Esprit Ventilator

K981072510(k) TraditionalCleared Nov 12, 1998233-day reviewClass II21 CFR 868.5895Recalls 11 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Respironics, Inc. · CBK Ventilator, Continuous, Facility Use

Cleared on Nov 12, 1998 after a 233-day review. The FDA summary names its predicate without a number. 45 of the 146 devices in its family, counting itself, have a Class I recall. 145 later devices descend from it. It has been recalled 11 times, including a Class I recall.

Decided
Nov 12, 1998
Review
233 days
Class
Class II
Product code
CBK
Recalls
11 (1 Class I)
Predicates
0

145 later devices descend from it, 10 citing it directly. 45 of the 146 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2004: 4 descendants cleared, 1 with a Class I recall12007: 4 descendants cleared, 1 with a Class I recall12012: 14 descendants cleared, 4 with a Class I recall42003: 1 descendant cleared2009: 8 descendants cleared, 3 with a Class I recall32010: 6 descendants cleared, 2 with a Class I recall22005: 1 descendant cleared2011: 13 descendants cleared, 8 with a Class I recall82006: 3 descendants cleared2014: 6 descendants cleared, 1 with a Class I recall12008: 6 descendants cleared, 1 with a Class I recall12026: 7 descendants cleared2017: 10 descendants cleared, 4 with a Class I recall42016: 8 descendants cleared, 3 with a Class I recall32019: 8 descendants cleared, 3 with a Class I recall32023: 3 descendants cleared, 1 with a Class I recall12022: 2 descendants cleared2015: 8 descendants cleared, 1 with a Class I recall12024: 6 descendants cleared2013: 10 descendants cleared, 4 with a Class I recall42025: 4 descendants cleared2018: 3 descendants cleared, 1 with a Class I recall12021: 6 descendants cleared, 3 with a Class I recall32020: 4 descendants cleared, 3 with a Class I recall31995200020052010201520202025
  1. This deviceK981072 Esprit VentilatorRespironics, Inc. · 1998
  2. Descendant · cites it directly · 114 of its ownK032451 Bird Tbird Vela VentilatorBird Products Corp. · 2004 · Recalled
  3. Descendant · 2 generations on · 112 of its ownK052554 Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2005
  4. Descendant · 2 generations on · 87 of its ownK053270 IVENT 201Versamed Medical Systems, Inc. · 2006
  5. Descendant · 3 generations on · 86 of its ownK061627 Modification to I Vent 201Versamed Medical Systems, Inc. · 2006
  6. Descendant · 4 generations on · 61 of its ownK063404 Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007

Showing 5 of 145 descendants. See all descendants

Descendants cleared per year

2003: 12004: 42005: 12006: 32007: 42008: 62009: 82010: 62011: 132012: 142013: 102014: 62015: 82016: 82017: 102018: 32019: 82020: 42021: 62022: 22023: 32024: 62025: 42026: 7
2003high 14 (2012) · low 1 (2003)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 145 · See all
  • K032451Bird Tbird Vela VentilatorBird Products Corp. · 2004 · cites it directly114descendants
  • K052554Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc. · 2005 · 2 generations on112descendants
  • K053270IVENT 201Versamed Medical Systems, Inc. · 2006 · 2 generations on87descendants
  • K061627Modification to I Vent 201Versamed Medical Systems, Inc. · 2006 · 3 generations on86descendants
  • K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB · 2007 · 4 generations on61descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K981072Esprit VentilatorRespironics, Inc.1998—
+1K032451Bird Tbird Vela VentilatorBird Products Corp.20041
+2K052554Modification to: Ivent 201 Portable VentilatorVersamed Medical Systems, Inc.2005—
+2K053270IVENT 201Versamed Medical Systems, Inc.2006—
+3K061627Modification to I Vent 201Versamed Medical Systems, Inc.2006—
+4K063404Servo-I Ventilator System, Models 64, 87 and 800Maquet Critical Care AB2007—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200441
200510
200630
200741
200861
200983
201062
2011138
2012144
2013104
201461
201581
201683
2017104
201831
201983
202043
202163
202220
202331
202460
202540
202670
FIG. 1 — Lineage of K981072, 1998–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

