DigitalDiagnost C50

K163410510(k) SpecialCleared Jan 4, 201730-day reviewClass II21 CFR 892.1680Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Healthcare (Suzhou) Co., Ltd. · KPR System, X-Ray, Stationary

Cleared on Jan 4, 2017 as substantially equivalent to K141381 (Philips Healthcare (Suzhou) Co., Ltd., 2014) after a 30-day review. Its family tree reaches back 7 generations to 2000. 1 later device descends from it. It has been recalled 4 times.

Decided
Jan 4, 2017
Review
30 days
Class
Class II
Product code
KPR
Recalls
4
Predicates
1

Its longest line of descent runs back 7 generations to K001201 (2000). 22 ancestors in all, the oldest K862120 (1986). 1 later device descends from it, 1 citing it directly.

  1. Root · 7 generations backK001201 Multix Compact K and Multix L Radiographic X-Ray SystemsSiemens Medical Solutions USA, Inc. · 2000
  2. 6 generations backK012546 Optima Urs Universal Radiographic SystemSedecal USA, Inc. · 2001 · Recalled
  3. 5 generations backK072659 Rad Vision, Models E and EUAlmana Medical Imaging · 2007
  4. 4 generations backK082064 Radvision ET Diagnostic X-Ray SystemAlmana Medical Imaging · 2008
  5. 3 generations backK083109 Suinsa Diagnostic X-Ray SystemSuinsa · 2008
  6. 2 generations backK123005 Ddrversa MotionSwissray Medical AG · 2012
  7. 1 generation backK141381 DuradiagnostPhilips Healthcare (Suzhou) Co., Ltd. · 2014 · Recalled
  8. This deviceK163410 DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Ltd. · 2017
  9. Descendant · cites it directly · 0 of its ownK201725 DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Lit. · 2020 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K201725DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Lit. · 2020 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K001201Multix Compact K and Multix L Radiographic X-Ray SystemsSiemens Medical Solutions USA, Inc.2000—
−6K012546Optima Urs Universal Radiographic SystemSedecal USA, Inc.20012
−5K072659Rad Vision, Models E and EUAlmana Medical Imaging2007—
−4K082064Radvision ET Diagnostic X-Ray SystemAlmana Medical Imaging2008—
−3K083109Suinsa Diagnostic X-Ray SystemSuinsa2008—
−2K123005Ddrversa MotionSwissray Medical AG2012—
−1K141381DuradiagnostPhilips Healthcare (Suzhou) Co., Ltd.20147
0 (this device)K163410DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Ltd.2017—
+1K201725DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Lit.20203

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198610
199420
199710
199840
199910
200220
200310
200410
201120

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
FIG. 1 — Lineage of K163410, 2000–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPR device

  1. #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K141381Duradiagnost
Summary page 6
…Trade Name: DuraDiagnost Manufacturer: Philips Healthcare (Suzhou) Co., Ltd. 510(k) Clearance: K141381 (June 12, 2014) Classification 21 CFR, Part 892.1680 Regulation:…
Philips Healthcare (Suzhou) Co., Ltd.20147
†K161459Nexus DRTM Digital X-ray Imaging System (with PaxScan 4336Wv4)Varian Medical Systems, X-Ray Products-Infimed2016—

Also cleared against K141381: 1 other device →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K201725DigitalDiagnost C50Philips Healthcare (Suzhou) Co., Lit.20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 9, 2023Class IIZ-1383-2023Equipment maintenanceIf the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear…
Mar 9, 2023Class IIZ-1382-2023Equipment maintenanceIf the PM for the telescopic carriage is not performed annually as defined in the Instructions for Use (IFU), the tension force of the spring balancer inside of the telescopic carriage will unavoidably decay through normal wear and tear…
Feb 25, 2022Class IIZ-1563-2022Software Design ChangeFollowing a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in the firmware of the…
Dec 21, 2018Class IIZ-2498-2019Radiation Control for Health and Safety ActDuring recent evaluations of the Philips DigitalDiagnost C50 system, Philips identified a missing label. This notification is intended to provide you with information regarding what the issue is, and relevant affected product, the actions…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 6 companies; a new generation was cleared every 1.5 years on median.
  • 8 of the 22 ancestors have a recall of any class.
  • Of its descendants, K201725 (Philips Healthcare (Suzhou) Co., Lit., 2020) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K141381.

How does its category compare?

All KPR devices

KPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · KPR

8in 2025

2007: 122008: 152009: 92010: 92011: 92012: 112013: 52014: 102015: 182016: 112017: 62018: 152019: 122020: 162021: 62022: 82023: 92024: 112025: 82026 (partial year): 6
2007high 18 (2015) · low 5 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPR

155in 2025

2007: 31 days median2008: 48 days median2009: 35 days median2010: 126 days median2011: 152 days median2012: 58 days median2013: 163 days median2014: 108 days median2015: 81 days median2016: 84 days median2017: 75 days median2018: 30 days median2019: 33 days median2020: 39 days median2021: 55 days median2022: 43 days median2023: 76 days median2024: 162 days median2025: 155 days median2026 (partial year): 197 days median
2007high 197 (2026) · low 30 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPR

7in 2025

2007: 102008: 102009: 52010: 32011: 52012: 12013: 52014: 82015: 12016: 42017: 32018: 42019: 12020: 12022: 22023: 42024: 12025: 7
2007high 10 (2007) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPR

39in 2025

2007: 1202008: 952009: 612010: 822011: 1032012: 742013: 242014: 342015: 592016: 182017: 312018: 362019: 772020: 412021: 372022: 272023: 392024: 342025: 39
2007high 120 (2007) · low 18 (2016)2025

MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123110120
200815481095
2009935561
20109126382
201191525103
20121158174
20135163524
201410108834
20151881159
20161184418
2017675331
20181530436
20191233177
20201639141
2021655037
2022843227
2023976439
202411162134
20258155739
202661970—

What is it indicated for?

The DigitalDiagnost C50 system is intended for use in generating radiographic images of human anatomy by qualified/trained doctor or technician. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K163410: DigitalDiagnost C50.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K163410

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026