510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Brivo XR385

K103448510(k) TraditionalCleared Aug 12, 2011261-day reviewClass II21 CFR 892.1680Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, LLC · KPR System, X-Ray, Stationary

Cleared on Aug 12, 2011 as substantially equivalent to K982955 (GE Medical Systems, 1998) and 2 other predicates after a 261-day review. Its family tree reaches back 4 generations to 1998. 4 later devices descend from it. It has been recalled once.

Decided
Aug 12, 2011
Review
261 days
Class
Class II
Product code
KPR
Recalls
1
Predicates
3

Its longest line of descent runs back 4 generations to K981556 (1998). 9 ancestors in all, the oldest K862120 (1986). 4 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2004: 1 other ancestor1998: 2 other ancestors2002: 1 other ancestor1986: 1 other ancestor2014: 1 descendant cleared2017: 1 descendant cleared2020: 2 descendants clearedK981556 · Canon X-Ray Digital Camera Cxdi-11 · Canon USA, Inc. · 1998K992547 · Canon X-Ray Digital Camera, Model Cxdi-22 · Canon USA, Inc. · 1999K023750 · Cannon Digital Radiography Model Cxdi-40G · Canon USA, Inc. · 2002K031447 · Canon, Digital Radiography, Model Cxdi-50G · Canon, Inc. · 20031990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 2 other ancestors3 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 1 descendant K031447 · Canon, Digital Radiography, Model Cxdi-50G · Canon, Inc. · 2003K023750 · Cannon Digital Radiography Model Cxdi-40G · Canon USA, Inc. · 2002K992547 · Canon X-Ray Digital Camera, Model Cxdi-22 · Canon USA, Inc. · 1999K981556 · Canon X-Ray Digital Camera Cxdi-11 · Canon USA, Inc. · 19980
  1. Root · 4 generations backK981556 Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc. · 1998
  2. 3 generations backK992547 Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc. · 1999
  3. 2 generations backK023750 Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc. · 2002
  4. 1 generation backK031447 Canon, Digital Radiography, Model Cxdi-50GCanon, Inc. · 2003
  5. This deviceK103448 Brivo XR385Ge Medical Systems, LLC · 2011
  6. Descendant · cites it directly · 3 of its ownK141381 DuradiagnostPhilips Healthcare (Suzhou) Co., Ltd. · 2014 · Recalled
  7. Descendant · 2 generations on · 1 of its ownK163410 DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Ltd. · 2017 · Recalled
  8. Descendant · 3 generations on · 0 of its ownK201725 DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Lit. · 2020 · Recalled
  9. Descendant · 2 generations on · 0 of its ownK201640 DuraDiagnostPhilips Healthcare (Suzhou) Co., Ltd. · 2020 · Recalled

Most influential descendants

Showing 4 of 4 · See all
  • K141381DuradiagnostPhilips Healthcare (Suzhou) Co., Ltd. · 2014 · cites it directly3descendants
  • K163410DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Ltd. · 2017 · 2 generations on1descendant
  • K201725DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Lit. · 2020 · 3 generations on0descendants
  • K201640DuraDiagnostPhilips Healthcare (Suzhou) Co., Ltd. · 2020 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K981556Canon X-Ray Digital Camera Cxdi-11Canon USA, Inc.1998—
−3K992547Canon X-Ray Digital Camera, Model Cxdi-22Canon USA, Inc.1999—
−2K023750Cannon Digital Radiography Model Cxdi-40GCanon USA, Inc.2002—
−1K031447Canon, Digital Radiography, Model Cxdi-50GCanon, Inc.2003—
0 (this device)K103448Brivo XR385Ge Medical Systems, LLC2011—
+1K141381DuradiagnostPhilips Healthcare (Suzhou) Co., Ltd.20147
+2K163410DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Ltd.20174
+3K201725DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Lit.20203
+2K201640DuraDiagnostPhilips Healthcare (Suzhou) Co., Ltd.20202

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199820
200210
200410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201710
202020
FIG. 1 — Lineage of K103448, 1998–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPR device

  1. #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982955Silhouette VR
Summary page 2
…ray imager (2ICFR § 892.1650) Product Code: KPR and MQB Predicate Device(s): Stationary x-ray system MODEL: Silhouette VR (K982955) GE Tethered Portable Digital Radiographic Detector (K041922) Canon d…
GE Medical Systems19981
*K041922GE Tethered Portable Digital Radiographic Detector
Summary page 2
…B Predicate Device(s): Stationary x-ray system MODEL: Silhouette VR (K982955) GE Tethered Portable Digital Radiographic Detector (K041922) Canon digital radiography CXDI-50G (K03 1447) Device Description: The Brivo XR385 Digita…
General Electric Co.2004—
*K031447Canon, Digital Radiography, Model Cxdi-50G
Summary page 2
…VR (K982955) GE Tethered Portable Digital Radiographic Detector (K041922) Canon digital radiography CXDI-50G (K03 1447) Device Description: The Brivo XR385 Digital Diagnostic Radiographic system provides stat…
Canon, Inc.2003—

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K141381DuradiagnostPhilips Healthcare (Suzhou) Co., Ltd.20143

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 16, 2016Class IIZ-2523-2018Component change controlThe placing of weight on the VRAD detector, of the Brivo XR385 digital radiographic system, can introduce image artifacts, which can result in patient exam retakes.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 3 companies; a new generation was cleared every 3.1 years on median.
  • 2 of the 9 ancestors have a recall of any class.
  • Of its descendants, K141381 (Philips Healthcare (Suzhou) Co., Ltd., 2014) has gone furthest, with 3 descendants of its own.

How does its category compare?

All KPR devices

KPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · KPR

8in 2025

2007: 122008: 152009: 92010: 92011: 92012: 112013: 52014: 102015: 182016: 112017: 62018: 152019: 122020: 162021: 62022: 82023: 92024: 112025: 82026 (partial year): 6
2007high 18 (2015) · low 5 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPR

155in 2025

2007: 31 days median2008: 48 days median2009: 35 days median2010: 126 days median2011: 152 days median2012: 58 days median2013: 163 days median2014: 108 days median2015: 81 days median2016: 84 days median2017: 75 days median2018: 30 days median2019: 33 days median2020: 39 days median2021: 55 days median2022: 43 days median2023: 76 days median2024: 162 days median2025: 155 days median2026 (partial year): 197 days median
2007high 197 (2026) · low 30 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPR

7in 2025

2007: 102008: 102009: 52010: 32011: 52012: 12013: 52014: 82015: 12016: 42017: 32018: 42019: 12020: 12022: 22023: 42024: 12025: 7
2007high 10 (2007) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPR

39in 2025

2007: 1202008: 952009: 612010: 822011: 1032012: 742013: 242014: 342015: 592016: 182017: 312018: 362019: 772020: 412021: 372022: 272023: 392024: 342025: 39
2007high 120 (2007) · low 18 (2016)2025

MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123110120
200815481095
2009935561
20109126382
201191525103
20121158174
20135163524
201410108834
20151881159
20161184418
2017675331
20181530436
20191233177
20201639141
2021655037
2022843227
2023976439
202411162134
20258155739
202661970—

What is it indicated for?

The Brivo XR385 Digital X-Ray System is intended for use on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patent sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K103448: Brivo XR385.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K103448

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026