510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Homechoice/Homechoice Pro Automated Personal Cycler Peritoneal Dialysis System Model 5C4471, 5C8310, 5C4471R and 5C8310R

K102936510(k) TraditionalCleared Mar 30, 2011177-day reviewClass II21 CFR 876.5630Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Baxter Healthcare Corp · FKX System, Peritoneal, Automatic Delivery

Cleared on Mar 30, 2011 as substantially equivalent to K053512 (Baxter Healthcare Corp, 2006) and 2 other predicates after a 177-day review. Its family tree reaches back 3 generations to 1994. 1 later device descends from it. It has been recalled 11 times.

Decided
Mar 30, 2011
Review
177 days
Class
Class II
Product code
FKX
Recalls
11
Predicates
3

Its longest line of descent runs back 3 generations to K923065 (1994). 1 later device descends from it, 1 citing it directly.

  1. Root · 3 generations backK923065 Baxter Personal Cycler Autommated Peritoneal DialBaxter Healthcare Corp · 1994 · Recalled
  2. 2 generations backK012988 Homechoice Personal Cycler Peritoneal Dialysis System, Models…Baxter Healthcare Corp · 2001 · Recalled
  3. 1 generation backK053512 Homechoice/Homechoice Pro Personal Cycler Peritoneal Dialysis…Baxter Healthcare Corp · 2006 · Recalled
  4. This deviceK102936 Homechoice/Homechoice Pro Automated Personal Cycler Peritoneal…Baxter Healthcare Corp · 2011
  5. Descendant · cites it directly · 0 of its ownK201867 Homechoice Claria APD System, Sharesource Connectivity Platform for…Baxter Healthcare Corporation · 2020 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K201867Homechoice Claria APD System, Sharesource Connectivity Platform for Use with the…Baxter Healthcare Corporation · 2020 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K923065Baxter Personal Cycler Autommated Peritoneal DialBaxter Healthcare Corp19944
−2K012988Homechoice Personal Cycler Peritoneal Dialysis System, Models 5C8310, 5C8302…Baxter Healthcare Corp20019
−1K053512Homechoice/Homechoice Pro Personal Cycler Peritoneal Dialysis System, Models…Baxter Healthcare Corp20063
0 (this device)K102936Homechoice/Homechoice Pro Automated Personal Cycler Peritoneal Dialysis System…Baxter Healthcare Corp2011—
+1K201867Homechoice Claria APD System, Sharesource Connectivity Platform for Use with…Baxter Healthcare Corporation20202

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
FIG. 1 — Lineage of K102936, 1994–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FKX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FKX device

  1. #1 K243371 Byonyks X-1 APD Cycler; Byonyks Automated PD…score 11.1 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K053512Homechoice/Homechoice Pro Personal Cycler Peritoneal Dialysis System, Models 5C4471 and…
Summary page 1
…l Cycler COMMON NAME Automated Peritoneal Dialysis (APD) Cycler CLASSIFICATION NAME Peritoneal dialysis system and accessories per 21 CFR 876.5630 PREDICATE DEVICE K0535 12 HomeChoiee/HomeChoice PRO Automated Personal Cycler K012988 HomeChoice/HomeChoice PRO Automated Perso…
Baxter Healthcare Corp20063
*K012988Homechoice Personal Cycler Peritoneal Dialysis System, Models 5C8310, 5C8302, 5C4471…
Summary page 1
…CATION NAME Peritoneal dialysis system and accessories per 21 CFR 876.5630 PREDICATE DEVICE K0535 12 HomeChoiee/HomeChoice PRO Automated Personal Cycler K012988 HomeChoice/HomeChoice PRO Automated Personal Cycler K923065 HomeChoice/HomeChoice PRO Automated Pers…
Baxter Healthcare Corp20019
*K923065Baxter Personal Cycler Autommated Peritoneal Dial
Summary page 1
…PREDICATE DEVICE K0535 12 HomeChoiee/HomeChoice PRO Automated Personal Cycler K012988 HomeChoice/HomeChoice PRO Automated Personal Cycler K923065 HomeChoice/HomeChoice PRO Automated Personal Cycler…
Baxter Healthcare Corp19944

