V60 Ventilator with PPV and Auto-Trak+ Software Options

K102985510(k) TraditionalCleared Apr 5, 2011180-day reviewClass II21 CFR 868.5895Recalls 9 (8 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Respironics, Inc. · MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator

Cleared on Apr 5, 2011 as substantially equivalent to K082660 (Respironics California, Inc., 2009) after a 180-day review. Its family tree reaches back 5 generations to 1983. 3 of the 23 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 9 times, including 8 Class I recalls.

Decided
Apr 5, 2011
Review
180 days
Class
Class II
Product code
MNT
Recalls
9 (8 Class I)
Predicates
1

Its longest line of descent runs back 5 generations to K833786 (1983). 18 ancestors in all, the oldest K811102 (1981). 4 later devices descend from it, 2 citing it directly. 3 of the 23 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2007: 1 other ancestor1998: 1 other ancestor, 1 with a Class I recall12002: 1 other ancestor2001: 3 other ancestors2000: 2 other ancestors1997: 1 other ancestor1995: 1 other ancestor1991: 1 other ancestor1988: 1 other ancestor1981: 1 other ancestor2023: 1 descendant cleared2022: 1 descendant cleared2026: 1 descendant cleared2025: 1 descendant clearedK833786 · Microprocessor Ventilator 7200 · Puritan Bennett Corp. · 1983×2: 2 generations cleared close together, K951264 (1996), K955324 (1996) K982454 · Bipap Vision Ventilatory Support · Respironics, Inc. · 1998 · recalledK082660 · V60 Ventilator, Model V8000 · Respironics California, Inc. · 2009 · Class I recall19801990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors, 1 with a Class I recall13 generations back: 3 other ancestors4 generations back: 7 other ancestors5 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 2 descendants K082660 · V60 Ventilator, Model V8000 · Respironics California, Inc. · 2009 · Class I recallK982454 · Bipap Vision Ventilatory Support · Respironics, Inc. · 1998 · recalledK955324 · Bipap S/T-D 30 System · Respironics, Inc. · 1996K951264 · Bipap Model S/T-D · Respironics, Inc. · 1996K833786 · Microprocessor Ventilator 7200 · Puritan Bennett Corp. · 1983−50
  1. Root · 5 generations backK833786 Microprocessor Ventilator 7200Puritan Bennett Corp. · 1983
  2. 4 generations backK951264 Bipap Model S/T-DRespironics, Inc. · 1996
  3. 3 generations backK955324 Bipap S/T-D 30 SystemRespironics, Inc. · 1996
  4. 2 generations backK982454 Bipap Vision Ventilatory SupportRespironics, Inc. · 1998 · Recalled
  5. 1 generation backK082660 V60 Ventilator, Model V8000Respironics California, Inc. · 2009 · Recalled (Class I)
  6. This deviceK102985 V60 Ventilator with PPV and Auto-Trak+ Software OptionsRespironics, Inc. · 2011
  7. Descendant · cites it directly · 2 of its ownK213521 Nihon Kohden NKV-330 Ventilator SystemNihon Kohden Orangemed, Inc. · 2022
  8. Descendant · 2 generations on · 0 of its ownK252182 SV70 Ventilator (SV70)Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2026
  9. Descendant · 2 generations on · 0 of its ownK242931 HFT750UMekics Co., Ltd. · 2025
  10. Descendant · cites it directly · 0 of its ownK230173 Servo-air Lite Ventilator SystemMaquet Critical Care AB · 2023

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K213521Nihon Kohden NKV-330 Ventilator SystemNihon Kohden Orangemed, Inc. · 2022 · cites it directly2descendants
  • K252182SV70 Ventilator (SV70)Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · 2026 · 2 generations on0descendants
  • K242931HFT750UMekics Co., Ltd. · 2025 · 2 generations on0descendants
  • K230173Servo-air Lite Ventilator SystemMaquet Critical Care AB · 2023 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K833786Microprocessor Ventilator 7200Puritan Bennett Corp.1983—
−4K951264Bipap Model S/T-DRespironics, Inc.1996—
−3K955324Bipap S/T-D 30 SystemRespironics, Inc.1996—
−2K982454Bipap Vision Ventilatory SupportRespironics, Inc.19982
−1K082660V60 Ventilator, Model V8000Respironics California, Inc.2009Class I
0 (this device)K102985V60 Ventilator with PPV and Auto-Trak+ Software OptionsRespironics, Inc.2011—
+1K213521Nihon Kohden NKV-330 Ventilator SystemNihon Kohden Orangemed, Inc.2022—
+2K252182SV70 Ventilator (SV70)Shenzhen Mindray Bio-Medical Electronics Co., Ltd.2026—
+2K242931HFT750UMekics Co., Ltd.2025—
+1K230173Servo-air Lite Ventilator SystemMaquet Critical Care AB2023—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198110
198810
199110
199510
199710
199811
200020
200130
200210
200710

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202210
202310
202510
202610
FIG. 1 — Lineage of K102985, 1983–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MNT device

