I Microporporous Membrane Oxgenator Series and Qudrox-ID Diffusion Membrane…Maquet Cardiopulmonary, AG20110

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 18, 2023Class IIZ-1640-2023Pending1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile…
May 18, 2023Class IIZ-1660-2023Under Investigation by firmProduct packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.
Oct 25, 2018Class IZ-0436-2019Nonconforming Material/ComponentThere is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.
Oct 25, 2018Class IZ-0437-2019Nonconforming Material/ComponentThere is a potential for blood leakage at the blood inlet connector when using the reducing screw adapter 1/4 x 3/16 that may result in clinically relevant loss of prime before patient connection or blood loss during patient connection.
Feb 23, 2016Class IIZ-1371-2016Unknown/Undetermined by firmMaquet Cardiopulmonary (MCP) is conducting a voluntary recall based on the potential risk of exposure to endotoxin for various products.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 3 companies; a new generation was cleared every 17.7 years on median.
  • None of the 4 ancestors has a recall.
  • Of its descendants, K132829 (Maquet Cardiopulmonary, AG, 2013) has gone furthest, with 4 descendants of its own.
  • 7 other devices also cite its predicate K022115.

How does its category compare?

All DTZ devices

DTZ Oxygenator, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · DTZ

2in 2025

2007: 42008: 32009: 92010: 22011: 42012: 92013: 92014: 52015: 92016: 32017: 22018: 32019: 52020: 52021: 52022: 22023: 52024: 32025: 22026 (partial year): 3
2007high 9 (2009) · low 2 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTZ

124in 2024

2007: 55 days median2008: 88 days median2009: 210 days median2011: 160 days median2012: 90 days median2013: 28 days median2014: 28 days median2015: 134 days median2016: 62 days median2018: 72 days median2019: 90 days median2020: 93 days median2021: 336 days median2023: 145 days median2024: 124 days median2026 (partial year): 137 days median
2007high 336 (2021) · low 28 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTZ

1in 2025

2007: 72008: 52010: 12011: 132012: 142014: 22016: 12018: 2, 2 Class I2019: 12020: 32021: 1, 1 Class I2022: 62025: 1
2007high 14 (2012) · low 0 (2009)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · DTZ

976in 2025

2007: 1342008: 1052009: 522010: 1192011: 1902012: 2152013: 3462014: 7642015: 6582016: 4432017: 3882018: 3202019: 4892020: 4422021: 5952022: 5592023: 7382024: 5352025: 976
2007high 976 (2025) · low 52 (2009)2025

MAUDE reports naming product code DTZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074557134
20083885105
20099210052
20102—1119
2011416013190
201299014215
20139280346
20145282764
201591340658
20163621443
20172—0388
20183722 (2)320
20195901489
20205933442
202153361 (1)595
20222—6559
202351450738
202431240535
20252—1976
202631370—

What is it indicated for?

The membrane oxygenator QUADROX-i Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. It is designed for a blood flow rate of 0.2-1.5 l/min and is intended for the treatment of pediatric (neonate and infant) patients. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide and regulates the blood temperature. The QUADROX-i Neonatal (HMO 11000) model with…

Read the full statement (105 words)

The membrane oxygenator QUADROX-i Neonatal is intended for use in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. It is designed for a blood flow rate of 0.2-1.5 l/min and is intended for the treatment of pediatric (neonate and infant) patients. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide and regulates the blood temperature.

The QUADROX-i Neonatal (HMO 11000) model with integrated arterial filter also filters air bubbles and particles larger than 33 pm.

The utilization period of this device is restricted to six hours.

Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102464: Quadrox-I Neonatal Microporous Membrane Oxygenator with Integrated Arterial Filter with Softline/Bioline Coating.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102464

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026