Also cleared against K053512: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K201867Homechoice Claria APD System, Sharesource Connectivity Platform for Use with the…Baxter Healthcare Corporation20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 10, 2018Class IIZ-0030-2020Mixed-up of materials/componentsPotential presence of leaks from the cassettes, which may be present due to an incomplete seal of the sheeting to the cassette. A leaking cassette could lead to contamination of the sterile fluid path with micro-organisms. This may…
Jul 5, 2017Class IIZ-3041-2017Use errorBaxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis…
Jul 5, 2017Class IIZ-3040-2017Use errorBaxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis…
Dec 16, 2016Class IIZ-1056-2017Device DesignCustomer complaints received for the presence of leaks
Jun 29, 2015Class IIIZ-0291-2016Device DesignLoud operating sounds, which was unacceptable to the end users when the device was powered on. The HomeChoice devices continue to perform as intended, as the issue is only associated with the noise coming from the device and not the…
Dec 8, 2014Class IIZ-1039-2015Device DesignBaxter is issuing a recall for one lot of Integrated Automated Peritoneal Dialysis set with Cassette 3 prong due to complaints received for leakage.
Oct 21, 2014Class IIZ-1031-2015Nonconforming Material/ComponentThe keypad buttons on HomeChoice devices may be activated without the operator pressing them.
Oct 6, 2014Class IIZ-0939-2015Labeling designThere are additional and updated warning and cautions that are not in the Patient At-Home Guide for HomeChoice Automated PD Systems and HomeChoice PRO Automated PD Systems.
Apr 3, 2014Class IIZ-1245-2015Error in labelingAdditional warning in the AMIA APD System labeling for patients vulnerable to unintended Increased Intraperitoneal Volume (IIPV), also known as overfill. The patients include neonates, infants, small children, and any patient with…
Oct 16, 2013Class IIZ-1292-2015Process designThe occluder in affected devices 1) may have sharp edges that could cut the pneumatic tubing that connects the occluder blade to the manifold causing an unrecoverable system error, and/or 2) blade may separate from the back plate and…
Oct 7, 2013Class IIZ-2609-2016Labeling designIn systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial Drain. This is to…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 1 company; a new generation was cleared every 5.1 years on median.
  • 3 of the 3 ancestors have a recall of any class.
  • Of its descendants, K201867 (Baxter Healthcare Corporation, 2020) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K053512.

How does its category compare?

All FKX devices

FKX System, Peritoneal, Automatic Delivery: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · FKX

2in 2025

2011: 22013: 22014: 12015: 12017: 12018: 22020: 12022: 42025: 2
2007high 4 (2022) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FKX

209in 2022

2022: 209 days median
2007high 209 (2022)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FKX

1in 2025

2007: 22010: 52011: 22012: 22013: 12014: 42015: 22016: 12018: 32020: 12023: 22025: 1
2007high 5 (2010) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FKX

3,128in 2025

2007: 1472008: 7022009: 3402010: 4,9662011: 11,3042012: 18,0342013: 22,0932014: 28,2892015: 24,6632016: 6,1372017: 3,1272018: 5,9272019: 5,9722020: 4,5982021: 5,0762022: 4,5032023: 3,9132024: 3,7712025: 3,128
2007high 28,289 (2014) · low 147 (2007)2025

MAUDE reports naming product code FKX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—2147
20080—0702
20090—0340
20100—54,966
20112—211,304
20120—218,034
20132—122,093
20141—428,289
20151—224,663
20160—16,137
20171—03,127
20182—35,927
20190—05,972
20201—14,598
20210—05,076
2022420904,503
20230—23,913
20240—03,771
20252—13,128
20260—0—

What is it indicated for?

The HomeChoicelHomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102936: Homechoice/Homechoice Pro Automated Personal Cycler Peritoneal Dialysis System Model 5C4471, 5C8310, 5C4471R and 5C8310R.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102936

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026