Recalled · Class II · 2015

Camino 110-4B Olm Intracranial Pressure Mnitoring Kit; Camino 110-4G Post Craniotomy Subdural Pressure Monitoring Kit; C

K102875510(k) TraditionalCleared Dec 20, 201080-day reviewClass II21 CFR 882.1620Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Integra LifeSciences Corporation · GWM Device, Monitoring, Intracranial Pressure

Cleared on Dec 20, 2010 as substantially equivalent to K853864 (Camino Laboratories, Inc., 1986) and 2 other predicates after an 80-day review. Its family tree reaches back 2 generations to 1982. 2 later devices descend from it. It has been recalled 3 times.

Decided
Dec 20, 2010
Review
80 days
Class
Class II
Product code
GWM
Recalls
3
Predicates
3

Its longest line of descent runs back 2 generations to K813459 (1982). 7 ancestors in all. 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor1986: 1 other ancestor1992: 1 other ancestor1989: 1 other ancestor1984: 1 other ancestor2017: 1 descendant cleared2026: 1 descendant clearedK813459 · Reusable & Disposable Temp. Probes · Electromedics, Inc. · 1982K962928 · Combined Intracranial Pressure-Temperature Sensing System (Icp/) · Camino Neurocare, Inc. · 1996 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant K962928 · Combined Intracranial Pressure-Temperature Sensing System (Icp/) · Camino Neurocare, Inc. · 1996 · recalledK813459 · Reusable & Disposable Temp. Probes · Electromedics, Inc. · 19820
  1. Root · 2 generations backK813459 Reusable & Disposable Temp. ProbesElectromedics, Inc. · 1982
  2. 1 generation backK962928 Combined Intracranial Pressure-Temperature Sensing System (Icp/)Camino Neurocare, Inc. · 1996 · Recalled
  3. This deviceK102875 Camino 110-4B Olm Intracranial Pressure Mnitoring Kit; Camino 110-4G…Integra LifeSciences Corporation · 2010
  4. Descendant · cites it directly · 1 of its ownK162108 Pressio 2 ICP Monitoring SystemSophysa SA · 2017 · Recalled
  5. Descendant · 2 generations on · 0 of its ownK252241 Pressio 3 Multi-parameter Neuromonitoring SystemSophysa · 2026

Most influential descendants

Showing 2 of 2 · See all
  • K162108Pressio 2 ICP Monitoring SystemSophysa SA · 2017 · cites it directly1descendant
  • K252241Pressio 3 Multi-parameter Neuromonitoring SystemSophysa · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K813459Reusable & Disposable Temp. ProbesElectromedics, Inc.1982—
−1K962928Combined Intracranial Pressure-Temperature Sensing System (Icp/)Camino Neurocare, Inc.19962
0 (this device)K102875Camino 110-4B Olm Intracranial Pressure Mnitoring Kit; Camino 110-4G Post…Integra LifeSciences Corporation2010—
+1K162108Pressio 2 ICP Monitoring SystemSophysa SA20173
+2K252241Pressio 3 Multi-parameter Neuromonitoring SystemSophysa2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198410
198610
198910
199210
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202610
FIG. 1 — Lineage of K102875, 1982–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GWM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GWM device

  1. #1 K252241 Pressio 3 Multi-parameter Neuromonitoring Systemscore 13.0 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K853864Intracranial Pressure Monitoring Kit 070
Summary page 1
…acranial Pressure Device Class: Class 11, under 21 CFR 882.1620 Classification Panel: Neurology Product Code: GWM Predicate Devices: K853864 Intracranial Pressure Monitoring Kit, Model 070 (Camino I10-413 and Camino 11l0-4G)…
Camino Laboratories, Inc.1986—
*K962928Combined Intracranial Pressure-Temperature Sensing System (Icp/)
Summary page 1
…Predicate Devices: K853864 Intracranial Pressure Monitoring Kit, Model 070 (Camino I10-413 and Camino 11l0-4G) K962928 Combined Intracranial Pressure-Temperature Sensing (Camino I110-413T) K022553…
Camino Neurocare, Inc.19962
*K022553Camino Intracranial Pressure Monitoring Catheter with Licox Imc Bolt Fitting, Model 110-4L
Summary page 1
…Camino 11l0-4G) K962928 Combined Intracranial Pressure-Temperature Sensing (Camino I110-413T) K022553 Camino® Intracranial Pressure Monitoring Catheter with LICO)? 1MGBolt Fitting, Model I110-4L (Camino 1…
Integra Neurosciences2002—