  1. #1 K233643 F&P Airvo 3 NIV (PT311US)score 13.5 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K082660V60 Ventilator, Model V8000
Summary page 1
…Devices: Manufacturer Device Name 510(k) Number Respironics California, Inc. V60 Ventilator K082660 Intended Use of the Device: The V60 Ventilator with Proportional Pressure Ventilation (PPV) and Auto-Trak+ Software Options is a…
Respironics California, Inc.2009Class I

Also cleared against K082660: 2 other devices →

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K213521Nihon Kohden NKV-330 Ventilator SystemNihon Kohden Orangemed, Inc.20222
K230173Servo-air Lite Ventilator SystemMaquet Critical Care AB20230

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 25, 2023Class IZ-2631-2023Nonconforming Material/ComponentPower Management PCBAs may malfunction, causing a power failure leading to ventilator loss of function.
Feb 28, 2022Class IZ-0908-2022Device DesignVentilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.
Feb 28, 2022Class IZ-0907-2022Device DesignVentilator units have an issue impacting an internal electrical circuit within the ventilators, that in some cases could result in the ventilator ceasing to operate.
Jan 24, 2022Class IZ-0662-2022Process controlA limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient.…
Jan 24, 2022Class IZ-0663-2022Process controlA limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient.…
Apr 24, 2017Class IZ-2061-2017Device DesignOver time, low-frequency vibrations can cause the pins within the female connectors on the Motor Controller to Data Acquisition Board ribbon cable to become partially displaced causing momentary high resistance that prevents or restricts…
Feb 1, 2017Class IIZ-1687-2017Device DesignThe V60 Ventilator with Version 2.20 software installed may falsely detect that the blower motor has stalled. If this condition occurs, the software will cause the ventilator to shut down (Vent Inop) and display Error Code 100E.…
Jun 4, 2013Class IZ-1537-2013Software designRespironics California, Inc. has initiated a recall on the V60 ventilator because of an issue with the software on the V60 Power Management Board Assembly. If a component fails on the Power Management Board Assembly, it may cause…
Jul 31, 2012Class IZ-2260-2012Nonconforming Material/ComponentRespironics is recalling the V60 ventilator because certain blower motor assemblies in the V60 ventilators may not meet specifications. Specifically, the impeller may not be properly seated on the blower motor shaft. This could…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K082660V60 Ventilator, Model V8000Respironics California, Inc. · 2009 · Direct predicate · ◆ on the line of descent in FIG. 13 Class I recalls
  • K981072Esprit VentilatorRespironics, Inc. · 1998 · 2 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 28 years and passes through 3 companies; a new generation was cleared every 2.2 years on median.
  • 7 of the 18 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K213521 (Nihon Kohden Orangemed, Inc., 2022) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K082660.

How does its category compare?

All MNT devices

MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · MNT

1in 2025

2008: 12009: 22011: 22012: 22013: 12017: 12020: 12022: 12023: 12024: 22025: 12026 (partial year): 1
2007high 2 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MNT

Too few MNT 510(k)s in any year since 2007 to give a median.

Recalls per year · MNT

5in 2022

2007: 12009: 12010: 12019: 1, 1 Class I2020: 332022: 5, 5 Class I
2007high 33 (2020) · low 0 (2008)2026

Class I (6 since 2007) Class II/III not classified

Adverse event reports per year · MNT

3,545in 2025

2007: 592008: 762009: 292010: 632011: 1702012: 2572013: 3222014: 8872015: 1,6572016: 2,5152017: 2,8782018: 2,2032019: 9,3502020: 4,1702021: 3,7902022: 3,2772023: 4,5652024: 4,8082025: 3,545
2007high 9,350 (2019) · low 29 (2009)2025

MAUDE reports naming product code MNT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—159
20081—076
20092—129
20100—163
20112—0170
20122—0257
20131—0322
20140—0887
20150—01,657
20160—02,515
20171—02,878
20180—02,203
20190—1 (1)9,350
20201—334,170
20210—03,790
20221—5 (5)3,277
20231—04,565
20242—04,808
20251—03,545
20261—0—

What is it indicated for?

The V60 Ventilator with Proportional Pressure Ventilation and Auto-Trak+ Software Options is an assist ventilator and is intended to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation: patients with respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional setting under the direction of a physician. The ventilator is intended to…

Read the full statement (131 words)

The V60 Ventilator with Proportional Pressure Ventilation and Auto-Trak+ Software Options is an assist ventilator and is intended to augment patient breathing. It is intended for spontaneously breathing individuals who require mechanical ventilation: patients with respiratory failure, chronic respiratory insufficiency, or obstructive sleep apnea in a hospital or other institutional setting under the direction of a physician.

The ventilator is intended to support pediatric patients weighing 20 kg (44 Ibs) or greater to adult patients. It is also intended for intubated patients meeting the same selection criteria as the noninvasive applications. The ventilator is intended to be used by qualified medical professionals, such as physicians, nurses, and respiratory therapists.

The ventilator is intended to be used only with various combinations of Respironics recommended patient circuits, interfaces (masks), humidifiers and other accessories.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102985: V60 Ventilator with PPV and Auto-Trak+ Software Options.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K102985

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.