510(k) · Cleared
Recalled · 2011Expect Endoscopic Aspiration Needle
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Boston Scientific Corp · ODG Endoscopic Ultrasound System, Gastroenterology-Urology
Cleared on Jan 12, 2011 as substantially equivalent to K100712 (Boston Scientific Corporation, 2010) after a 7-day review. Its family tree reaches back 4 generations to 1990. 26 later devices descend from it. It has been recalled once.
- Decided
- Jan 12, 2011
- Review
- 7 days
- Class
- Class II
- Product code
- ODG
- Recalls
- 1
- Predicates
- 1
Its longest line of descent runs back 4 generations to K904667 (1990). 7 ancestors in all. 26 later devices descend from it, 2 citing it directly.
- Root · 4 generations backK904667 NA NeedlesOlympus Corp. · 1990
- 3 generations backK973128 Olympus NA-10J-1 Aspiration NeedleOlympus America, Inc. · 1997
- 2 generations backK023272 Disposable Aspiration Needle NA-200HThe Olympus Optical Co. · 2002 · Recalled
- 1 generation backK100712 Resound Endoscopic Ultrasound Aspiration NeedleBoston Scientific Corporation · 2010 · Recalled
- This deviceK110030 Expect Endoscopic Aspiration NeedleBoston Scientific Corp · 2011
- Descendant · cites it directly · 16 of its ownK112198 Expect Endoscopic Ultrasound Aspiration NeedleBoston Scientific Corporation · 2011
- Descendant · 2 generations on · 11 of its ownK123676 Cellvizio 100 Series and Cellvizio Systems with Confocal MiniprobesMauna Kea Technologies · 2013
- Descendant · 3 generations on · 10 of its ownK132389 Cellvizio 100 Series System and Cellvizio System with Confocal…Daniel & Daniel Consulting · 2014
- Descendant · 4 generations on · 9 of its ownK141358 Cellvizio 100 Series with System Confocal MiniprobesMauna Kea Technologies · 2014
- Descendant · cites it directly · 8 of its ownK133312 Expect Slimline Endoscopic Ultrasound Aspiration NeedleBoston Scientific Corporation · 2014
Showing 5 of 26 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 26 · See all- K112198Expect Endoscopic Ultrasound Aspiration NeedleBoston Scientific Corporation · 2011 · cites it directly16descendants
- K123676Cellvizio 100 Series and Cellvizio Systems with Confocal MiniprobesMauna Kea Technologies · 2013 · 2 generations on11descendants
- K132389Cellvizio 100 Series System and Cellvizio System with Confocal Miniprobes; Uroflex B and…Daniel & Daniel Consulting · 2014 · 3 generations on10descendants
- K141358Cellvizio 100 Series with System Confocal MiniprobesMauna Kea Technologies · 2014 · 4 generations on9descendants
- K133312Expect Slimline Endoscopic Ultrasound Aspiration NeedleBoston Scientific Corporation · 2014 · cites it directly8descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −4 | K904667 | NA Needles | Olympus Corp. | 1990 | — |
| −3 | K973128 | Olympus NA-10J-1 Aspiration Needle | Olympus America, Inc. | 1997 | — |
| −2 | K023272 | Disposable Aspiration Needle NA-200H | The Olympus Optical Co. | 2002 | 1 |
| −1 | K100712 | Resound Endoscopic Ultrasound Aspiration Needle | Boston Scientific Corporation | 2010 | 5 |
| 0 (this device) | K110030 | Expect Endoscopic Aspiration Needle | Boston Scientific Corp | 2011 | — |
| +1 | K112198 | Expect Endoscopic Ultrasound Aspiration Needle | Boston Scientific Corporation | 2011 | — |
| +2 | K123676 | Cellvizio 100 Series and Cellvizio Systems with Confocal Miniprobes | Mauna Kea Technologies | 2013 | — |
| +3 | K132389 | Cellvizio 100 Series System and Cellvizio System with Confocal Miniprobes… | Daniel & Daniel Consulting | 2014 | — |
| +4 | K141358 | Cellvizio 100 Series with System Confocal Miniprobes | Mauna Kea Technologies | 2014 | — |
| +1 | K133312 | Expect Slimline Endoscopic Ultrasound Aspiration Needle | Boston Scientific Corporation | 2014 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1993 | 1 | 0 |
| 1994 | 1 | 0 |
| 2000 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2011 | 1 | 0 |
| 2013 | 1 | 0 |
| 2014 | 3 | 0 |
| 2015 | 3 | 0 |
| 2016 | 2 | 0 |
| 2017 | 4 | 0 |
| 2018 | 2 | 0 |
| 2019 | 1 | 0 |
| 2020 | 2 | 0 |
| 2021 | 2 | 0 |
| 2022 | 1 | 0 |
| 2024 | 1 | 0 |
| 2025 | 3 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in ODG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new ODG device
- #1 K251859 EVIS EUS Ultrasound Gastrointestinal…score 14.0 · 4 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K100712 | Resound Endoscopic Ultrasound Aspiration NeedleSummary page 2…6. Technological Characteristics: The proposed Expect TmEUS-ENA device has a modified handle design as compared to the currently cleared Expect Tm EUS-FNA device K100712. 7. Performance Data: Bench Testing has been performed on the Expectm EUS-FNA device to demonstrate that the modified ha… | 2010 |
What descends from it? (26)
See all 262 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 11, 2011 | Not classified | Z-3215-2011 | Other | The firm has determined that a single lot of 25 ga Expect EUS/FNA devices were mislabeled. The carton label incorrectly described the devices as 22 ga. The inner pouch label incorrectly describes the device as 25 ga. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 21 years and passes through 5 companies; a new generation was cleared every 6.9 years on median.
- 4 of the 7 ancestors have a recall of any class.
- Of its descendants, K112198 (Boston Scientific Corporation, 2011) has gone furthest, with 16 descendants of its own.
- 1 other device also cites its predicate K100712.
How does its category compare?
All ODG devicesODG Endoscopic Ultrasound System, Gastroenterology-Urology: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.
FDA decisions per year · ODG
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · ODG
71in 2022
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · ODG
4in 2020
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · ODG
3,569in 2025
MAUDE reports naming product code ODG, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 0 | — | 0 | 1 |
| 2009 | 2 | — | 0 | 3 |
| 2010 | 0 | — | 0 | 2 |
| 2011 | 2 | — | 0 | 13 |
| 2012 | 0 | — | 0 | 6 |
| 2013 | 2 | — | 0 | 2 |
| 2014 | 1 | — | 0 | 1 |
| 2015 | 0 | — | 0 | 11 |
| 2016 | 1 | — | 0 | 50 |
| 2017 | 4 | 75 | 0 | 82 |
| 2018 | 3 | 30 | 0 | 66 |
| 2019 | 1 | — | 0 | 96 |
| 2020 | 2 | — | 4 | 445 |
| 2021 | 1 | — | 0 | 1,234 |
| 2022 | 4 | 71 | 0 | 1,832 |
| 2023 | 1 | — | 0 | 2,078 |
| 2024 | 1 | — | 0 | 1,904 |
| 2025 | 2 | — | 0 | 3,569 |
| 2026 | 2 | — | 4 | — |
What is it indicated for?
The Expect Needle is designed to sample targeted submucosal and extramural lesions through the accessory channel of a curvilinear echoendoscope.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K110030 · printed Nov 4, 2026
Cite this page
510(k) Report. “K110030: Expect Endoscopic Aspiration Needle.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K110030
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026