45 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CBK device

  1. #1 K252663 Inbentus Versatilescore 16.0 · 17 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—7200 ventilator
Named without a submission number
—
*—BIRD Tbird
Named without a submission number
—
*—Siemens SY300
Named without a submission number
—

What descends from it? (145)

See all 145

10 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K032451Bird Tbird Vela VentilatorBird Products Corp.2004114
K022750PLV Continuum Ventilator, Model P1000Respironics California, Inc.200329
K082660V60 Ventilator, Model V8000 Class IRespironics California, Inc.200914
K072450Esprit Ventilator Auto-Trak Sensitivity Option, Model V1000Respironics California, Inc.20076
K102054V200 VentilatorRespironics California, Inc.20102
K034032PLV Continuum Ventilator, Model P2000Respironics California, Inc.20041
K034040Esprit Ventilator, Model V1000 Class IRespironics, Inc.20041
K110083V200 Ventilator with Aprv Mode, Esprit Ventilator with Aprv ModeRespironics California, Inc.20120
K071212Esprit Ventilator Speaking Mode Option, Model V1000Respironics California, Inc.20070
K041412Esprit Ventilator with Nico-Esprit Interface Option, Model V1000Respironics California, Inc.20040

Has anything in its family been recalled? (45 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 29, 2014Class IIZ-0485-2015Device DesignIf the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the ventilator to switch to battery power if…
Sep 17, 2014Class IZ-0008-2015Unknown/Undetermined by firmA failure of a specific component may prevent the ventilators from operating on AC power or transitioning back to AC power when operating on battery. Additionally, if a battery is not present or is depleted the ventilators will not…
Apr 22, 2009Not classifiedZ-1813-2009Device DesignThe recall was initiated because through routine product monitoring Respironics discovered that under certain conditions, when operating on AC line power, a portion of the current from the External Battery charger is erroneously diverted…
Jan 25, 2008Not classifiedZ-2002-2008Device DesignThis action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize…
Jan 25, 2008Not classifiedZ-2003-2008Device DesignThis action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize…
Jan 25, 2008Not classifiedZ-2004-2008Device DesignThis action is being taken to address power supply failures on the Esprit Ventilator which have occurred in some units. These failures, which were discovered through routine product monitoring, have been reported in countries that utilize…
Oct 29, 2007Class IIZ-2084-2013Device DesignThe power to the Esprit Display's Backlight is interrupted, causing the GUI to flicker or go dark and making the display unreadable.
Aug 8, 2006Not classifiedZ-0047-2007OtherThis action is being initiated due to power supply snubber board failures on certain Esprit Ventilators in countries that utilize operating voltages of 200-240 Volts AC and will involve the replacement of the Esprit Ventilator power…
Feb 20, 2002Not classifiedZ-1230-04OtherFlow sensors would exhibit drift.
Jun 8, 2001Not classifiedZ-0951-03OtherMaterial hardness on original design check valve may cause premature failure.
Jun 1, 2001Not classifiedZ-0903-04OtherVentilator would spontaneously suspend ventilatory support to patients due to a hardware/software compatibility issue.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K034040Esprit Ventilator, Model V1000Respironics, Inc. · 2004 · Cites it directly · ◆ in the 2004 descendants bar of FIG. 11 Class I recall
  • K082660V60 Ventilator, Model V8000Respironics California, Inc. · 2009 · Cites it directly · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls
  • K083526Trilogy 100 VentilatorRespironics Inc., Sleep & Home Respiratory Group · 2009 · 2 generations on · ◆ in the 2009 descendants bar of FIG. 15 Class I recalls
  • K070594Palmtop Ventilator, Models PTV-8, PTV-10Pulmonetic Systems, Inc. · 2007 · 2 generations on · ◆ in the 2007 descendants bar of FIG. 12 Class I recalls
  • K113053Bipap a 30 Ventilatory Support SystemRespironics, Inc. · 2012 · 2 generations on · ◆ in the 2012 descendants bar of FIG. 15 Class I recalls
  • K102985V60 Ventilator with PPV and Auto-Trak+ Software OptionsRespironics, Inc. · 2011 · 2 generations on · ◆ in the 2011 descendants bar of FIG. 18 Class I recalls
  • K090888Newport HT70 Family of VentilatorsNewport Medical Instruments, Inc. · 2009 · 3 generations on · ◆ in the 2009 descendants bar of FIG. 13 Class I recalls
  • K093261MOVESThornhill Research, Inc. · 2010 · 3 generations on · ◆ in the 2010 descendants bar of FIG. 12 Class I recalls
  • K162877VOCSN Unified Respiratory SystemVentec Life Systems, Inc. · 2017 · 3 generations on · ◆ in the 2017 descendants bar of FIG. 114 Class I recalls
  • K162832731 Series VentilatorsZOLL Medical Corporation · 2017 · 3 generations on · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K093416Trilogy 200 VentilatorRespironics, Inc. · 2010 · 3 generations on · ◆ in the 2010 descendants bar of FIG. 14 Class I recalls
  • K111610Trilogy Series Ventilator with OximetryRespironics, Inc. · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 12 Class I recalls