Also cleared against K853864: 2 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K162108Pressio 2 ICP Monitoring SystemSophysa SA20171

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 27, 2015Class IIZ-1815-2015Under Investigation by firmIntegra LifeSciences is recalling the Camino Intracranial Pressure Monitoring Catheters because they may have been incorrectly assembled. Specifically, the o-ring in the bolt adapter may be in the wrong location and as a result, there is…
Jul 9, 2014Class IIZ-2072-2014No Marketing ApplicationIntegra LifeSciences is recalling the Camino Intracranial Pressure Monitoring Kits because of revised labelling to indicate they are unsafe in an MR environment.
Sep 10, 2013Class IIZ-0117-2014Under Investigation by firmIntegra LifeSciences has identified through an internal evaluation that in order to comply with US FDA requirements, certain Camino Intracranial Pressure Monitoring Kits require additional labelling to indicate they are UNSAFE in an MR…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 28 years and passes through 3 companies; a new generation was cleared every 14.7 years on median.
  • 2 of the 7 ancestors have a recall of any class.
  • Of its descendants, K162108 (Sophysa SA, 2017) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K853864.

How does its category compare?

All GWM devices

GWM Device, Monitoring, Intracranial Pressure: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · GWM

1in 2025

2007: 22008: 22009: 12010: 22011: 32012: 42013: 22014: 22016: 32017: 32018: 12019: 22020: 32021: 22024: 22025: 12026 (partial year): 1
2007high 4 (2012) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GWM

30in 2020

2011: 123 days median2012: 100 days median2016: 192 days median2017: 37 days median2020: 30 days median
2007high 192 (2016) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GWM

3in 2025

2008: 12010: 92011: 62012: 12013: 122014: 22019: 12023: 1, 1 Class I2024: 4, 2 Class I2025: 3
2007high 12 (2013) · low 0 (2007)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · GWM

441in 2025

2007: 1262008: 1052009: 1172010: 1042011: 1842012: 1512013: 1202014: 3602015: 4862016: 4192017: 4452018: 3022019: 2602020: 2682021: 2112022: 4242023: 7442024: 6412025: 441
2007high 744 (2023) · low 104 (2010)2025

MAUDE reports naming product code GWM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—0126
20082—1105
20091—0117
20102—9104
201131236184
201241001151
20132—12120
20142—2360
20150—0486
201631920419
20173370445
20181—0302
20192—1260
20203300268
20212—0211
20220—0424
20230—1 (1)744
20242—4 (2)641
20251—3441
20261—0—

What is it indicated for?

The use of the OLM Intracranial Pressure Monitoring Kit by a qualified neurosurgeon is indicated when direct measurement of intracranial pressure in the parenchyma or the subarachnoid space, is clinically important. The use of the Post Craniotomy Subdural Pressure Monitoring Catheter by a qualified neurosurgeon is indicated when direct pressure measurement in the subdural space, post craniotomy, is clinically important. The Camino Intracranial Pressure…

Read the full statement (147 words)

The use of the OLM Intracranial Pressure Monitoring Kit by a qualified neurosurgeon is indicated when direct measurement of intracranial pressure in the parenchyma or the subarachnoid space, is clinically important.

The use of the Post Craniotomy Subdural Pressure Monitoring Catheter by a qualified neurosurgeon is indicated when direct pressure measurement in the subdural space, post craniotomy, is clinically important.

The Camino Intracranial Pressure Temperature Monitoring Kit is indicated for use by qualified neurosurgeons for measurement of intracranial pressure and temperature in the parenchyma.

The Camino® Intracranial Pressure Monitoring Catheter with Licox IMC Bolt Fitting, Model 110-4L is indicated for use by a qualified Neurosurgeon when direct measurement of intracranial pressure in the parenchyma or subarachnoid space is clinically important. The Camino® Intracranial Pressure Monitoring Catheter with Licox IMC Bolt Fitting, Model 110-4L is intended to be used only through a Licox Brain Oxygen Monitoring System Bolt.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102875: Camino 110-4B Olm Intracranial Pressure Mnitoring Kit; Camino 110-4G Post Craniotomy Subdural Pressure Monitoring Kit; C.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K102875

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026