510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

Baxter Personal Cycler Autommated Peritoneal Dial

K923065510(k) TraditionalCleared Mar 4, 1994617-day reviewClass II21 CFR 876.5630Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Baxter Healthcare Corp · FKX System, Peritoneal, Automatic Delivery

Cleared on Mar 4, 1994 after a 617-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 18 later devices descend from it. It has been recalled 4 times.

Decided
Mar 4, 1994
Review
617 days
Class
Class II
Product code
FKX
Recalls
4
Predicates
0

18 later devices descend from it, 6 citing it directly.

Descendants, by year cleared 2006: 1 descendant cleared2011: 2 descendants cleared2005: 1 descendant cleared2001: 1 descendant cleared2003: 1 descendant cleared2013: 2 descendants cleared2022: 3 descendants cleared2018: 2 descendants cleared2020: 1 descendant cleared2017: 1 descendant cleared2014: 1 descendant cleared2015: 1 descendant cleared2025: 1 descendant cleared1990200020102020
  1. This deviceK923065 Baxter Personal Cycler Autommated Peritoneal DialBaxter Healthcare Corp · 1994
  2. Descendant · cites it directly · 17 of its ownK012988 Homechoice Personal Cycler Peritoneal Dialysis System, Models…Baxter Healthcare Corp · 2001 · Recalled
  3. Descendant · cites it directly · 13 of its ownK043363 Fresenius Liberty CyclerFresenius Medical Care North America · 2005 · Recalled
  4. Descendant · 2 generations on · 6 of its ownK123630 Fresenius Liberty CyclerFresenius Medical Care North America, Design Cente · 2013 · Recalled
  5. Descendant · 3 generations on · 5 of its ownK141145 Liberty PDX CyclerFresenius Medical Care Renal Therapies Group, LLC · 2014
  6. Descendant · cites it directly · 5 of its ownK053512 Homechoice/Homechoice Pro Personal Cycler Peritoneal Dialysis…Baxter Healthcare Corp · 2006 · Recalled

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

2001: 12003: 12005: 12006: 12011: 22013: 22014: 12015: 12017: 12018: 22020: 12022: 32025: 1
2001high 3 (2022) · low 0 (2002)2026

Most influential descendants

Showing 5 of 18 · See all
  • K012988Homechoice Personal Cycler Peritoneal Dialysis System, Models 5C8310, 5C8302, 5C4471…Baxter Healthcare Corp · 2001 · cites it directly17descendants
  • K043363Fresenius Liberty CyclerFresenius Medical Care North America · 2005 · cites it directly13descendants
  • K123630Fresenius Liberty CyclerFresenius Medical Care North America, Design Cente · 2013 · 2 generations on6descendants
  • K141145Liberty PDX CyclerFresenius Medical Care Renal Therapies Group, LLC · 2014 · 3 generations on5descendants
  • K053512Homechoice/Homechoice Pro Personal Cycler Peritoneal Dialysis System, Models 5C4471 and…Baxter Healthcare Corp · 2006 · cites it directly5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K923065Baxter Personal Cycler Autommated Peritoneal DialBaxter Healthcare Corp1994—
+1K012988Homechoice Personal Cycler Peritoneal Dialysis System, Models 5C8310, 5C8302…Baxter Healthcare Corp20019
+1K043363Fresenius Liberty CyclerFresenius Medical Care North America20059
+2K123630Fresenius Liberty CyclerFresenius Medical Care North America, Design Cente20131
+3K141145Liberty PDX CyclerFresenius Medical Care Renal Therapies Group, LLC2014—
+1K053512Homechoice/Homechoice Pro Personal Cycler Peritoneal Dialysis System, Models…Baxter Healthcare Corp20063

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200110
200310
200510
200610
201120
201320
201410
201510
201710
201820
202010
202230
202510
FIG. 1 — Lineage of K923065, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FKX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FKX device

  1. #1 K243371 Byonyks X-1 APD Cycler; Byonyks Automated PD…score 11.2 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (18)

See all 18

6 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K012988Homechoice Personal Cycler Peritoneal Dialysis System, Models 5C8310, 5C8302, 5C4471…Baxter Healthcare Corp200117
K043363Fresenius Liberty CyclerFresenius Medical Care North America200513
K053512Homechoice/Homechoice Pro Personal Cycler Peritoneal Dialysis System, Models 5C4471 and…Baxter Healthcare Corp20065
K103220Voyager Peritoneal Dialysis SystemDeka Research & Development Corp.20113
K102936Homechoice/Homechoice Pro Automated Personal Cycler Peritoneal Dialysis System Model…Baxter Healthcare Corp20111
K031676Homechoice Apd Set with Lineo Connector, Models 5C4469Q, 5C4468Q, 5C4531Q, 5C8302Q…Baxter Healthcare Corp20030

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 6, 2012Class IIZ-2132-2012Employee errorBaxter discovered that a Baxter technical service representative was providing incorrect instructions for inspection of solution bags for a blocked frangible prior to connection to the Automated PD Set with Cassette.
Nov 20, 2007Not classifiedZ-0946-2008Device DesignUnrecorded Infusion; If the system is powered down or a power failure occurs during a fill cycle, the HomeChoice system may not record the last 1-4 pump strokes infused into the patient when the power is restored. This situation may…
Jun 21, 2005Not classifiedZ-1018-05OtherThe screws that secure the Power Entry Module (PEM) to the base of the HomeChoice device are not reliably grounded to the device's ground system, posing an electric shock hazard.
May 24, 2004Not classifiedZ-1012-04OtherA software defect could result in a patient experiencing an overfill with software versions 8.5 and higher. The possibility of an overfill can occur when the Initial Drain Alarm Volume is set to OFF and a change to the volume is made at…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K012988 (Baxter Healthcare Corp, 2001) has gone furthest, with 17 descendants of its own.

How does its category compare?

All FKX devices

FKX System, Peritoneal, Automatic Delivery: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FKX

2in 2025

2011: 22013: 22014: 12015: 12017: 12018: 22020: 12022: 42025: 2
2007high 4 (2022) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FKX

209in 2022

2022: 209 days median
2007high 209 (2022)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FKX

1in 2025

2007: 22010: 52011: 22012: 22013: 12014: 42015: 22016: 12018: 32020: 12023: 22025: 1
2007high 5 (2010) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FKX

3,128in 2025

2007: 1472008: 7022009: 3402010: 4,9662011: 11,3042012: 18,0342013: 22,0932014: 28,2892015: 24,6632016: 6,1372017: 3,1272018: 5,9272019: 5,9722020: 4,5982021: 5,0762022: 4,5032023: 3,9132024: 3,7712025: 3,128
2007high 28,289 (2014) · low 147 (2007)2025

MAUDE reports naming product code FKX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—2147
20080—0702
20090—0340
20100—54,966
20112—211,304
20120—218,034
20132—122,093
20141—428,289
20151—224,663
20160—16,137
20171—03,127
20182—35,927
20190—05,972
20201—14,598
20210—05,076
2022420904,503
20230—23,913
20240—03,771
20252—13,128
20260—0—
Cite this page

510(k) Report. “K923065: Baxter Personal Cycler Autommated Peritoneal Dial.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K923065

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026