Nearest 12 of 44 shown.

What does this lineage say?

  • Of its descendants, K032451 (Bird Products Corp., 2004) has gone furthest, with 114 descendants of its own.

How does its category compare?

All CBK devices

CBK Ventilator, Continuous, Facility Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CBK

4in 2025

2007: 112008: 182009: 122010: 142011: 132012: 172013: 52014: 52015: 72016: 42017: 62018: 42019: 82020: 52021: 62023: 52024: 92025: 42026 (partial year): 3
2007high 18 (2008) · low 0 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CBK

163in 2025

2007: 142 days median2008: 115 days median2009: 122 days median2010: 150 days median2011: 192 days median2012: 207 days median2013: 277 days median2014: 439 days median2015: 266 days median2016: 323 days median2017: 280 days median2018: 183 days median2019: 435 days median2020: 176 days median2021: 267 days median2023: 309 days median2024: 203 days median2025: 163 days median2026 (partial year): 235 days median
2007high 439 (2014) · low 115 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CBK

14in 2024

2007: 172008: 232009: 82010: 92011: 122012: 2, 1 Class I2013: 5, 2 Class I2014: 5, 4 Class I2015: 6, 4 Class I2016: 4, 1 Class I2017: 12018: 3, 3 Class I2019: 14, 13 Class I2020: 42021: 21, 19 Class I2022: 8, 4 Class I2023: 4, 3 Class I2024: 14, 14 Class I2026 (partial year): 15, 13 Class I
2007high 23 (2008) · low 0 (2025)2026

Class I (81 since 2007) Class II/III not classified

Adverse event reports per year · CBK

13,991in 2025

2007: 1,5312008: 1,9332009: 1,7662010: 1,7992011: 1,8772012: 5,5092013: 7,9212014: 7,6002015: 11,8762016: 16,0562017: 14,4512018: 9,8492019: 9,1002020: 11,0962021: 13,3222022: 15,9742023: 24,9282024: 20,3362025: 13,991
2007high 24,928 (2023) · low 1,531 (2007)2025

MAUDE reports naming product code CBK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711142171,531
200818115231,933
20091212281,766
20101415091,799
201113192121,877
2012172072 (1)5,509
201352775 (2)7,921
201454395 (4)7,600
201572666 (4)11,876
201643234 (1)16,056
20176280114,451
201841833 (3)9,849
2019843514 (13)9,100
20205176411,096
2021626721 (19)13,322
20220—8 (4)15,974
202353094 (3)24,928
2024920314 (14)20,336
20254163013,991
2026323515 (13)—

What is it indicated for?

The ESPRIT ventilator is a microprocessor controlled, electrically powered, mechanical ventilator. It is intended for use by qualified medical personnel in providing continuous or intermittent ventilatory support for adult and pediatric patients as prescribed by a physician. The ESPRIT Ventilator is intended for use in either invasive or non-invasive applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K981072: Esprit Ventilator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K981072

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026