510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Medivators Advantage Plus Endoscope Reprocessor

K102996510(k) TraditionalCleared Jan 13, 201198-day reviewClass II21 CFR 876.1500Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medivators Reprocessing Systems · FEB Accessories, Cleaning, For Endoscope

Cleared on Jan 13, 2011 as substantially equivalent to K061899 (Advanced Sterilization Products, 2006) after a 98-day review. Its family tree reaches back 3 generations to 1994. 1 later device descends from it. It has been recalled 4 times.

Decided
Jan 13, 2011
Review
98 days
Class
Class II
Product code
FEB
Recalls
4
Predicates
1

Its longest line of descent runs back 3 generations to K914145 (1994). 4 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 3 generations backK914145 DSD-91(TM) Disinfector for Flexible EndoscopesMedivators, Inc. · 1994 · Recalled
  2. 2 generations backK040883 Evotech Integrated Endoscope Disinfection SystemAdvanced Sterilization Products · 2005 · Recalled
  3. 1 generation backK061899 Evotech Integrated Endoscope Disinfection SystemAdvanced Sterilization Products · 2006 · Recalled
  4. This deviceK102996 Medivators Advantage Plus Endoscope ReprocessorMedivators Reprocessing Systems · 2011
  5. Descendant · cites it directly · 0 of its ownK172677 Advantage Plus Pass-ThruMedivators, Inc. · 2017 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K172677Advantage Plus Pass-ThruMedivators, Inc. · 2017 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K914145DSD-91(TM) Disinfector for Flexible EndoscopesMedivators, Inc.19941
−2K040883Evotech Integrated Endoscope Disinfection SystemAdvanced Sterilization Products20053
−1K061899Evotech Integrated Endoscope Disinfection SystemAdvanced Sterilization Products20062
0 (this device)K102996Medivators Advantage Plus Endoscope ReprocessorMedivators Reprocessing Systems2011—
+1K172677Advantage Plus Pass-ThruMedivators, Inc.20171

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
FIG. 1 — Lineage of K102996, 1994–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FEB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FEB device

  1. #1 K243433 enspire 3000 Cleaning and Liquid Chemical…score 10.8 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061899Evotech Integrated Endoscope Disinfection System
Summary page 2
…e function has not been changed. It has been shown to be equivalent to the Advanced Sterilization Products' EvoTechTM Integrated Endoscope Disinfection System (K06 1899). The machines have the same indications for use and the same methods of providing disinfection. The ASP Evoech system…
Advanced Sterilization Products20062

Also cleared against K061899: 3 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K172677Advantage Plus Pass-ThruMedivators, Inc.20170

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 13, 2024Class IIZ-2537-2024Error in labelingIFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may…
Jun 13, 2024Class IIZ-2538-2024Error in labelingIFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may…
Jun 13, 2024Class IIZ-2536-2024Error in labelingIFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may…
Nov 13, 2023Class IIZ-0617-2024Process controlIdentification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately result in postponement…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 3 companies; a new generation was cleared every 4.2 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K172677 (Medivators, Inc., 2017) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K061899.

How does its category compare?

All FEB devices

FEB Accessories, Cleaning, For Endoscope: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · FEB

2in 2024

2007: 22008: 12009: 12010: 22011: 32012: 12013: 12014: 12016: 22017: 22018: 12019: 22020: 22022: 12023: 12024: 22026 (partial year): 1
2007high 3 (2011) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FEB

111in 2011

2011: 111 days median
2007high 111 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FEB

15so far in 2026, the most since 2007

2008: 32010: 12016: 22020: 12021: 12023: 12024: 12025: 32026 (partial year): 15
2007high 15 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FEB

356in 2025

2007: 392008: 1522009: 202010: 502011: 322012: 162013: 272014: 482015: 922016: 1032017: 692018: 882019: 622020: 4262021: 5752022: 6072023: 9982024: 8842025: 356
2007high 998 (2023) · low 16 (2012)2025

MAUDE reports naming product code FEB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—039
20081—3152
20091—020
20102—150
20113111032
20121—016
20131—027
20141—048
20150—092
20162—2103
20172—069
20181—088
20192—062
20202—1426
20210—1575
20221—0607
20231—1998
20242—1884
20250—3356
20261—15—

What is it indicated for?

Medivators Advantage Plus Endoscope Reprocessing System tests, cleans, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The Advantage Plus system is indicated to provide high level disinfection of beat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is not required prior to placement in the Advantage Plus system. The scopes must be precleaned immediately after use.…

Read the full statement (167 words)

Medivators Advantage Plus Endoscope Reprocessing System tests, cleans, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The Advantage Plus system is indicated to provide high level disinfection of beat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is not required prior to placement in the Advantage Plus system. The scopes must be precleaned immediately after use.

The Advantage Plus Endoscope Reprocessing System uses Rapicide PA High Level Disinfectant to provide high level disinfection of endoscopes when used according to the directions for use. The system uses Intercept Detergent in its cleaning cycle at a concentration of 0.5%.

Rapicide PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is above the minimum recommended concentration (MRC) of B5Oppm peracetic acid; this ensures that the disinfectant was above MRC during the entire disinfection cycle. Rapicide PA should be used in the Advantage Plus System under the following contact conditions:

Contact Time: 5 minutes

Temperature: 300C

MRC: S5Oppm

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K102996: Medivators Advantage Plus Endoscope Reprocessor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K102996

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026