510(k) Report

510(k) · Cleared

Recalled · Class III · 2018

Emit II PLUS6-Acetylmorphine Assay; Emit II Plus 6-AM/ Ecstasy Calibrator/ Control Level 1; Emit II Plus 6-AM / Ectasy C

K102779510(k) TraditionalCleared Mar 18, 2011175-day reviewClass II21 CFR 862.3650Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Healthcare Diagnostics, Inc. · DJG Enzyme Immunoassay, Opiates

Cleared on Mar 18, 2011 as substantially equivalent to K001178 (Microgenics Corp., 2000) and 1 other predicate after a 175-day review. Its family tree reaches back 3 generations to 2001. 1 later device descends from it. It has been recalled 2 times.

Decided
Mar 18, 2011
Review
175 days
Class
Class II
Product code
DJG
Recalls
2
Predicates
2

Its longest line of descent runs back 3 generations to K010496 (2001). 6 ancestors in all, the oldest K983159 (1998). 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor2001: 1 other ancestor1998: 1 other ancestor2015: 1 descendant cleared×2: 2 generations cleared close together, K010496 (2001), K012110 (2001) K043028 · Emit II Plus Ecstasy Assay · Dade Behring, Inc. · 20051995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 1 other ancestor 1 generation forward: 1 descendant K043028 · Emit II Plus Ecstasy Assay · Dade Behring, Inc. · 2005K012110 · Dri Ecstasy Enzyme Immunoassay · Microgenics Corp. · 2001K010496 · Cedia Dau Amphetamines/Ecstasy Assay · Microgenics Corp. · 20010
  1. Root · 3 generations backK010496 Cedia Dau Amphetamines/Ecstasy AssayMicrogenics Corp. · 2001
  2. 2 generations backK012110 Dri Ecstasy Enzyme ImmunoassayMicrogenics Corp. · 2001
  3. 1 generation backK043028 Emit II Plus Ecstasy AssayDade Behring, Inc. · 2005
  4. This deviceK102779 Emit II PLUS6-Acetylmorphine Assay; Emit II Plus 6-AM/ Ecstasy…Siemens Healthcare Diagnostics, Inc. · 2011
  5. Descendant · cites it directly · 0 of its ownK150275 Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay, Immunalysis…Immunalysis Corporation · 2015

Most influential descendants

Showing 1 of 1 · See all
  • K150275Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay, Immunalysis 6-Acetylmorphine…Immunalysis Corporation · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K010496Cedia Dau Amphetamines/Ecstasy AssayMicrogenics Corp.2001—
−2K012110Dri Ecstasy Enzyme ImmunoassayMicrogenics Corp.2001—
−1K043028Emit II Plus Ecstasy AssayDade Behring, Inc.2005—
0 (this device)K102779Emit II PLUS6-Acetylmorphine Assay; Emit II Plus 6-AM/ Ecstasy Calibrator/…Siemens Healthcare Diagnostics, Inc.2011—
+1K150275Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay, Immunalysis…Immunalysis Corporation2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
200010
200110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
FIG. 1 — Lineage of K102779, 2001–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DJG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DJG device

  1. #1 K232659 Dochek® Multi-Drug Urine Test Cup Rx, Dochek®…score 17.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K001178Cedia Dau 6-Acetylmorphine Assay
Summary page 4
…Proposed Device Acetylmorphine Assay Feature Emit® HI Plus 6-Acetylmorphine Assay (k001178) The Emit® 11 Plus 6-Acetylmorphine Assay The CEDIA® Heroin Metabolite…
Microgenics Corp.2000—
*K043028Emit II Plus Ecstasy Assay
Summary page 2
…Calibrator The Emit® II Plus 6-AM / Ecstasy Calibrators / Controls are substantially equivalent to the Emit® 11 Plus Ecstasy Calibrators / Controls cleared under k043028. Control The Emit®W 11 Plus 6-AM / Ecstasy Calibrators / Controls are substantially equivalent to the Emit® 11 Plus E…
Dade Behring, Inc.2005—

Also cleared against K001178: 3 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K150275Immunalysis 6-Acetylmorphine Urine Enzyme Immunoassay, Immunalysis 6-Acetylmorphine…Immunalysis Corporation20150

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 12, 2018Class IIIZ-1800-2018Under Investigation by firmIncorrect calibrator level listed in qualitative calibration steps
Mar 12, 2018Class IIIZ-1799-2018Under Investigation by firmIncorrect calibrator level listed in qualitative calibration steps

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 3 companies; a new generation was cleared every 3.4 years on median.
  • None of the 6 ancestors has a recall.
  • Of its descendants, K150275 (Immunalysis Corporation, 2015) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K001178.

How does its category compare?

All DJG devices

DJG Enzyme Immunoassay, Opiates: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · DJG

3in 2025

2007: 22008: 22009: 22010: 62011: 62012: 52013: 52014: 92015: 122016: 22017: 42018: 52019: 32020: 52021: 32022: 22023: 102024: 52025: 32026 (partial year): 3
2007high 12 (2015) · low 2 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DJG

31in 2025

2010: 187 days median2011: 233 days median2012: 95 days median2013: 60 days median2014: 62 days median2015: 141 days median2017: 101 days median2018: 134 days median2019: 235 days median2020: 30 days median2021: 422 days median2023: 140 days median2024: 150 days median2025: 31 days median2026 (partial year): 157 days median
2007high 422 (2021) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DJG

2in 2025

2011: 12019: 12025: 2
2007high 2 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DJG

1in 2025

2011: 42012: 12013: 42015: 32016: 22017: 12018: 12021: 22022: 12024: 32025: 1
2007high 4 (2011) · low 0 (2007)2025

MAUDE reports naming product code DJG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20082—00
20092—00
2010618700
2011623314
201259501
201356004
201496200
20151214103
20162—02
2017410101
2018513401
2019323510
202053000
2021342202
20222—01
20231014000
2024515003
202533121
202631570—

What is it indicated for?

The Emit® 11 Plus 6-Acetylmorphine Assay is a homogeneous enzyme immunoassay with 10 ng/mL cutoff. The assay is intended for use in laboratories for the qualitative and/or semiquantitative analyses of 6-acetylmorphine (6-AM), a heroin metabolite, in human urine. Emit® 11 Plus assays are designed for use with a number of chemistry analyzers. Semiquantitative test results may be used to assess assay performance as part of a quality control program and to…

Read the full statement (185 words)

The Emit® 11 Plus 6-Acetylmorphine Assay is a homogeneous enzyme immunoassay with 10 ng/mL cutoff. The assay is intended for use in laboratories for the qualitative and/or semiquantitative analyses of 6-acetylmorphine (6-AM), a heroin metabolite, in human urine. Emit® 11 Plus assays are designed for use with a number of chemistry analyzers.

Semiquantitative test results may be used to assess assay performance as part of a quality control program and to estimate a dilution of the specimen for confirmation by GC/MS.

The Emit® 11 Plus 6-Acetylmorphine Assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Calibrators are used in the calibration of the Emit® 11 Plus 6-Acetylmorphine and Emit® 11 Plus Ecstasy Assays.

Controls may be used as quality control materials based on the specific Emit® 11 Plus 6-Acetylmorphine and Emit® 11 Plus Ecstasy Assay cutoffs.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102779: Emit II PLUS6-Acetylmorphine Assay; Emit II Plus 6-AM/ Ecstasy Calibrator/ Control Level 1; Emit II Plus 6-AM / Ectasy C.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102779

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026